http://www.itm-lab.com/show/134.html
4Pages

{{requestButtons}}

Catalog excerpts

http://www.itm-lab.com/show/134.html - 1

Windless Natural Convection Constant Temperature Chamber Model:RS-600LH 1.Use of equipment This equipment is used for the testing of radiator modules, power supplies, servers, etc. It provides a cavity for the test environment under the condition of natural convection and no external air supply. The temperature control is achieved by the balance between the heater and the natural convection of the external environment. test environment. 2.The heating principle: The infrared heater installed in the cavity heats the entire test environment by means of "thermal radiation" and "natural convection" heat transfer. By controlling the input voltage and adjusting the output power ratio of the heater, the purpose of constant temperature control in the cavity is achieved. With the optimized structural design, when the thermal equilibrium state is reached, the cavity test space has a uniform temperature and a very low wind speed. It can also isolate the wind speed interference of the external environment, which is analogous to the operation of the product under the condition of natural convection of indoor air and strict high temperature environment.

Open the catalog to page 1
http://www.itm-lab.com/show/134.html - 2

3. Features: 1. Simulate the environmental test of the constant temperature chamber under the condition of natural convection and no external air supply. 2. The temperature inside the chamber is adjusted in PID mode. 3. Chamber temperature range: room temperature +3°C~70°C. 4. With a test piece placement platform with adjustable position, grid structure. 4. Objects to be tested: Radiator modules, servers, desktops, laptops, game consoles, Power supply, LED and other electronic products. 5.Technical Parameter 1. The chamber body is made of metal frame and 20 mmt transparent plastic plywood,...

Open the catalog to page 2
http://www.itm-lab.com/show/134.html - 3

composed of 10 independent horizontal bars, In the horizontal position 30 cm stage height adjustable position setting, the horizontal bar is detachable. 6. Chamber heating heater, with thermal isolation safety cover. 7. Room temperature +3°C~70°C. 8. A set of 2-point average temperature display and control unit is attached. 9. A 9-point temperature measurement report is attached when leaving the factory. Temperature uniformity: The heater is above 40 cm, 50% of the central space, and the standard deviation between the measured temperature at each point and the average temperature at 9...

Open the catalog to page 3