QAL 181 QAL 1 APPROVED PARTICULATE MEASUREMENT SYSTEM ■ Forward-scatter technology with a minimum detection limit of <0.05 mg/m3 ■ ProScatter™ technology provides enhanced measurement due to reduced cross-sensitivity to particle type and size compared to other scatter monitors ■ Robust and rugged for challenging high-temperature 500°C stack conditions and Ex hazardous zones - Automatic zero and span checks that fully challenge the system's ability to measure forward-scattered light and satisfy daily drift checks . TCP/IP Ethernet, RS485, RS232, 4x4 - 20 mA outputs, 1x 4-20 mA input, 4x Relay outputs, 1x Relay input for increased choice of integration with your DCS or DAHS (controller specific).
Open the catalog to page 1Elevating standards. For life. QAL 181 PARTICULATE MEASUREMENT SYSTEM TECHNICAL SPECIFICATIONS IECEx Gas Zone Certificate number Certificate code Dust Zone Certificate number Certificate code Enclosure protection rating Local output 1.5Kg standard sensor RS-485 (connection to control unit) Connection Required on Duct Hole pattern to suit DN80 PN10/PN16 or 3" 150 lb ANSI (hole ID at least 88 mm) Power Requirements Cable Entries Air Purge Requirements Dimensions 24 V DC provided by the control unit 2 x M20 cable entry glands - Dia 6-12mm cable Requires continuous air purge of 50 L/min supplied...