1. Catalogs
  2. ENVEA
  3. E-Series - AF22E SO2 monitor

E-Series - AF22E SO2 monitor

E-Series - AF22E SO2 monitor

E-Series - AF22E SO2 monitor

Product catalog summary
Overview: The AF22e is an eco-designed ultraviolet fluorescence analyzer for measuring sulfur dioxide (SO2) concentrations in ambient air, compliant with EN 14212 standards. It offers excellent metrological performance, innovative photomultiplier module design, and real-time calibration capabilities.
Features and Benefits:
  • High sensitivity and signal stability due to innovative photomultiplier module.
  • Real-time calibration graph and animated synoptic for operational monitoring.
  • Service assistance with predictive maintenance and step-by-step service guides.
  • Plug & play principle with automatic recognition of electronic boards and devices.
  • Embedded communication protocol for DAS with automatic recognition and configuration.
  • Local and remote control capabilities for configuration, calibration, and diagnostics.
  • Eco-friendly design with ultra-low power consumption and modular concept for robustness and ease of service.
Technical Specifications:
  • Measurement range: 0-1 ppm / 0-10 ppm (user selectable or auto-ranging).
  • Detection limit: <0.3 ppb, Zero noise: <0.2 ppb, Span noise: 0.4% of reading (0.3 ppb minimum).
  • Zero drift: <0.5 ppb / 24h, <0.5 ppb / 7 days; Span drift: <0.5% of F.S over 24h, <0.5% of F.S over 7 days.
  • Response time: < 60s, Repeatability: 0.4% of reading (0.3 ppb minimum), Linearity: 1% of F.S.
  • Sample flow-rate: 20 L/h, Power supply: 100-250 VAC, 50/60Hz, 24VDC (option).
  • Power consumption: 70W at start-up, 30W nominal.
  • Connectivity: Ethernet, USB, serial RS232/RS485 (optional).
  • Optional analog output with external opto-isolated I/O interface.
  • Communication protocols: Modbus, Mod4, UDP, Baeyrn.
  • Dimensions: 483 x 545 x 133 mm, Weight: 9.8 kg.
  • Operating temperature: 5°C to +40°C.
Principle of Operation: The analyzer uses ultraviolet fluorescence to measure SO2 concentrations. SO2 absorbs UV energy and emits fluorescence, which is detected by a photomultiplier tube. This method ensures accurate measurements by eliminating hydrocarbon interferences.
Main Options:
  • Built-in permeation bench with SO2 tube.
  • Internal converter for H2S monitoring and external converter module for TRS compounds.
  • Enhanced temperature range up to +50°C for use without air conditioning.
  • Pressure and temperature compensation, internal solenoid valve block, and internal sampling pump.
  • Built-in web-server for remote emulation of the analyzer.
See more

Catalog excerpts

E-Series - AF22E SO2 monitor-1

UV FLUORESCENCE SULFUR DIOXIDE ANALYZER envea A fully eco-designed, ultraviolet fluorescence based analyzer for measuring SO2 concentrations in ambient air (EN 14212) ■ Excellent metrological performances for SO2 measurements ■ Innovative conception of the photomultiplier module for excellent sensitivity and signal stability ■ Real-time calibration graph, animated synoptic, auto-diagnostic, control and maintenance data screen displayed while instrument is operating ■ Service assistance inside: detects early warning signs, allows predictive maintenance, identifies the needed service and step by step guides for service operations ■ Automatic recognition of plugged electronic boards or optional devices: plug & play principle ■ Local and remote control through digital port (configuration, calibration, test and diagnosis parameters for maintenance support) e-Series analyzers are fully eco-designed, with consideration to minimise carbon footprint through the whole lifecycle: ■ Ultra low power consumption: an environmentally-friendly and cost-saving analyzer ■ Breakthrough mechanical design for weight and power saving as well as thermal insulation & reliability ■ Modular concept makes the product more robust, and simpler to service Includes embedded communication protocol for all DAS with automatic recognition & configuration

 Open the catalog to page 1
E-Series - AF22E SO2 monitor-2

Elevating standards. For life. AF22e UV FLUORESCENCE SULFUR DIOXIDE ANALYZER Measurement range Measurement units Zero drift Span drift Response time Repeatability Linearity Sample flow-rate (user selectable or auto-ranging) Detection limit (LDL) <0.3 ppb Zero noise <0.2 ppb Span noise 0.4% of reading (0.3 ppb minimum) <0.5% of F.S over 24h <0.5% of F.S over 7 days Power supply 230VAC power supply: 70W at start-up, 30W Power nominal consumption 24 VDC power supply: 70W at start-up, 30W The AF22e is a criteria pollution monitor based on the ultraviolet fluorescence, which is the standard method...