Catalog excerpts
Product Data Sheet X2GP - General Purpose Gas Analyzer Up to four component gas analyzer featuring NDIR/UV/VIS photometer, paramagnetic and electrochemical 02, thermal conductivity and moisture sensors Enhanced performance with IntrinzX technology Extended ambient temperature range: Plantweb connectivity and WinControl data X-STREAM X2GP - General Purpose Gas Analyzer TheX-STREAM full 19" general purpose analyzercombines powerful analytical technology with modern communication functionality to face your analytical problem. The X-STREAM platform enables the combination of up to four channels of non-dispersive infrared, ultraviolet, visible photome- ters (NDIR/UV/VIS), thermal conductivity (TCD), trace moisture (tH20), paramagnetic and electrochemical oxygen (p02/e02) With the X-STREAM photometer technology, the analyzer provides a measuring accuracy that allows improving your process while also reducing the total cost of ownership by: ■ Large dynamic ranges ■ Very low temperature dependency ■ Outstanding long-term stability X-STREAM analyzers provide four status signal relay outputs and RTU over serial (RS232/485) communication. The X-STREAM X2 analyzer provides: ■ One to four analog outputs ■ Optional digital inputs and relay outputs ■ Serial interface with Modbus communication ■ Easy integration into DeltaV systems A pre-engineered DeltaV module features easy integration of X-STREAM X2 into your DeltaV environment via ModbusRTU over serial interface. ProfibusDP is also supported when using ■ Data acquisition with XTR WinControl This optional PC software supports online and offline data processing as well as data export to external devices. Calculator channels enable extended processing capabilities for measurement data. The configurable user interface enables vizualization of sampling points by including a plant image as background for measured values. ROSGMOUNT' EMERSON. Analytical Process Management
Open the catalog to page 1Ease of use The instrument has a LCD alphanumeric display and is operated manually by six keys. Clear text messages (available in 5 languages) and front panel LEDs provide information about the measurement and analyzer status. Options in a 19" Enclosure Sample gas pump Flow measurement with alarm Valve block Pressure sensor Digital Input/Output cards XTR WinControl: Data acquisition with configurable layout. Worldwide Approvals CE, CSA-C/US and C-Tick approvals allow global installation of X-STREAM general purpose analyzers. Process-approved Sensors Solvent-resistant, corrosion-resistant,...
Open the catalog to page 2Specifications Lowest and Highest Ranges Available for Different Gases (Excerpt) In total, the X-STREAM family of process gas analyzers can detect more than 60 gases. The following table is an example of the most commonly used gases. Contact your Emerson representative for information on configurations or gases that are not listed. Table 1 Gas Components and Measuring Ranges, examples Special Specs or Conditions Gas component Acetone 1 Acetone 1 Acetylene Ammonia Argon Carbon dioxide Carbon monoxide Chlorine Ethane Ethanol 1 Ethylene Helium Hexane 1 Hydrogen 4 Hydrogen Sulfide Hydrogen...
Open the catalog to page 3Standard Performance Specifications Table 2 NDIR/UV/VIS, TCD – Standard Measurement Performance Specifications NDIR/UV/VIS ≤1% ≤1% ≤ 2 % per week ≤ 0.5 % per week ≤1% 4 s ≤ t90 ≤ 7 s 5 0.2–1.5 l/min. ≤ 0.5 % ≤ 1500 hPa abs. (≤ 7 psig) Detection limit (4 σ) 1 4 Linearity 1 4 Zero-point drift 1 4 Span (sensitivity) drift 1 4 Repeatability 1 4 Response time (t90) 3 Permissible gas flow Influence of gas flow 1 4 Maximum gas pressure 8 Influence of pressure 2 – At constant temperature – With pressure compensation 7 Permissible ambient temperature Influence of temperature 1 13 (at constant...
