1. Catalogs
  2. eralytics GmbH
  3. ERAOIL Brochure

ERAOIL Brochure

ERAOIL Brochure

ERAOIL Brochure

Product catalog summary
Introduction
The RDE-OES Spectrometer, branded as eraoil, is a cutting-edge device designed for elemental analysis of wear metals, additives, and contaminants in various operating liquids such as lubricating oils, fuels, and coolants. It is compliant with ASTM standards D6595 and D6728 and offers excellent correlation to ICP-AES (ASTM D5185).
Specifications
eraoil can measure up to 32 elements automatically within 30 seconds, with a sub-ppm detection limit for most elements. It features a high-end CMOS detector and automatic background correction. The device is designed for ease of use, even by untrained operators, and includes a built-in PC, touchscreen, electrode sharpener, and ventilation system.
Applications
The spectrometer is ideal for analyzing wear metals in lubricants from engines, transmissions, and hydraulic systems. It also serves applications in coolants, metalworking fluids, and other process fluids. The device supports multi-elemental analysis, providing significant advantages for industrial and transport operations.
Features
  • Compliant with ASTM D6595 and D6728 standards.
  • Measures up to 32 elements in 30 seconds with sub-ppm LOD.
  • Stand-alone design with built-in PC and touchscreen.
  • No sample preparation or carrier/plasma gas required.
Technical Specifications
The spectrometer operates within a temperature range of 0 °C to 40 °C and up to 90% non-condensing humidity. It is factory calibrated for 26 elements, with optional calibration for up to 32 elements. The device requires a sample volume of 2 mL and does not need solvents, reagents, or gases.
Performance and Precision
eraoil offers superior precision data, outperforming ASTM D6595 standards. It provides a wide detection range for various elements in lubricants, fuels, and coolants.
Design and Usability
The spectrometer features a compact design suitable for lab and on-site applications. It includes a multilingual 10” color touchscreen and interfaces for Ethernet, USB, and RS232. The device supports external monitor connections and data input via keyboard, mouse, or barcode reader.
Software and Data Management
eraoil uses Windows® software for data transfer, spectrum viewing, and result analysis. It can store over 100,000 detailed test reports and spectra in its internal memory.
Power and Consumables
The device operates on 90 – 270 V, AC 50/60 Hz, 500 W power. Recommended consumables for 1000 tests include graphite discs, electrodes, and sample cups.
See more

Catalog excerpts

ERAOIL Brochure-1

RDE-OES SPECTROMETER FOR ELEMENTAL ANALYSIS Standards ASTM D6595, ASTM D6728 Excellent correlation to ICP-AES ASTM D5185

 Open the catalog to page 1
ERAOIL Brochure-2

eraoil – Fast & Easy Wear Metal Analysis . Wear metals, additives, contaminants eraoil is the first and most advanced stand-alone rotating disc electrode spectrometer analyzing wear metals, additives, and contaminants in all kinds of operating liquids, like lubricating oils, fuels, heavy fuel oil (HFO), coolants, process water, etc. Up to 32 elements are measured fully automatically within 30 seconds. eraoil’s high end CMOS detector in combination with the automatic back-ground correction allow for a sub-ppm detection limit (LOD) for most elements. Smallest footprint with built-in PC eraoil’s...

 Open the catalog to page 2
ERAOIL Brochure-3

eraoil is fully compliant with ASTM D6595 (Oil) and ASTM D6728 (Fuel). ASTM D6595 is the standard for identifying wear metals and contaminants in used lubricating oils and used hydraulic fluids by rotating disc electrode atomic emission spectrometry. For determining contaminants in gas turbine and diesel engine fuel, ASTM D6728 is the standard of choice. To protect finished fuels and heavy fuel oil (HFO) against trace contamination eraoil is available with the special Low Detection Fuel Module to detect e.g. Potassium, Lithium, Sodium and Vanadium at LOD levels of as low as 0.1 ppm. Typical eraoil...

 Open the catalog to page 3
ERAOIL Brochure-5

Technical Specifications of eraoil Available Test Methods ASTM D6595: Wear metals, contaminants and additives in used lubricating oils and hydraulic fluids ASTM D6728: Contaminants in gas turbine and engine fuel Spectrometer Type Rotating Disc Electrode Spectrometer (190 - 810 nm) Thermally stabilized Paschen-Runge mounting with CMOS detectors Elemental concentrations in ppm (mg/kg) Determination of additives, wear metals and contaminants in lubricating oils, hydraulic fluids, gas turbine & diesel engine fuel, heavy fuel oil (HFO), crude oil, glycol coolants, process water, mineral water, and...

 Open the catalog to page 5
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.