RESIDUAL CHLORINE ANALYZER

RESIDUAL CHLORINE ANALYZER

RESIDUAL CHLORINE ANALYZER

Product catalog summary
Overview: The Model CLF-100 Residual Chlorine Analyzer is designed to measure residual chlorine concentration in city water and sewerage effluent post-chlorination. It can measure both residual and free available chlorine concentrations using specific reagents.
Features: The analyzer features a long-term stable measurement system with a contact-free swing rotary residual chlorine electrode and a beads cleaning system. It includes a transmitter with a built-in microcomputer for diagnostics, such as flow rate fault detection and calibration errors. Optional auto-calibration and auto-cleaning systems are available for enhanced reliability.
Specifications:
  • Measurement Method: Polarography using an eccentric rotary micro-electrode.
  • Sensor: CLR-31 swing rotary with built-in temperature compensation.
  • Measurement Range: Various ranges from 0~0.5 to 0~10 mg/L, selectable manually or remotely.
  • Output Signal: 4~20mA DC, isolated output.
  • Sample Conditions: Temperature 0~40°C, Pressure 0.02~0.3 MPa.
  • Power Requirements: 100V/110V AC, 50/60Hz.
Reagents: The analyzer uses specific reagents for chlorine measurement, with compositions detailed in tables for residual and free chlorine measurement.
Options:
  • Automatic Cleaning: Water or ozone cleaning, with manual, auto, or remote start modes.
  • Automatic Calibration: Zero and span calibration methods, with manual, auto, or remote start modes.
Performance:
  • Repeatability: ±2%FS.
  • Linearity: ±3%FS.
  • Response Time: Within 2 minutes for 90% response.
Principle of Operation: The system includes a sensor, transmitter, sampler, and solution pump. The sample reacts with the reagent to liberate I2 or Br2, which is reduced by the sensing electrode, and the resulting current is amplified to indicate chlorine concentration.
Related Equipment: Includes a 50-liter reagent tank and a sand filter for sample preparation.
Contact Information: DKK-TOA Corporation, Tokyo, Japan, and European Representative Office in Cambridge, UK.
See more

Catalog excerpts

RESIDUAL CHLORINE ANALYZER-1

This instrument is used for measuring the residual chlorineconcentration of city water and sewerage effluent after chlorination (both residual chlorine concentration and free available chlorine concentration can be measured by using the appropriate reagents). FEATURES Long term, stable measurement using a sensor which contains acontact-free swing rotary residual chlorine electrode (includes beads cleaning system).Օ Transmitter with built-in microcomputer including various diagnosticfunctions such as flow rate fault detection and calibration error etc. Stable and reliable operation can be further enhanced by adding anoptional auto-calibration system and optional auto-cleaning system. STANDARD SPECIFICATIONS Product Name :Residual chlorine analyzer. Model :CLF-100. Measurement Object :Residual chlorine (total chlorine) ofwater or free available chlorine (free chlorine). Measurement Method :Polarography by eccentric rotarymicro-electrode. Sensor :CLR-31 swing rotary (with built-in Pt-Pt temperature compensationresistor) sensing electrode, type 2133. Measurement Range :Select one of the following:-0~0.5/1.0 (Residual chlorine measurement only).0~1/2, 0~2/5, 0~3/6, 0~5/10 mg/ Reagents Composition :Residual chlorine measurement: Table 1.Free chlorine measurement: Table 2.Consumption: Approx 1.1 m (ppm units are also available ifspecified). /min(Approx 1.6 Measurement Range :Manual or remote switching (client to /day). Switching specify). Indication :Digital (LCD 0.00~10.00). > Table 1: Residual chlorine measurement (composition of 50 liters). Output Signal :4~20mA DC, maximum load 600W,isolated output. > Extra pure grade reagent Measurement range0~6mg/ or Less0~7mg/ or GreaterExtra pure grade reagentMeasurement range 0~10mg/ Sample Conditions Contact Switching Inputs Contact Switching Outputs :Temperature: 0~+40 :Range indication, undermaintenance, power fail, low sample flow, low reagent, flow error,temperature fault, in cleaning mode,in calibration mode, calibration error, (contact rating: 30V DC, 0.1A). High concentration, low concentration (contact rating: 30V DC, 0.1A). :Range switching (open contacts=lowrange, closed contacts=high range)(contact rating: 30V DC 0.1A). Start cleaning, start calibration (contact closed for 100mS or longer) (contact rating: 30V DC, 0.1A). > KI 250g500g NaCH 3 CO 2 Anhydride 62.5g 125g C 2 H 4 O 2 500ml 1000ml Table 2: Free effective chlorine measurement (composition of 50 liters). or less. KBr 1500g NaCH 3 CO 2 Anhydride 500gC 2 H 4 O 2 500ml Wetted Materials :Hard PVC, Teflon tubing, polyethylenetubing. Liquid Transfer :Sample and reagent solution aretransferred to flow cell by solution pumps. Power Requirements :100V/110V AC ձ10%, 50/60Hz. Power Consumption :Approx. 60 VA (main body ofinstrument; approx. 40VA, with cleaner; approx. 50 VA, withcalibration; approx. 60VA). Construction :Indoor, self-standing, drip-proof. > o C (no freezing).Pressure: 0.02~0.3 MPa (0.2~3kgf/cm Operating Temp Range :-5~+50 > 2 ).Consumption: 1~3 /min. (flow rate ofsample to be introduced into analyzer: 50m > o C. /min). Operating Humidity Range :<85% RH. >

 Open the catalog to page 1
RESIDUAL CHLORINE ANALYZER-2

tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:5.00pt "Arial", sans-serif; } .font1 { font:6.00pt "Arial", sans-serif; } .font2 { font:7.00pt "Arial", sans-serif; } .font3 { font:8.00pt "Arial", sans-serif; } Main Materials: Transmitter: ADC12 (Al. die cast). Sensor: A1050 (Al.). Frame: A1050P (Al.). : Transmitter/sensor: Pantone 537C (Equivalent to Munsell 5PB8/1). Frame: Grey (Equivalent to Munsell N6). : Approx. 20kg. : Sample inlet: VP16 pipe, reagent inlet: V16 pipe, Drain: VP25 pipe, cleaning water: VP16 pipe (option). : Four 6mm~12mm cable entries. Calibration Period...

 Open the catalog to page 2
RESIDUAL CHLORINE ANALYZER-3

SymbolDescription CV1Check valveCV2Check valveOSReagent optical flow sensorPRPressure regulatorP1Sample pump P2Reagent pumpBV1Sample adjustment valveBV2Reagent stop valveBV3Sample supply valveBV4Span supply valve BV5Span stop valveBV6Cleaning water stop valveBV7Zero sample stop valvePV1Pinch valve for stopping sampleSV1Solenoid valve SV2Solenoid valve *Option > 3 GeneratorFAir filterSV3Solenoid valve >

 Open the catalog to page 3
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.