Wireless Data Logger EBI 10 for Validation
6Pages

{{requestButtons}}

Catalog excerpts

Wireless Data Logger EBI 10 for Validation - 1

Wireless Data Loggei EBI 10 for Validation PRQCESS VALIDATIoN AccQRDING TQ Iso 17665 / Iso 15883 mdical | cssd a xylem brand

Open the catalog to page 1
Wireless Data Logger EBI 10 for Validation - 2

Only 3 Easy Steps ...ebro® Validation Systems Setup 1 Evaluate Report 3 2 Measurement systems: The ebro Winlog validation system is a combination of high accuracy dataloggers and powerful reporting and evaluation software. The system consists of: EBI sries Data Loggers: Datalogger to measure and record process conditions (temperature, pressure, time). TGV validation, FDA 21 CFR Part 11 conform, IQ / OQ documentation, allows evaluation according to international standards (EN 285, ISO 17665, ISO 15883) Winlog.validation Easy-to-use software for programming, evaluation, calibration and report...

Open the catalog to page 2
Wireless Data Logger EBI 10 for Validation - 3

Technical Data Type Cable Logger EBI 10-T and EBI 10-TP Sensors 2, 3 or 4 (probes are factory replaceable) Temperature measurement range -50 °C ... +150 °C (-58 °F ... +302 °F) Pressure measurement range 1 mbar ... 4,000 mbar (0.015 psi ... 58 psi) Temperature resolution 0.025° C (0.045 °F) Pressure resolution 1 mbar (0.015 psi) Temperature accuracy 0.1° C (0.045 °F) Pressure accuracy 10 mbar (0.15 psi) Memory 100,000 measurements (Total) Sampling rate 250 ms ... 24 h Measurement mode endless, start / stop time, start immediately, start immediately until end of memory Battery 3.6 V...

Open the catalog to page 3
Wireless Data Logger EBI 10 for Validation - 4

Accessories EBI 10 Cable Logger Probe AL 171 PT1000 PTFE coated Probe length = 3 cm Probe diameter = 1.2 mm Cable length = 1,200 mm ocess validation in washer-disinfectors nd steam steam-sterilizers Temprature and pressure loggers complying with the requirements of the standard EN ISO 15883-1 must be used during the process validation in washer-disinfectors. The sensors are placed between the instruments and in proximity to the washer-disinfector's own measurement points. In addition, measurement points must be selected on the load carriers and on the chamber walls of the...

Open the catalog to page 4
Wireless Data Logger EBI 10 for Validation - 5

Process validation in steam-sterilizers Laws and Standards Various laws and standards call for appropriate validated processes in order to avoid hazards for patients and hospital staff emanating from processed medical products. These regulations are based on the standards ISO EN 17665, EN 285 (steam-sterilizers), or pr DIN 58929 / DIN EN 13060 (small steam-sterilizers) as well as ISO 15883 (washer-disinfectors). The aim of the technical validation process is to provide a documented verification that each particular steam-sterilizer supplies the sterilization results required. Thermoelectric...

Open the catalog to page 5
Wireless Data Logger EBI 10 for Validation - 6

a xylem brand ebro® Electronic GmbH PeringerstraBe 10 D-85055 Ingolstadt Tel. +49 (0) 841 - 9 54 78-0 Fax +49 (0) 841 - 9 54 78-80 Internet: www.ebro.com e-Mail: info@ebro.com All names are registered tradenames or trademarks of ebro® Electronic GmbH -Specifications are subject to change without notice-