Model VREF

Model VREF

Product catalog summary
Overview: The document provides information on the VREF module, a 10V calibration reference tool used for high-level devices, specifically for transferring Epsilon's calibration to customer electronics. It is compatible with Epsilon's Model 3648 high-temperature capacitive extensometers and the 3603 and 3604 signal conditioners.
Specifications: The VREF module offers a high-precision 10V reference to ensure accurate calibrations. It can be used to calibrate multiple high-level output extensometers by applying the 10V extension value for each.
Calibration Procedure:
  1. Connect the VREF module to the testing machine's strain channel input, set VREF to 0V, and zero the strain channel for the zero reference.
  2. Set VREF to 10V and calibrate the full-scale extensometer value on the strain channel for the calibrated full scale.
  3. If necessary, return VREF to 0V and zero the strain channel again.
  4. Reconnect the extensometer's signal conditioner to the testing machine.
Conclusion: The VREF module simplifies the calibration process, ensuring precise and reliable calibration for high-level output extensometers.
See more

Catalog excerpts

Model VREF-1

VREF Modules O Epsilon technology corp 10V calibration reference for high level devices. VREF VREF module Like the shunt calibration module for strain gaged extensometers, the VREF module is used to transfer Epsilon’s calibration to a customer’s electronics. VREF works with Epsilon’s Model 3648 high temperature capacitive extensometers, and is compatible with both the 3603 and 3604 signal conditioners. The high-precision 10V reference in the VREF module ensures calibrations are spot on every time. One VREF module may be used to calibrate several high level output extensometers by using the 10V extension value for each extensometer. VREF Module Calibration only requires these simple steps: 1. With the VREF module connected to the testing machine’s strain channel input, set VREF to 0V and zero the testing machine’s strain channel. This is the zero reference. 2. Set VREF to 10V and calibrate the full scale extensometer value on the testing machine’s strain channel. This is the calibrated full scale. 3. If your testing machine requires returning to 0V after setting full scale, set VREF back to 0V and zero the testing machine’s strain channel. 4. Reconnect the extensometer’s signal conditioner to the testing machine and you’re ready to go. Visit our website at www.epsilontech.com