Series 2040

Series 2040

Series 2040

Product catalog summary
Description
The Environics Series 2040 is a computerized system designed for blending and diluting gases within a single chassis. It is capable of producing precise gas calibration standards and gaseous atmospheres for analytical and production purposes. The system supports up to six mass flow controllers and features a serpentine pre-mix zone and a zero dead-space final mixing zone, all constructed with electropolished stainless steel and gas-compatible elastomers. The system is controlled via a user-friendly interface with a backlit LCD display and can be operated remotely through optional RS-232 communications.
Product Features and Benefits
  • Wide range of dilution ratios (1:1 to 10,000:1) reduces the need for multiple gas cylinders.
  • Multi-component capability allows for the creation of complex standards with minimal inventory.
  • Automatic calculation of dilution and span gas flows enhances efficiency and accuracy.
  • High capacity memory stores up to 200 multi-component recipes.
  • Modular design allows for future expansion and protects against obsolescence.
  • Optional RS-232 interface enables remote operation and integration with data stations.
Software
  • Concentration Mode: Displays target and actual gas concentrations.
  • Flow Mode: Displays target and actual flow rates for each gas component.
  • Maintain Ports: Manages component gas information and connections.
  • Divider Mode: Functions as a computerized ten-step gas divider.
  • Automatic Sequencer: Allows for programmable, unattended operation.
  • Purge Mode (Optional): Purges gas circuits and mixing zones.
  • Status (Optional): Enables remote activation of system modes and external devices.
Specifications
  • Performance: Accuracy and flow repeatability are within ±1.0% and ±0.05% of setpoint, respectively.
  • Mechanical: Features external Swagelok™ inlets and outlets, with operating pressures ranging from 10 to 75 psig.
  • Materials: Wetted surfaces are made of electropolished 316 stainless steel, with various seal options available.
  • Operating Temperatures: 32°F to 122°F (0°C to 50°C).
  • Dimensions: Available in portable and rack configurations.
  • Electrical: Standard 115 VAC with optional 220 VAC, 3 Amps maximum current.
  • Electronics: Inmos T 400 series processor with 12 Bit A/D and D/A conversion.
Options
  • RS-232 Serial Data Interface
  • Status Board
  • Solenoid Valve on Output
  • Extra Component Gas Port
  • Pressurization Package
  • Humidification Package
Disclaimer
Environics Inc. provides this information based on data considered accurate but does not warrant results. Specifications are subject to change without notice. Users should verify all information and consider it in conjunction with Environics' conditions of sale.
See more

Catalog excerpts

Series 2040-1

Environics Product Data ® Series 2040 Computerized Multi-Component Gas Blending/Dilution System Internally-stored mass flow controller calibration data improves accuracy. High capacity memory permits storage and recall of up to 200 multi-component "recipes", saving time and reducing errors. Twenty-five line by eighty character display permits viewing of data in worksheet form. Modular design allows user to add additional gas circuits later, reducing initial investment and protecting against obsolescence. The Series 2040 consists of a single chassis supporting up to six (6) mass flow controllers , a serpentine pre-mix zone and a zero dead-space final mixing zone. All gas wetted surfaces are electropolished stainless steel. Seals are gas compatible elastomers. Optional RS-232 Serial Data Interface permits remote operation and complete integration with a data station. The Series 2040 employs mass flow controllers calibrated to a NIST (National Institute of Standards and Technology) traceable primary standard. The Series 2040 has seven primary routines accessible through "soft" keys, located immediately below their on-screen labels. Commands are entered from the front panel and displayed on a backlit twenty-five line by eighty character liquid crystal display. Calibration sequences are stored in the internal microprocessor for recall by the keypad, optional RS-232 communications or optional status interface. Concentration Mode: User enters target output gas concentration for the span gas. The actual concentration is displayed during mixing. Flow Mode: User enters target flow rate (cc's per minute) for each component gas. Actual flow rates are displayed after mixing is initiated. Maintain Ports: User enters the name of the component gas in the source cylinder, its concentration and the port to which it is connected. Divider Mode: Allows the user to operate the instrument as a computerized ten step gas divider. Automatic Sequencer: Permits unattended automatic operation of the instrument on a programmable seven-day schedule. Purge Mode (Optional): Purge component gas circuits and mixing zone. Status (Optional): Allows user to remotely activate different modes of the system and also activate external devices DESCRIPTION ® The Environics Series 2040 offers the unique capability to blend and dilute gases in a single chassis system. The gas mixes can be used in generating precise gas calibration standards, creating gaseous atmospheres or producing gas mixes for analytical research or production purposes. The Series 2040 can produce gas concentrations from percent to ppb standards to your analytical instrumentation at highly accurate levels with our user friendly computer controlled system. The Series 2040 is available in either a bench top configuration or an optional 19” rack mount. SOFTWARE PRODUCT FEATURES AND BENEFITS Broad range of dilution ratios (1:1 - 10,000:1 depending on configuration) allows the user to significantly reduce the number of gas cylinders needed to perform compliance tests. Multi-component capability permits the user to generate a wide range of complex standards with a minimum inventory of expensive, long lead-time, multi-component gas cylinders. Automatic calculation of dilution and span gas flows based on the concentration commanded by the user eliminates the need for manual computation and allows rapid transitition from point to point and scale to scale.

 Open the catalog to page 1
Series 2040-2

SPECIFICATIONS Performance (as a percent of setpoint) * From 10 to 100% of Full Scale Flow ± 1.0% ± 1.0% ± 0.05% Accuracy Concentration: Flow: Flow Repeatability (as % of setpoint) * Mass flow controllers are calibrated using a NIST traceable Primary Flow Standard, using a Reference Temperature of 0 o C (32oF) and a Reference Pressure of 760mm Hg (29.92 in. Hg) Warm up time: 30 minutes Mechanical Inlets Balance: One external ¼" Swagelok™* Purge: One external ¼" Swagelok™* Analyte: One external ¼" Swagelok™* Outlet One external ¼" Swagelok™* Operating Pressures at inlets (flow dependent) Minimum:...

 Open the catalog to page 2

All Environics catalogs and technical brochures

  1. Series 4020

    2  Pages

  2. Series 4040

    2  Pages

  3. Series 2020

    2  Pages

  4. Series 2014

    2  Pages

  5. Series 2010

    2  Pages

  6. Series 3000