LP980 Spectrometer Brochure

Product catalog summary
Overview: The LP980 Transient Absorption Spectrometer is a modular, computer-controlled system designed for measuring transient absorption and laser-induced emission kinetics. It offers two modes: Kinetic Mode (LP980-K) for time-resolved measurements at a single wavelength, and Kinetic and Spectral Mode (LP980-KS) for time-gated spectral measurements.
Technical Specifications: The spectrometer uses a nanosecond pulsed laser source to generate transient species, with a xenon lamp as the probe source. It supports measurements in liquid, gaseous, and solid samples, with different setups for each to optimize signal-to-noise ratios.
Measurement Modes:
  • Kinetic Mode: Utilizes a PMT detector and oscilloscope for high-resolution transient kinetics, covering 230 nm to 2550 nm.
  • Spectral Mode: Uses an ICCD detector for full transient spectral features, with time resolutions down to 3 ns and spectral coverage from 230 nm to 930 nm.
Signal Detection and Data Acquisition: The system includes a red-sensitive PMT detector with a high dynamic range and linearity. Detection limits are ΔOD = 0.002 for fast detectors and ΔOD = 0.0005 for slow detectors, with improvements through data averaging.
Software: The L900 software package controls the spectrometer, offering various measurement and correction methods, data averaging, and analysis routines. It supports multiple measurement modes and provides comprehensive data manipulation and display options.
Measurement Examples: Examples include transient absorption and photobleaching, triplet-triplet annihilation, and singlet oxygen luminescence decays. These demonstrate the system's capability to handle complex transient dynamics and spectral dependencies.
Applications: The LP980 is suitable for high-quality research and routine applications, providing flexibility and ease of use for studying transient chemical and biological species.
Key Features:
  • Data Conversion and Correction: The system can convert kinetic decay data into spectral data and vice versa. It includes automated fluorescence background subtraction and probe subtraction to ensure accurate TA behavior representation.
  • Measurement Conditions: Examples include measurements of Ruthenium bipyridine in water and Anthracene in cyclohexane, with specific conditions for pump and probe wavelengths, energy, and slicing intervals.
  • Software Capabilities: The LP980-K software allows for standard curve fitting and advanced analysis using the optional FLASH software for batch and global fitting of multiple decay curves.
Technical Specifications:
  • Sensitivity: Minimum detectable optical density (ΔOD) is 0.002 for fast detectors and 0.0005 for slow detectors.
  • Time Resolution: Instrument response function (FWHM) varies from 5 ns to 100 ns depending on the detector option.
  • Laser Excitation Source: Includes options for Nd:YAG lasers and tuneable dye lasers, with specific wavelength ranges and pulse energies.
  • Probe Source: Utilizes a pulsed or steady-state xenon arc lamp with specific operational parameters.
  • Monochromator: Features a 325 mm focal length Czerny-Turner design with a triple grating turret and integrated filter wheel for second-order light removal.
  • Detector Options: Includes PMT and InGaAs detectors covering a wide spectral range, with additional options for NIR-PMTs and ICCD cameras for spectral mode.
Accessories:
  • Stopped Flow Accessory: Allows for rapid kinetic analysis with multi-mixing capabilities.
  • Temperature-Controlled Sample Holders: Options for TE-cooled and cryostat holders for precise temperature control during measurements.
  • Laser Energy Meters: Measures excitation pulse energy and energy transmitted by the sample.
Conclusion: The LP980-K spectrometer is a versatile and flexible system designed for precise TA and emission measurements, with extensive software and hardware options to accommodate various research needs.
Contact Information: Edinburgh Instruments Ltd. is based in the UK, with customer support available worldwide. Contact details for both UK and US support are provided.
See more

Catalog excerpts

LP980 Spectrometer Brochure-1

EDINBURGH INSTRUMENTS TRANSIENT ABSORPTION SPECTROMETER

 Open the catalog to page 1
LP980 Spectrometer Brochure-2

2 LP980-K TRANSIENT ABSORPTION SPECTROMETER - KINETIC MODE Spectrometer system for measurements of transient absorption and laser-induced emission kinetics with the ability to automatically generate transient spectra. 10 LP980-SYSTEM OVERVIEW TRANSIENT ABSORPTION SPECTROMETER Configuration, layout, and specifications for the turn-key, modular and unrivalled spectrometer. 14 LP980-KS TRANSIENT ABSORPTION SPECTROMETER - KINETIC AND SPECTRAL MODE Spectrometer system for direct measurements of time-gated transient absorption and laser-induced emission spectra as well as kinetic measurements. ABOUT...

 Open the catalog to page 2
LP980 Spectrometer Brochure-3

TRANSIENT ABSORPTION / LASER FLASH PHOTOLYSIS TECHNICAL OVERVIEW Transient Absorption / Laser Flash Photolysis is a technique for studying the transient chemical and biological species generated by a short, intense light pulse from a nanosecond pulsed laser source (pump pulse). This intense light pulse creates short lived photoexcited intermediates such as excited states, radicals, and ions. These intermediates are generated in concentrations large enough for chemical and physical interaction to occur and for direct observation of the associated temporally changing absorption characteristics....

