1. Catalogs
  2. DJB Instruments UK Ltd
  3. A/120/CR Calibration Reference IEPE Accelerometer

A/120/CR Calibration Reference IEPE Accelerometer

A/120/CR Calibration Reference IEPE Accelerometer

Product catalog summary
Specifications:
The A/120/CR is a general-purpose Konic Shear IEPE accelerometer designed for calibration or validation checks. It requires a constant current supply and features a standard 10/32 tapped thread for back-to-back calibration. The device is available in different sensitivities, with a maximum test accelerometer mass of 25 grams.

Electrical Characteristics:
  • Voltage/Charge Sensitivity: ±10% at 1.0mV/(m/s²) or 10.2mV/(m/s²) (Metric), 10mV/g or 100mV/g (Imperial)
  • Resonant Frequency: 28 kHz
  • Typical Frequency Response: 1Hz – 8kHz (Metric), 0.7Hz – 9kHz (Imperial)
  • Cross Axis Error: ≤5%
  • Temperature Range: -50/+125°C (-58/+257°F)
  • Charge Sensitivity Deviation: -5% at -50°C, +5% at +125°C
  • Shock Level: 9807m/s² (1000g)
  • Base Strain Sensitivity: 0.001g/µstrain
  • Supply Voltage: 15/35 V DC
  • Supply Current: 2/20 mA
  • Bias Voltage: 10/14 VDC
  • Setting Time to 90% Final Value: <3 sec

Physical Characteristics:
  • Case Material: Titanium
  • Mounting: Base tapped hole, 10-32 UNF x 4mm deep (Metric), 0.16in deep (Imperial)
  • Weight: 24g (0.85oz)
  • Connector: 10-32 UNF Microdot
  • Case Seal: Welded hermetic connector
  • Size: 14.3 (A/F) x 25mm (Metric), 0.56 (A/F) x 0.98in (Imperial)

Performance:
  • Typical Spectral Noise (100mV/g): 1Hz - 732 µg/√Hz, 10Hz - 82.1 µg/√Hz, 100Hz - 16.2 µg/√Hz, 1kHz - 4.2 µg/√Hz, 10kHz - 3.1 µg/√Hz

Options:
  • A/120/CR
  • A/120/V
  • A120VI
  • A120VT
  • A120VTC
  • A/120/VTI

Note: The data provided is for information and reference only and should not be used as pass/fail criteria for calibration purposes.
See more

Catalog excerpts

A/120/CR Calibration Reference IEPE Accelerometer-1

A/120/CR Calibration Reference IEPE Accelerometer 10mV/g or 100mV/g ±10% General purpose Konic Shear IEPE accelerometer for use as a reference device in calibration or validation checks. Requiring a constant current supply. With a standard 10/32 tapped thread in the top of the accelerometer this allows back to back calibration of other accelerometers ensuring both devices operate along the same axis. Available with a range of stud adaptors to allow the connection of numerous types of accelerometer to the integral 10-32 UNF tapped hole. The A/120CR can be supplied in a range of different sensitivities to suit your requirement. Please let our sales team know the stud size and sensitivity you require when ordering. Maximum test accelerometer mass 25grams Temperature Response Options: Typical Frequency Response Typical Spectral Noise (100mV/g) Voltage/Charge Sensitivity 1.0mV/(m/s2) or 10.2mV/(m/s2) 10mV/g or 100mV/g ±10% Resonant frequency 28 kHz Typical Frequency ±5% 1Hz – 8kHz Response (unloaded) ±10% 0.7Hz – 9kHz Cross axis error ≤5% Temperature range -50/+125°C -58/+257°F Charge sensitivity -5% @ -50°C -5% @ -58°F deviation (20°C/68°F) +5%@ +125°C +5%@ +257°F Shock level 9807m/s2 1000g Base Strain Sensitivity 0.001g/µstrain Supply voltage 15/35 V DC Supply Current 2/20 mA Bias voltage 10/14 VDC Setting time to 90% final val. <3 sec Case material Titanium Mounting Base tapped hole, 10-32 UNF x Base tapped hole, 104mm deep 32 UNF x 0.16in deep Weight 24g 0.85oz Connector Case Seal Size 10-32 UNF Microdot Welded hermetic connector 14.3 (A/F) x 25mm 0.56 (A/F) x 0.98in Please note: For information and reference only. Data should not be used as pass / fail criteria for calib

 Open the catalog to page 1