Technical Specifications
The EC4-ClO2-5-01 is an electrochemical 3-electrode sensor designed for detecting chlorine dioxide gas. It is suitable for industrial safety, leakage detection, emission monitoring, sewage/water treatment plants, ClO2 generators, the semiconductor industry, and warehouse logistics monitoring. Key features include high selectivity, double-sealed housing for leakage protection, stable zero, linear output, high sensitivity, low power consumption, and a wide operating temperature range from -20°C to +40°C. The sensor is RoHS compliant and has a typical warm-up time of less than 60 seconds.

Performance
The sensor has a sensitivity of -325 nA/ppm ± 200 nA/ppm, a zero current of ± 20nA, and a range of 0-5ppm with a maximum overload of 10ppm. It offers a resolution of less than 0.01 ppm and response times of T50 < 20s and T90 < 60s. The repeatability is 2%, and the output is linear.

Operation
Operating on an amperometric principle with a bias voltage of 0 mV, the recommended load resistor is 100 Ω. The sensor's warm-up time is less than 60 seconds.

Environmental Conditions
The sensor operates within a temperature range of -20 to +40℃, a humidity range of 15-95 %RH (non-condensing), and a pressure range of 800 to 1200 hPa. The storage temperature is optimal between 4 to 6℃.

Lifetime and Warranty
The sensor has a long-term drift of less than 1% per month, an expected lifetime of over 2 years, and a zero drift in clean air of less than 1 ppm. The storage life is 6 months, and it comes with a 12-month warranty.

Cross Sensitivity
The sensor shows positive readings for gases like bromine, chlorine, iodine, nitrogen dioxide, ozone, and fluorine, while sulfur dioxide shows a negative reading. Methane and carbon dioxide do not affect the sensor reading.

Disclaimer and Warning
The performance data is based on test conditions and may change without notice. Users must follow guidelines for storage, assembly, and operation to ensure proper use. The sensor should not be cleaned with strong solvents, and its response should be checked before use. Disposal should follow local electronic waste regulations.

Contact Information
EC Sense GmbH in Germany and Ningbo AQSystems Technology Co., Ltd. in China are the main business centers for Europe, the rest of the world, and Asia respectively.
See more

Catalog excerpts

EC4-ClO2-5-01 Chlorine Dioxide Gas Sensor-1

Electrochemical 3-Electrode Sensor EC4-ClO2-5-01 Chlorine Dioxide Gas Sensor Datasheet Easy Gas Sensor Solutions

 Open the catalog to page 1
EC4-ClO2-5-01 Chlorine Dioxide Gas Sensor-2

EC4-ClO 2 -5-01 Chlorine Dioxide Gas Sensor Datasheet EC4-ClO2-5 Chlorine Dioxide Gas Technical Specifications Performance Features Detects with high selectivity a wide variety of gases Double sealed housing for advanced leakage protection Stable zero Zero Current Maximum Overload Response Time Environment Operating Temperature Range High sensitivity Operating Humidity Range Operating Pressure Range Storage Temperature Linear output Wide operation temperature range of -20°C to +40°C Typical warm-up time in minutes RoHS compliant Typical Applications Industrial Safety Leakage Detection Operation...

 Open the catalog to page 2
EC4-ClO2-5-01 Chlorine Dioxide Gas Sensor-3

EC4-ClO 2 -5-01 Chlorine Dioxide Gas Sensor Datasheet Test Concentration Sensor Reading Carbon Dioxide Nitrogen Dioxide Sulfur Dioxide *Positive Reading Expected **Negative Reading Expected Note: 1) The above interference factors may vary due to different sensors and service life, please refer to the actual test results. 2) This table is not complete for all cross gases. Please contact us for other gases. 3) The above parameters are the test results at a temperature of 25°C, a relative humidity of 50%RH and a normal pressure environment. The performance of the sensor varies under different environmental...

 Open the catalog to page 3
EC4-ClO2-5-01 Chlorine Dioxide Gas Sensor-4

EC4-ClO 2 -5-01 Chlorine Dioxide Gas Sensor Datasheet Disclaimer The EC Sense performance data stated above is based on data obtained under test conditions using the EC Sense gas distribution system and AQS test software. In the interest of continuous product improvement, EC Sense reserves the right to change design features and specifications without notice. We are not responsible for any loss, injury or damage caused by this. EC Sense assumes no responsibility for any indirect loss, injury or damage resulting from the use of this document, the information contained therein or any omissions...

 Open the catalog to page 4
EC4-ClO2-5-01 Chlorine Dioxide Gas Sensor-5

Business Centre Europe and the rest of the world Business Centre Asia EC Sense GmbH Wangener Weg 3 82069 Hohenschäftlarn, Germany Tel: +49(0)8178-99992-10 Fax: +49(0)8178-99992-11 Email: [email protected] www.ecsense.com, www.ecnose.de Ningbo AQSystems Technology Co., Ltd. F4-17 Building, Zhong Wu Technology Park No.228, Jin Gu Bei Road, Yinzhou District NingBo, Zhejiang Provence, P.R. China Post Code: 315100 Tel: +86(0)574 88097236, 88096372 Email: [email protected] www.ecsense.cn, www.ecnose.de EC4-ClO2-5 Chlorine Dioxide Gas Sensor_Datasheet_V1.1_20230607 Copyright@2023 EC Sense GmbH

 Open the catalog to page 5
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.