 Open the catalog to page 3
LP980 Spectrometer Brochure-4

OVERVIEW The LP980 Transient Absorption (TA) Spectrometer uses the same basic optical-electrical setup for operation in both the kinetic and the spectral mode, i.e. laser excitation source, probe source, sample compartment (including optics, attenuators, laser shutter, and probe shutter), monochromator / spectrograph, and control electronics. The difference between the two modes is the detector and the data acquisition electronics. In kinetic mode a PMT detector is used and the transients are acquired using a fast, high resolution oscilloscope. The LP980-K has been designed to meet the stringent...

 Open the catalog to page 4
LP980 Spectrometer Brochure-5

SIGNAL DETECTION AND DATA ACQUISITION The LP980-K is supplied as standard with a red sensitive PMT which covers the spectral range from 200 nm 870 nm. The detector housing also accommodates the high voltage power supply and the voltage divider circuit. The PMT detector system is designed to achieve a high dynamic range with exceptional current linearity. This is an essential requirement in measuring small signals on a large signal background. The voltage divider and power supply are configured to support this high linearity current mode by operating in a five stage dynode configuration. The PMT...

 Open the catalog to page 5
LP980 Spectrometer Brochure-6

Control Features • Wavelength / slit control • Filter wheel for second order light rejection • Grating selection • Pump laser flashlamp trigger • Pump laser Q-switch trigger • Probe source pulse current • Pump and probe shutters • Oscilloscope trigger • Oscilloscope time base • Oscilloscope voltage scale • Signal offset • Time shift / delay • Optional temperature controlled sample holder • Optional cryostat mounting • Stopflow synchronisation • Optional Energy Meters SOFTWARE The LP980-K spectrometer system is fully computer-controlled by means of the comprehensive L900 software package. A variety...

 Open the catalog to page 6
LP980 Spectrometer Brochure-7

MEASUREMENT EXAMPLES Transient Absorption and Photobleaching In TA, the presence of the transient species can cause the sample to have either increased or decreased levels of absorption relative to the absorption of the ground state species (positive ΔOD and negative ΔOD respectively). While an increased absorption is associated with triplettriplet or singlet-singlet transitions, a reduction in the measured optical density is associated with either ground state depletion or sample emission. Generally these effects can be separated spectrally, or by means of their lifetimes. In some special cases...

 Open the catalog to page 7
LP980 Spectrometer Brochure-8

Spectrally Dependent Transient Kinetics Time-resolved TA spectra can provide substantially more information than kinetic measurements alone. The measurement of benzophenone in cyclohexane shows two distinct absorption bands with maxima at 330 nm and 530 nm. The delay characteristics (measured with a PMT) show different decay kinetics in each band. Spectral measurements (measured with an ICCD camera) reveal that the longer wavelength band shifts towards the near-infrared spectral range with time. Sample: Benzophenone in cyclohexane Measurement Conditions: Top: Apump = 355 nm, Epump = 8 mJ, pulsed probe...

 Open the catalog to page 8
LP980 Spectrometer Brochure-9

TIME−RESOLVED ABSORPTION SPECTRA The LP980-K Spectrometer hardware is computercontrolled and enables the user to generate timeresolved absorption spectra in a two-fold process: firstly, a series of TA measurements over a pre-defined range of probe wavelengths is recorded, and, secondly, this data is sliced at desired time windows and delays from the laser pulse excitation. By automatically scanning through the spectral range the probe background level can change. The changing background level does not have an effect on the value of the optical density, but it has an effect on the noise of the...

 Open the catalog to page 9
LP980 Spectrometer Brochure-10

SYSTEM The LP980-K is a system for the measurement of laser-induced TA and emission decay kinetics with the ability to automatically generate temporally resolved TA and emission spectra. Optional: Thin-film geometry, co-linear excitation, diffuse reflectance geometry, and fluorescence and phosphorescence lifetime measurements accessories are all available. Edinburgh Instruments Ltd. has a policy of continuous product development and reserve the right to amend specifications without prior notice (Aug 2018)

 Open the catalog to page 10
LP980 Spectrometer Brochure-11

LASER PUMP SOURCE In TA / Laser Flash Photolysis, transient species are generated using a short pulse, high peak power laser, known as the pump pulse. Suitable lasers include fixed wavelength lasers, particularly Nd:YAG lasers (fundamental wavelength 1064 nm) and their harmonics (at wavelengths 532 nm, 355 nm and 266 nm), or tuneable lasers, particularly optical parametric oscillators (OPOs). Other lasers sometimes used include nitrogen, excimer lasers and dye lasers. The LP980 has been designed with the ultimate flexibility in mind. It can be supplied either as a turn-key, fully tested and performance...

 Open the catalog to page 11

All Edinburgh Instruments catalogs and technical brochures

  1. MicroPL

    4  Pages

  2. VPLED

    2  Pages

  3. VPL

    2  Pages

  4. HPL

    2  Pages

  5. AGILE

    2  Pages

  6. RM5

    7  Pages

  7. EPLED Series

    2  Pages

  8. FLS1000

    24  Pages

  9. LifeSpec II

    4  Pages

  10. DS5

    4  Pages

  11. RMS1000

    16  Pages

  12. 295-FIR

    2  Pages

  13. 395-FIR

    2  Pages

  14. PL Series

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.