ECS01

ECS01

Product catalog summary
Overview
The ECS01 Quaternary Gradient Analytical System is a high-performance HPLC system designed for analytical applications. It includes a UV-VIS detector, column oven, analytical pump, and gradient box, all of which communicate via RS232 or Ethernet. The system is supported by Clarity Chromatography software, which meets CFR21GLP requirements.

Main Components
  • ECD2800 UV-VIS Detector: Offers a wavelength range of 190-800 nm with low noise levels (±3·10-6 AU) and high-speed sampling up to 100 Hz. It includes an AC 05 analytical cell and features wavelength autocalibration.
  • ECO2080 Column Oven: A peltier heating/cooling oven that can accommodate up to three columns of 25 cm length. It features an Overheating-Security-Fuse and operates within a temperature range of 0-80 °C.
  • ECP2010 Analytical HPLC Pump: Functions as a quaternary gradient pump with the ECB2004B Gradient Box. It has a flow rate range of 0.01-10 ml/min and a pressure limit of 40 MPa. The pump includes a learning algorithm for pulsation suppression.
  • ECB2004B Gradient Box with Degasser: Handles liquids in ECOM gradient systems, featuring a vacuum degasser and a four-way gradient valve controlled by the ECP2010 pump.

System Requirements
The system requires a PC equipped with at least three serial RS232 ports or one Ethernet port. The Clarity software requirements should also be checked.

Included Components
  • ECD2800 UV-VIS Detector (P/N: ADA0000X)
  • ECB2004B Gradient Box with Degasser (P/N: ABA0000X)
  • ECP2010 Analytical HPLC Pump (P/N: ACA0000X)
  • ECO2080 Column Oven (P/N: AOA0000X)
  • Clarity Station for 1 instrument (P/N: AP000C50)
  • Clarity module for HPLC control (P/N: AP000A24)
  • Analytical injection valve with 20ul sample loop (P/N: AVE02006)
  • Accessories for analytical system series 2000 (P/N: ASA00180)
See more

Catalog excerpts

ECS01-1

High effective HPLC system containing UV-VIS detector, column oven, analytical pump and gradient box. Detector with continuously variable wavelength 190-800 nm and very low noise level offers high-speed sampling up to 100 Hz. Communication of all units is through RS232 or Ethernet (LAN). Additional PC A/D converter card is not required for this system. System contains manual injector and Clarity software. All units are supported in Clarity software from installation. System contains Detector with continuously variable wavelength in the range of 190-800 nm and a noise level ±3-10-6 AU. Unit offers high-speed sampling up to 100 Hz and wavelength autocalibration. AC 05 analytical cell is included. ECO2080 Column Oven with heating and cooling The peltier heating/cooling column oven with a big compartment room can accommodate up to 3 columns with length up to 25 cm. Overheating-Security-Fuse prevents destruction of temperature sensitive columns. Temperature range is 0-80 °C ECP2010 Analytical HPLC Pump This isocratic pump works with Gradient Box ECB2004B as a quaternary gradient pump. Flow rate range is 0.01-10 ml/min and pressure limit at 40 MPa. Unit software includes new learning algorithm for pulsation suppression and many testing and diagnostic functions. ECB2004B Gradient Box with Degasser Box is suited for handling liquids in ECOM gradient systems. It accommodates a container for solvent bottles, built-in high-efficiency vacuum degasser and four-way gradient valve. Up to four solvent lines may be degassed simultaneously by one unit. Four-way gradient valve is controlled from the pump ECP2010. This system is fully controlled by Clarity Chromatography software, which is also a part of delivery. Clarity software complies CFR21GLP requirements. Note: Column and computer are not included. PC system has to be equipped for handling 3 serial RS232 or 1 Ethernet port at least. Check also Clarity software requirements. Clarity Chromatography Software ECOM spol. s r.o. Trebonicka 239, CZ252 19 Chrastany u Prahy, Czech Republic

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.