1. Catalogs
  2. ebro
  3. USB DATA LOGGERS FOR COLD CHAIN MONITORING
video corpo

USB DATA LOGGERS FOR COLD CHAIN MONITORING
1 /12Pages

USB DATA LOGGERS FOR COLD CHAIN MONITORING

USB DATA LOGGERS FOR COLD CHAIN MONITORING
1 /12Pages

Catalog excerpts

USB DATA LOGGERS FOR COLD CHAIN MONITORING-1

EBI 300/310/330 USB DATA LOGGERS FOR COLD CHAIN MONITORING

 Open the catalog to page 1
USB DATA LOGGERS FOR COLD CHAIN MONITORING-2

The easy to use data loggers with USB connection monitor the temperature and/or humidity during transport and storage of sensitive goods like medicine, food, serums etc.. Measurement reports are created automatically as PDF files when you connect the logger to a PC. The EBI 300 and EBI 310 USB data loggers are suitable for multi-use, the EBI 330 data loggers are single-use versions which can be ordered preconfigurated and are used especially when returning a more expensive multi-use logger to the sender after a shipment is difficult. Please contact us for more information. Program | Measure •...

 Open the catalog to page 2
USB DATA LOGGERS FOR COLD CHAIN MONITORING-3

Automatic PDF report generation Programmable at www.ebi300.com no special software for programming and readout required but available Indication of alarm status via flashing LED Data integrity Conforms with FDA 21 CFR Part 11 and ATP The data loggers help you to comply with GMP and VO (EG) 37/2005 Free firmware updates at your place via software Excellent value-for-money Digital interface • Digital interface between loggers and external probes (at EBI 300 TE, EBI 300 TH, EBI 310 TE, EBI 310 TH and EBI 310 TX). Data logger functions as data collector with optional internal sensor Easy exchange of...

 Open the catalog to page 3
USB DATA LOGGERS FOR COLD CHAIN MONITORING-4

Every EBI 3x0 USB data logger have the afore mentioned properties. Depending on the application, claim and your purse, there are different requirements for which we have the right devices.The following overview shall help making the decision. * with connected, exchangeable sensors

 Open the catalog to page 4
USB DATA LOGGERS FOR COLD CHAIN MONITORING-5

General technical specifications: valid for both EBI 330 data logger types* Memory capacity 20,000 measurements Resolution Storage temperature Sample rate Measurement modes ♦ Start with key press ♦ Start immediately until end of memory Maximum start delay Housing material Protection class Batch calibration certificate available on request * Please find the exact technical data of each EBI 330 data logger type on this page. EBI 330-T30 Single-Use USB Data Logger Standard version Technical Data Measurement rang / operating Battery lifetime ± 0.8 °C for the remaining measurement range NTC Lithium...

 Open the catalog to page 5
USB DATA LOGGERS FOR COLD CHAIN MONITORING-6

Memory capacity 40,000 measurements Storage temperature Sample rate Measurement modes Maximum start delay Housing material Certificate Battery Battery lifetime • Start / Stop ♦ Measurement until end of memory • Start with key press Value, MIN / MAX, until end of memory, alarm on / off 72 h Factory calibration certificate Lithium button cell (CR 2450), 3 V Up to 2 years, depending on applications * Please find the exact technical data of each EBI 300 data logger type on this double page. Technical Data Measurement range/operating -30 °C ... +70 °C (-22 °F ... +158 °F) temperature By connecting...

 Open the catalog to page 6
USB DATA LOGGERS FOR COLD CHAIN MONITORING-7

Technical Data Measurement range external -35 °C ... +70 °C (-31 °F ... +158 °F) Measurement range internal -30 °C ... +70 °C (-22 °F ... +158 °F) temperature / operating temperature Accuracy (internal and external) ± 0.5 °C (- 20 °C ... +40 °C / - 4 °F ... +104 °F) ± 0.8 °C for the remaining measurement range Probe NTC, Stainless steel, 0 4mm, L = 50mm, pointed Cable PVC, L = 1m, waterproof, oilproof and food safe • Internal temperature probe usable additionally Technical Data Measurement range temperature / -30 °C ... +70 °C (-22 °F ... +158 °F) operating temperature ± 0.8 °C for the remaining...

