Series 4020
2Pages

{{requestButtons}}

Catalog excerpts

Series 4020 - 1

Environics ® Product Data Series 4020 Computerized Gas Mixing and Dilution System A low cost, computerized gas blending and gas dilution system with an advanced Windows user interface. The Environics® Series 4020 is a computerized gas mixing/dilution system which offers the unique capability of mixing or diluting gases in a single chassis. The gas mixes can be used in generating precise gas calibration standards, creating gaseous atmospheres or producing gas mixes for analytical research or production purposes. The Series 4020 can produce gas concentrations from percent to ppb levels for single or multi-point calibration. The Series 4020 can mix up to three (3) component gases simultaneously to generate precise gas calibration standards and gaseous atmospheres in a broad dynamic range. The dilution capabilities of the 4020 allows from 1:1 to 1,000:1 depending on the configuration of the system. The system consists of two components: the Series 4020 instrument and the user's personal computer. The user interface is a Microsoft Windows application that communicates with the Environics system via an RS232 serial interface. The Series 4020 consists of a single chassis supporting up to four mass flow controllers. The mass flow controllers are factory calibrated using a primary flow standard traceable to the United States National Institute of Standards and Technology (NIST). Each flow controller utilizes an 11 point calibration table with linear interpolation, to increase accuracy and reduce flow controller nonlinearity. PRODUCT FEATURES AND BENEFITS Broad range of dilution ratios (up to 1,000:1) allows the user to significantly reduce the number of cylinders needed to perform compliance tests. INSTRUMENT OPERATION The Series 4020 has four basic modes of operation. Concentration Mode: Allows user to create a blend by entering target gas concentrations for each source gas cylinder, and the desired total output flow for the mix. Divider Mode: Allows the user to operate the instrument as an automated ten step gas divider. Allows multi-point calibration of analyzers. Automatic calculation of dilution and span gas flows. Internally-stored Mass Flow Controller calibration tables . Modular design allows for additional gas circuits to be added. Flow Mode: Allows user to specify the flow rate of each gas cylinder. Program Mode: Provides the ability to program the instrument for unattended operation. Programs can be recalled and run in any sequence, at various times/dates. User definable cylinder library allows for easy selection of frequently used gas cylinders. Each cylinder may contain an unlimited number of component gases with automatic K-factor calculation.

Open the catalog to page 1
Series 4020 - 2

SPECIFICATIONS Performance Accuracy Concentration: ± 1.0% setpoint Flow: ± 1.0% setpoint Repeatability: ± 0.05% setpoint *Performance specifications are valid when all Mass Flow Controllers are operating between 10% and 100% of full scale flow. Mass flow controllers are calibrated using a NIST traceable Primary Flow Standard, using a Reference Temperature of 0° C (32°F) and a Reference Pressure of 760 mm Hg (29.92 in. Hg) Warm up time: 30 minutes Mechanical Inlets External 1/4" Swagelok (or compatible fitting) Outlet External 1/4" Swagelok (or compatible fitting) Operating Pressures at...

Open the catalog to page 2

All Environics catalogs and technical brochures

  1. Series 2040

    2 Pages

  2. Series 4040

    2 Pages

  3. Series 2020

    2 Pages

  4. Series 2014

    2 Pages

  5. Series 2010

    2 Pages

  6. Series 3000

    2 Pages

  7. Series 4000

    2 Pages

  8. Series 2000

    2 Pages

  9. Series 9100

    3 Pages

  10. Series 6103

    2 Pages

  11. Series 6100

    2 Pages

Archived catalogs

  1. ROBD2

    2 Pages