Model 3590VHR

Model 3590VHR

Model 3590VHR

Product catalog summary
Overview
The Epsilon 3590VHR is a high-resolution calibrator designed for extensometers, featuring a resolution of 20 nanometers and a measuring range of 50 mm. It is equipped with software for displaying readings and can be paired with an optional netbook PC or digital display. The calibrator meets ASTM E83 Class B-1 and ISO 9513 Class 0.5 standards.

Specifications
  • Measuring Range: 50 mm (2 inches) with a maximum range of 125 mm.
  • Accuracy: ±0.19 µm over any 0.12 mm range.
  • Resolution: 20 nm.
  • Temperature Range: 18 °C to 28 °C for operation, -10 °C to 60 °C for storage.
  • Connectivity: USB 2.0 interface.
  • Weight: 3.6 kg.
  • Adapters: Ø9.52 mm adapter set included.
  • Compliance: EMC Compliance BS EN 61326-1: 2006, RoHS Directive 2011/65/EU.

Features
  • Scale calibration traceable to NIST and NPL.
  • Software for Windows-based PC included.
  • Autozero function for easy calibration.
  • Wide range of adapters available for various extensometer models.
  • Optional digital readout display and extension posts for long gauge lengths.

Options
  • Various adapter sets for different specimen types and extensometer models.
  • Extension posts for long gauge length extensometers.
  • Custom adapters available upon request.

Calibration and Compliance
Epsilon offers full system calibrations traceable to NPL, ensuring compliance with ISO 9513 and ASTM E83 standards. The in-house laser interferometer system allows for precise calibrations, eliminating reliance on micrometer head or gauge block calibrations.

Conclusion
The Epsilon 3590VHR calibrator is a versatile and precise tool for calibrating extensometers, offering high resolution and compliance with international standards. It is supported by a range of adapters and optional accessories to meet diverse testing requirements.
See more

Catalog excerpts

Model 3590VHR-1

3590VHR Very High Resolution Digital Electronic Calibrators Epsilon’s very high resolution calibrator features 20 nanometers (0.8 microinches) resolution and has 50 mm (2 inches) of measuring range. The 3590VHR is provided with software to display the calibrator readings and optional netbook PC or digital display. The 3590VHR calibrator meets the accuracy and resolution requirements for ASTM E83 Class B-1 for a 6 mm gauge length and greater extensometers and ISO 9513 Class 0,5 over the full measuring range of the calibrator. Model 3590VHR calibrator with digital readout display option Coarse and fine adjustment knobs for fast and accurate positioning Model 3590VHR calibrator storage case The 3590VHR uses a coarse adjustment screw for large displacements, coupled with a fine adjustment screw that allows fine adjustment to within approximately 0.025 microns (about 1 micro inch). When it comes time to calibrate your Model 3590VHR calibrator with software (computer optional) extensometers, these calibrators are easier to use and faster than any others on the market. They have enough measuring range to calibrate nearly all extensometers, and enough resolution to calibrate units with low measuring ranges. The software is included and provides readouts in millimeters or inches. The autozero function is a great help in calibrating. Just activate it to start readings from zero. The digital display reduces operator error. A wide range of adapters are available. The calibrator comes with smooth round adapters of 9.52 mm diameter (0.375 inches). For special adapters, contact Epsilon. For very long gauge length extensometers, an optional extension post is available. The calibrator is provided with a calibration traceable to NPL (UK standards organization). See the Model 3590VHR extensometer videos Model 3590 Adapters Enables use with a wide variety of Supplied with ISO 17025 accredited calibration extensometer models. Visit our website at www.epsilontech.co

 Open the catalog to page 1
Model 3590VHR-2

Features • Scale calibration traceable to NIST. • System fully calibrated traceable to NPL (UK standards organization). • Software for Windows based PC provided, a PC with a USB port is required. • Readings taken on the PC with included software may be save to a data file. • Uses the same adapters as the model 3590 and 3590HR calibrators. • Provided with a foam lined case for storage. • Optional Netbook PC or stand-alone digital readout display. Measuring Range: 50 mm (2 in) Maximum displacement Maximum Range: 125 mm (0-5 in) (Maximum gauge length + displacement)* Accuracy: ±0.19 pm (±7.5 pin...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.