 Open the catalog to page 7
USB DATA LOGGERS FOR COLD CHAIN MONITORING-8

EBI 310 Multi-Use USB Data Loggers General technical specifications: valid for all EBI 310 data logger types* Memory capacity E n d le s s m e a s u re m e n t Start / Stop Measurement until end of memory Start with key press Value, MIN / MAX, until end of memory, alarm on / o Factory calibration certificate * Please find the exact technical data of each EBI 310 data logger type on on the next pages. EBI 310 Multi-Use USB Data Logger High precision version Technical Data Measurement range/operating temperature … 22 °F ... +167 °F) By connecting an external probe, the temperature measurement range...

 Open the catalog to page 8
USB DATA LOGGERS FOR COLD CHAIN MONITORING-9

Technical Data Measurement range external -200°C ... +250°C (- 328 °F ... +482 °F) Measurement range internal -30 °C ... +75°C (- 22 °F ... +167 °F) temperature / operating temperature ± 0.5 °C for the remaining measurement range Probe Pt 1000, Stainless steel, 0 5mm, L = 50mm, blunt Cable PTFE, L = 1m, waterproof, oilproof and food safe Battery Lithium button cell (CR 2450), 3V Battery life time Up to 2 years, depending on applications

 Open the catalog to page 9
USB DATA LOGGERS FOR COLD CHAIN MONITORING-10

Technical Data Measurement range temperature / -30 °C ... +75°C (- 22 °F ... +167 °F) operating temperature ± 0.5 °C for the remaining measurement range ± 0.8 °C for the remaining measurement range Probe temperature Pt 1000 Accuracy humidity ± 2 %between 10% rH ... 90 %rH (at 25 °C/+77 °F) ± 4 % for the remaining measurement range Probe humidity capacitive Battery Lithium button cell (CR 2450), 3V Battery life time Up to 2 years, depending on applications Technical Data Measurement range external -200°C ... + 400°C (-328 °F ... +752 °F), temperature dependent on probe type Measurement range internal...

 Open the catalog to page 10
USB DATA LOGGERS FOR COLD CHAIN MONITORING-11

• Measurement range: -200 °C ... +200 °C (-328 °F ... +392 °F) • Needle: L = 45 mm, Ø = 5 mm, blunt • Cable: PTFE, L = 3 m • Measurement range: -50 °C ... +180 °C (-58 °F ... +356 °F) • Needle: L = 130 mm, Ø = 3 mm, blunt • Cable: PTFE, L = 3 m • Measurement range: -50 °C ... +180 °C (-58 °F ... +356 °F) • Needle: L = 130 mm, Ø = 3 mm, blunt • Cable: PTFE, L = 1 m • Measurement range: +100 °C ... +400 °C (+212 °F ... +752 °F) • Needle: L = 50 mm, Ø = 1.5 mm, blunt • Cable: metal wrapped, L = 3 m, not waterproof

 Open the catalog to page 11
USB DATA LOGGERS FOR COLD CHAIN MONITORING-12

Single-use USB data logger, package unit: 10 pieces Standard USB Data Logger USB Data logger with external temperature probe USB Data logger with external humidity probe High Precision USB Data Logger USB Data logger with external precision temperature probe USB Data logger with external humidity probe USB Data logger with temperature-two-channel-adapter Let's Solve Water Xylem Analytics Germany Sales GmbH & Co. KG, ebro ebro is a registered trademark of Xylem Inc. or one of its subsidiaries. Technical changes reserved. © 2014 Xylem Analy tics Germany Sales GmbH & Co. KG.

 Open the catalog to page 12
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.