- Process volume capacity 700 L per cycle. STAAT NDARD Our patented process is designed to shred and sterilize infectious waste by saturated steam. Shredding and sterilization are achieved in one fully enclosed and automated, stainless steel, system with no intermediate waste handling. The contaminated waste is loaded into the upper chamber where a heavy-duty shredder reduces the waste into small pieces. Using gravity, the processed material drops into the lower treatment chamber. After shredding, waste and all the inner parts of the machine are steam heated to 138°C (280 F) and pressurized to 3,5 bar (51 psi). Sterilization is achieved by maintaining a temperature of 138°C (280 F) at the core of the waste for recycle or to dispose as ordinary waste. Waste are both sterilized (Microbial inactivation = 108) and volumereduced by up to 80 %. The 30 minutes duration of an average cycle process is fully automated and totally monitored. A computerized control system scans the process and automatically prints and records on a memory card a batch report at the end of each cycle with the essential sterilizing parameters for accurate record keeping. Approved by the Higher Council of Public Health of France under the direction of the Ministry of Health and the Ministry of the Environment.
Open the catalog to page 1General Characteristics Operating Characteristics Average Cycle Time Process Volume Capacity Average Waste Density Average Process Weight Capacity Microbial Inactivation Waste Volume Reduction