Product catalog summary
Overview: The document details the specifications and features of the HDM-135A and HDM-136A pH transmitters, which are designed for precise pH measurement and control in industrial settings. These devices are robust, user-friendly, and suitable for on-site installation.
Key Features:
  • Calibration: Simplified process with up to five pH standard solutions stored in memory, including automatic electrode quality determination.
  • Temperature Measurement: Displays temperature using a pH electrode with an internal sensor, offering both manual and automatic temperature compensation.
  • Output Control: Features an output hold during maintenance to prevent disruptions, with adjustable measured pH values for operational control.
  • Diagnostics: Self-diagnostic functions detect hardware failures and electrode degradation, providing error messages and switching output to notify faults.
  • System Configuration: Available in 2-wire (HDM-135A) and 4-wire (HDM-136A) systems, with various power supply options and transmission outputs.
Specifications:
  • Measurement Range: pH: -1.00 to 14.00, mV: -600 to 600 mV, Temperature: 0 to 100ºC.
  • Display: Digital LCD with precision of pH; 0.01, mV; 0.1, Temperature; 0.1ºC.
  • Output Signal: 4~20mADC, isolated, with adjustable range between -1~14pH.
  • Power Supply: 24VDC for 2-wire, 90~132VAC or 180~264VAC for 4-wire systems.
  • Construction: Outdoor installation, IP55 rated, with a compact aluminum case.
Installation and Wiring:
  • Includes detailed wiring diagrams for both 2-wire and 4-wire systems, with options for direct connection or via junction box and extension cable.
  • Optional features include external input for output hold command and various surface finishes and coatings.
Optional Accessories:
  • Junction box and extension cable for installations where standard electrode lead length is insufficient.
  • Power supply unit and bar graph meter relay for additional functionality and integration with control systems.
Specifications and Applications: The document outlines various models of pH electrodes and their applications based on temperature and installation conditions. Models such as HC-G70, HC-G80P, and HC-G82P are designed for general process use and effluent treatment, with specific features like weather-proofing, heat-proofing, and pressure resistance.
Material and Features: The wetted part materials include PVC, PP, SUS316, PVDF, and PFA, each chosen for specific environmental resistances such as high temperature, chemical resistance, and pressure resistance.
Classification: The electrodes are classified into immersion and flow-through types, with specific models designed for high alkali resistance and hydrofluoric acid resistance.
Supported Detectors: Detectors compatible with the HDM-135A/136A controller are listed, with recommendations to select based on immersion type, flow-through type, material, and measurement conditions.
Contact Information: DKK-TOA Corporation provides international operations support, with contact details for further inquiries.
Note: Information is subject to change without notice, and the document was issued on 5/10/2012.
See more

Catalog excerpts

HDM-135A (2-wire system) HDM-136A (4-wire system)-1

The pH transmitter is a user-friendly device that features a compact and robust aluminum case, making it ideal for on-site installation. This device comes in two different models: the 2-wire system (24VDC power supply) and the 4-wire system (adjustable-voltage AC power supply). Features Simplified calibration with standard solutions The data for up to five different pH standard solutions can be stored in the internal memory for single-action calibration. Stability judgement function provides accurate calibration with standard solutions, free from operator error. Automatic determination of electrode quality The transmitter judges the electrode quality from its characteristics during calibration with standard solutions. Degradation of electromotive force at pH7, degradation o felectromotive force per pH and other information is displayed in the form of error messages. Characteristic data of each electrode can be called out to determine the extent of degradation as required. Temperature display The analyzer/transmitter measures and displays the temperature of samples and standard solutions using a pH electrode equipped with an internal temperature sensor (5600, GSS-304B, or other model). Output hold while performing maintenance work When the transmitter enters maintenance (ST-BY) mode, the previous output value is held. This helps to prevent disruptions to the control system. Measured value shift Measured pH values can be shifted for operational control. (Shift width: ±1.0pH) Manual temperature compensation Manual temperature compensation function (0-100ºC) can be provided for use with electrodes which do not have compensation functions. pH temperature compensation The transmitter compensates the pH temperature characteristics of samples (such as pure water and boiler water). Setting range of temperature compensation coefficient : ±0.1pH/ºC Standard conversion temperature: 25ºC Self-diagnostics The transmitter detects damage on the glass membrane, problem with temperature compensation resistance, hardware failures and abnormal data. If a fault is detected, the transmission output is switched to the upper or lower limit to notify the fault (burn-out function). Automatic reversion to measurement mode The analyzer/transmitter automatically switches back to measurement mode if it is left in maintenance mode for more than two hours. External input for “hold” feature (option) The transmitter can receive a “hold” command signal from the cleaning devices to hold output during the cleaning. System configuration Transmitter HDM-135A Instrument room Electrode lead Indicator Recorder Controller DCS, etc. Indicator Recorder Controller DCS, etc. Electrode lead Instrument room

