Overview: The DEEneo is a digital signal conditioner designed for inductive sensors (LVDT) and is suitable for DIN rail mounting. It supports a wide range of LVDT sensors and offers digital signal processing via a microcontroller, with features such as signal linearization, cable break detection, and alarm signaling.
Technical Specifications:
  • Output Signals: 0...20 mA, 4...20 mA, 0...5 V, ±5 V, 0...10 V, ±10 V.
  • Power Supply: 9...36 VDC, with power consumption of 70 mA at 24 VDC and 130 mA at 12 VDC.
  • Resolution: 16-bit digital signal processing.
  • Operating Temperature: -40 to +85 °C.
  • EMC Compliance: EN IEC 61326-1:2021.
  • Dimensions: 77 x 75 x 26 mm.
Functions:
  • Cable Break Detection: Alerts via LED and deactivates analog signal output upon detection.
  • Switching Output: Programmable with configurable direction and threshold using the SET button or eddySetup software.
  • Alarm Output: Can be used as a second switching output.
Software Configuration: The eddySetup software allows for detailed configuration of the DEEneo, including sensor supply settings, output signal selection, filter frequency, and alarm output activation. It is compatible with Windows 10 and 11.
Connection Details: The DEEneo supports various LVDT sensor configurations, including 4-wire, 5-core, and 6-wire connection cables. Specific terminal assignments are provided for each configuration.
Ordering Information: The DEEneo can be ordered with different output signal configurations and is available with either sensor cable or connector output options.
See more

Catalog excerpts

DEEneo digital signal conditioner-1

Digital signal conditioner for inductive sensors (LVDT) for DIN rail mounting ■ Supporting a wide range of inductive sensors (LVDT) ■ Digital signal processing by microcontroller ■ Digitalization via 16 bit AD converter ■ Linearisation of sensor signal ■ Signal adjustment via SET button or software ■ Cable break detection ■ Alarm signal and switc

 Open the catalog to page 1
DEEneo digital signal conditioner-2

2 TECHNICAL DATA | DRAWING DEEneo The DEEneo signal conditioner was developed for operating inductive LVDT sensors (full bridge). The electronics supply the sensor and convert the sensor signal into a standardized, analogue output signal with the help of a microcontroller. DEEneo supports the operation of all LVDT sensors. The primary coil of the connected sensor is excited with a sinusoidal signal. To determine the sensor position, the secondary coil signals are calculated differentially and are used to generate an analog output signal. The internal, ratiometric signal processing is digital...

 Open the catalog to page 2
DEEneo digital signal conditioner-3

CONNECTION DEEneo is designed to operate all commercially available LVDT sensors. ■ LVDT sensors with 4-wire connection cable (e.g. eddylab LVDTs): Please refer to the table below or the diagram on the right. ■ LVDT sensors with 5-core connection cable: The center tap (ends of both secondary windings) does not need to be connected. Alternatively, it can also be connected to terminal 1 or 7. ■ LVDT sensors with 6-wire connection cable: The ends of both secondary windings (see sensor data sheet) must be connected. These can then be connected to terminal 1 or 7. CONNECTION DATA CLAMP

 Open the catalog to page 3
DEEneo digital signal conditioner-4

The DEEneo electronics have an integrated cable break detection based on an impedance measurement of the LVDT's primary coil. If the sensor cable is cut, the impedance at the electronics changes independently of the core position and the cable break detection is triggered. This requires the connections of the primary coil of the sensor to be severed. A partial break only of the connections to the secondary coils does not activate this function. To use the cable break detection function, an alarm-emitting device (signal lamp, acoustic warning device) is connected to terminal 13 (alarm S2) or the...

 Open the catalog to page 4
DEEneo digital signal conditioner-5

DEEneo I ORDER CODE 5 ORDER CODE DEEneo- X Possible combinations: 51 + DEEneo: Sensor with connector output, cable 52 + DEEneo: Sensor with cable output, electronics DEEneo POSSIBLE COMBINATIONS ■ Sensor with connector output, cable K4PxM, electronics DEEneo Sensor with cable output, electronics DEEneo Email: [email protected] Internet: www.eddylab.de

 Open the catalog to page 5

All Eddylab GmbH catalogs and technical brochures

  1. SM-F18

    8  Pages

  2. AX Series

    8  Pages

  3. MAGNESCALE

    11  Pages

  4. Serie REX-D

    8  Pages

  5. REX-Series

    8  Pages

  6. PMX-24

    4  Pages

  7. PAXI

    6  Pages

  8. PAX P, DP

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.