Open the catalog to page 4Table 4 Oxygen – Standard Measurement Performance Specifications Oxygen Sensors Detection limit (4 σ) 1 4 Linearity 1 4 Zero-point drift 1 4 Span (sensitivity) drift 1 4 Repeatability 1 4 Response time (t90) 3 Permissible gas flow Influence of gas flow 1 4 Maximum gas pressure 7 Influence of pressure 2 – At constant temperature – With pressure compensation 6 Permissible ambient temperature Influence of temperature 1 12 (at constant pressure) – On zero point – On span (sensitivity) Thermostat control Warm-up time Related to full scale Related to measuring value From gas analyzer inlet at gas...
Open the catalog to page 5Special Performance Specifications for Gas Purity Measurements (ULCO & ULCO2) Table 5 Special Performance Specifications for Gas Purity Measurements Detection limit (4 σ) 1 2 Linearity 1 2 Zero-point drift 1 2 3 Span (sensitivity) drift 1 2 4 Repeatability 1 2 Response time (t90) 7 Permissible gas flow Influence of gas flow 1 2 Maximum gas pressure 10 Influence of pressure 5 – At constant temperature – With pressure compensation 8 Permissible ambient temperature Influence of temperature 6 (at constant pressure) – On zero point – On span (sensitivity) Thermostat control 1 2 3 4 Related to...
Open the catalog to page 6Signal In- & Outputs, Interfaces Analog signal outputs: 1–4, individually optically isolated 4(0)–20 mA (RB ≤ 500 Ω) Relay outputs: 4 status relays acc. NAMUR NE 107 or e.g. concentration thresholds, valve status notification dry contacts: 1 A, 30 V Communication interface: RS 485 / 232C with Modbus RTU optional: Ethernet with Modbus TCP 7/14 digital inputs (for remote control); max. 30 VDC, 2.3 mA, common ground 9/18 additional relay outputs (e.g. concentration thresholds, valve status notification, flow alarm, range ID) dry contacts: 1 A, 30 V All dimensions in mm [inches in brackets]...
Open the catalog to page 7X-STREAM X2CP Product Data Sheet jjjg www.youtube.com/user/RosemountAnalytica Emerson Process Management Rosemount Analytical Process Gas Analyzer Center of Excellence Emerson Process Management Asia Pacific Private Limited Emerson Process Management Rosemount Analytical Gas Chromatograph Center of Excellence 10241 West Little York, Suite 200 MIDDLE EAST AND AFRICA Emerson Process Management Jebel AN Free Zone Dubai, United Arab Emirates ©2013 Emerson Process Management. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount Analytical is a...
Open the catalog to page 8All Emerson Automation Solutions - ROSEMOUNT catalogs and technical brochures
-
VPS 53 Vapor Pressure Sensor
2 Pages
-
EasyHeat™ XD Cable Kits
2 Pages
-
FRL Accessories
14 Pages
-
TankRadar Pro
32 Pages
-
Rosemount™ 2140 Level Detector
22 Pages
-
Roxar subsea PIG
2 Pages
-
Roxar subsea Sand monitor
2 Pages
-
Roxar subsea CM10K
2 Pages
-
Roxar PTPT15K
2 Pages
-
Rosemount TankRadar OFC
2 Pages
-
SW-300
4 Pages
-
DL8000
12 Pages
-
DeltaV Analyze
8 Pages
-
AMS Machine Works
4 Pages
-
DeltaV Executive Portal
7 Pages
-
1500XA Gas Chromatograph
8 Pages
-
Rosemount 975UR
4 Pages
-
Rosemount 975HR
4 Pages
-
Rosemount 975UF
4 Pages
-
Rosemount 975MR
4 Pages
-
CT5400 Process Gas Analyzer
4 Pages
-
CT2211 Leak Detection System
4 Pages
-
Desalination of Seawater
2 Pages
-
700XA Process Gas Chromatograph
12 Pages
-
SafeGuard Alarm Controller
2 Pages
-
Safety Monitoring
6 Pages
-
AMS Suite
16 Pages
-
Drinking Water Sht 2V2
2 Pages
-
Line Card
2 Pages
-
Drinking Water Industry
16 Pages
-
Explosion Proof Video Camera
2 Pages
-
Net Safety UV Flame Detector
4 Pages
Archived catalogs
-
UV/IRS Flame Detector
4 Pages
-
UVS Flame Detector
4 Pages