 Open the catalog to page 1
HDM-135A (2-wire system) HDM-136A (4-wire system)-2

Features Product name Model pH transmitter HDM-135A (2-wire system) HDM-136A (4-wire system) pH: -1.00~14.00 mV: -600~600 mV Measurement range Temperature: 0~100ºC Display only. No output-signal. Digital LCD Display pH; 0.01, mV; 0.1, Temperature; 0.1˚C Precision Within ±0.03pH (at equivalent input) Performance Linearity Within ±0.02pH or less (at equivalent input) (excluding detector) Repeatability Output signal 4~20mADC, isolated. Load resistance: Max 650Ω or less. Adjustable range between -1~14pH (0.1pH steps). Output range Minimum width of 2 pH. Power supply 24VDC ±10% 90~132VAC 50/60Hz or...

 Open the catalog to page 2
HDM-135A (2-wire system) HDM-136A (4-wire system)-3

Mounting bracket (for 50A pipe) (Brackets fit either side) Optional terminal cable port G 3/4 (PF 3/4 F) (with gasket for ø11 cable) Power/transmission cable port G 3/4 (PF 3/4) with gasket for ø11 cable Electrode lead port G 3/4 (PF 3/4 F) with gasket for ø6 lead Mounting bracket (for 50A pipe) (Brackets fit either side) Power cable port G 3/4 (PF 3/4 F) (with gasket for ø11 cable) Transmission output cable port G 3/4 (PF 3/4) with gasket for ø11 cable Electrode lead port G 3/4 (PF 3/4 F) with gasket for ø6 lead Hood (option) Recommended when installing the instrument in a location exposed to...

 Open the catalog to page 3
HDM-135A (2-wire system) HDM-136A (4-wire system)-4

Wiring diagrams for 2-wire system Electrode lead <When electrode is directly connected to transmitted -y^rr\ Power / transmission output line (4~20mADC) <When the electrode is connected via a junction box and extension cable> Junction box cr in Fvioncinn r-ahio Transmitter Electrode lead <Wiring example when using the device together with a RHC-7C chemical cleaner> When the transmitter is equipped with the external input for the output hold command Electrode lead Power / transmission output line Transmission output Load resistance: 650Q or less Output hold command line Chemical cleaner Cleaning...

 Open the catalog to page 4
HDM-135A (2-wire system) HDM-136A (4-wire system)-5

Wiring diagrams for 4-wire system Electrode lead When the transmitter is equipped with the input for the external output hold command (terminals #10~#11 (option)) Electrode lead HDM-136A transmitter 1 2 3 4 5 E Transmission output line Output hold command line Class D Chemical cleaner Control unit Output 4~20mADC is held during cleaning or when the mode selector switch is set to ST-BY. AC power supply (100VAC 50/60Hz) Class D Liquid transfer unit NC NO Air supply Pressure reducing valve with gauge (setting range 0.06~0.1 MPa): Option Liquid transfer tube 3 Connection terminals description Legends...

 Open the catalog to page 5
HDM-135A (2-wire system) HDM-136A (4-wire system)-6

Product code Custom spec. code: Numeric digit: 9 Alphabet: Z Transmission output range (4~20mADC) 0~14 pH 0~10 pH 0~8 pH 2~12 pH 4~14 pH 4~10 pH 6~14 pH Custom spec.*1 Input for external output hold command*2 None Equipped Surface finish (coating)*3 Standard coating High performance coating Arrester*4 None Included Assembly with cable port adapter None, G3/4 (PF3/4) standard G1/2 (PF1/2) SUS304 NPT1/2 SUS304 NPT3/4 SUS304 Electrode lead port Direct connection EC-10 Extension cable (connection with FC-4) Hood (sun shade) None Equipped (No. 544493K) Markings Japanese (standard) English Official...

 Open the catalog to page 6
HDM-135A (2-wire system) HDM-136A (4-wire system)-7

Extension cable The extension cable is a special cable specifically manufactured for a pH/ORP analyzer. It connects the transmitter and junction box. Conductor (0.3mm2) PE insulation (natural color) Semi-conductive PVC Shielding PET tape winding PE core sheath (black) Conductor (0.3mm2) PE insulation (green) PE insulation (red) Filler Cloth tape winding Shielding PVC sheath PE insulation (blue) Model : EC-10 Outside diameter : ø8 Insulation : Polyethylene and PVC Sheath : PVC Insulation resistance between core conductors : 105MΩ or greater/100m. Maximum cable length : 100m, no cable splicing....

 Open the catalog to page 7