Overview: The HLX22-T Series Temperature Transmitter features interchangeable sensing probes with connectors, allowing for easy replacement without affecting performance. Calibration data is stored in the probes, ensuring consistent accuracy across the temperature range.
Specifications: The transmitter offers high accuracy due to precise calibration methods and advanced microprocessor technology. It supports high-temperature applications up to +80°C (+176°F) and can be connected via cables of varying lengths (2m, 5m, 10m) without compromising accuracy. Output options include voltage (0-1V, 0-10V) and current (4-20mA), with scalable temperature output.
Installation and Configuration: The HLX22-T is suitable for wall mounting and DIN rail installation. An optional duct mounting kit is available. The device can be configured with or without a display to show current temperature values.
Technical Data: The sensing probe uses a Pt1000 sensor element with a working range of -40 to 80°C. The device features a response time of less than 6 minutes and a temperature dependence of electronics at ±0.007°C/°C. The housing is available in metal or polycarbonate, with an IP65 protection class.
Applications: The HLX22-T is ideal for use in the pharmaceutical industry, clean rooms, storage rooms, greenhouses, and cooling chambers, offering accuracy of ±0.1°C at 20°C.
Ordering Guide: The transmitter can be customized with different housing materials, output types, and probe configurations. Cable lengths and additional accessories such as display covers and mounting kits are available.
Dimensions: The housing dimensions vary slightly between polycarbonate and metal options, designed for robust use in industrial environments.
See more

Catalog excerpts

HLX22-T Series-1

Temperature Transmitter with interchangeable probes Unique for the HLX22-T series are the interchangeable sensing probes with connector. The calibration data is stored in the probes, which are therefore interchangeable and probe replacement does not affect the performance of HLX22-T. The outstanding accuracy over the entire temperature range is based on very precise calibration methods and on the latest microprocessor technology. Well-proven humidity sensor elements ensure excellent long-term stability. For high temperature applications (up to +80°C / +176°F) or in case of limited space availability, the sensing probes can be connected to HLX22-T housing with cables (2m, 5m or 10m / 6.6ft, 16.4ft or 32.8ft) without any repercussions for the overall accuracy of the instrument. Voltage 0 - 1 / 10V or current 4 - 20mA (2 wire) outputs are available, of which the temperature output can be scaled according to the application (see ordering guide). polycarbonate housing HLX22-T is suitable for direct wall mounting and for installation on rails according to DIN EN 50022. For easy duct mounting a duct mounting kit is available as an option. An optional display indicates the actual T values. metal housing Probe Dimensions (mm) with plugable T probe HLX22-xTx1x with remote T probe HLX22-xTx1x +HAxxxx

 Open the catalog to page 1
HLX22-T Series-2

Typical Applications accuracy ±0,1°C at 20°C interchangeable probes remote sensing probe up to 10m (32.8ft) measuring range -40...80°C (-40...176°F) optional display traceable calibration cost saving, easy loop-calibration of T probes pharmaceutical industry clean rooms storage rooms green houses cooling chambers Technical Data Measuring values of sensing probe Temperature Sensor element Working range sensing probe Accuracy Pt1000 (tolerance class A, DIN EN 60751) fixed sensing probe: -40...60°C (-40...140°F) remote sensing probe: -40...80°C (-40...176°F) ∆° C Temperature dependence of electronics...

 Open the catalog to page 2
HLX22-T Series-3

Housing Dimensions (mm) polycarbonate housing For use in harsh industrial environments all models of HLX22-T series are available in a robust metal housing. The smooth surface and the rounded outlines allow for the use in clean room applications. Position 1 - Transmitter Hardware Configuration Housing metal housing polycarbonate housing temperature 0-1V 0-10V 4-20mA wall mounting - cable gland M16x1.5 wall mounting - rear cable outlet 1 probe T without display with display without plug 1 plug for power supply and outputs Type Output Model Probe Display Plug Software Configuration T-Unit Scaling...

 Open the catalog to page 3