SP325 & TP325 Logger Operation - 1

Rev. 05/09 DICKSON SP325 & TP325 Logger Operation Contents: Product Applications and Useful Features Product Specifications Getting Started DicksonWare Software Specifications Product Accessories Frequently Asked Questions Calibrations Troubleshooting Warranty / Factory Service & Returns Product Applications & Useful Features Product DICKSON Specifications SP325 Self Contained Temperature TP325 Temperature & Humidity Troubleshooting /Warranty & Returns Product Accessories & FAQs Getting Started Getting Started DicksonWare Calibrations Specifications

Open the catalog to page 1
SP325 & TP325 Logger Operation - 2

Rev. 05/09 Product Applications Dickson’s Pro Series Temperature Data Logger now feature a LCD display and push-to-start! The new SP325 & TP325 feature high speed USB connectivity, user replaceable battery, a current min/max display, and convenient push-to-start functionality setting a new standard in accurate, easy -to -use, affordable PC based recording. Product Features • 1 year battery life • Stores 32,000 sample points • Push-to-Start • Digital Display • Compact size, fits in the palm of your hand Product Applications & Useful Features Product DICKSON Specifications Troubleshooting...

Open the catalog to page 2
SP325 & TP325 Logger Operation - 3

Rev. 05/09 Product Specifications SP325 Operating Range (unit): -4 to +158°F (-20 to +70°C) (non-condensing) Internal Temp. Sensor: Internal digital temperature sensor Internal Temp. Accuracy: ±1.2°F (±0.67°C) IP Rating IP20 Storage Capacity: 32,512 points Resolution: 1°F/1% RH Dimensions: 1.34” x 3.08” x 2.25” (3.4 cm x 7.8 cm x 5.7 cm) Minimum DicksonWare Version Required: 11.0 TP325 Operating Range (unit): -4 to +158°F (-20 to +70°C) (non-condensing) Internal Temp. Sensor: Internal digital temperature sensor Internal Temp. Accuracy: ±0.8 °F (±0.5°C) over range +20 to +120°F (-6 to...

Open the catalog to page 3
SP325 & TP325 Logger Operation - 4

Rev. 05/09 Product Applications & Useful Features Product Specifications Troubleshooting /Warranty & Returns DICKSON Getting Started To setup your SP325 or TP325 you’ll need to be at your PC. 1. Install DicksonWare v.11 or higher Software on your PC. a. Windows 98 or higher is required. b. DicksonWare v.11 or higher is required. If you currently have DicksonWare on your PC, check the version by choosing Help and About from the menu bar. c. To install and start DicksonWare Software onto your PC, insert the CD into the CD drive. Installation will start automatically. 2. Connect the cable...

Open the catalog to page 4
SP325 & TP325 Logger Operation - 5

Rev. 05/09 Product Applications & Useful Features Product DICKSON Specifications Display/Push to Start The display will remain blank when Press Button to Start has been selected in Setup but logging has not yet begun. If the display is flashing the unit is not logging. Min/Max Display • There are two display options: Current Readings only or Current/Min/Max scroll, which can be selected under Setup/Display. • To reset Min/Max readings on the display, a paper clip or toothpick should be used to depress the Min/Max Clear button located on the back of the unit. This will not clear logged data....

Open the catalog to page 5
SP325 & TP325 Logger Operation - 6

Rev. 05/09 Product Applications & Useful Features Product DICKSON Specifications DicksonWare™ Software Specifications • Microsoft Windows® compatible • Allows for simple viewing and zooming of logged data • Easy set-up of Dickson Data Loggers including: • User selectable sample intervals from 1 seconds to 24 hours • Display temperature in °C or °F • Delayed logger start times • Logger data capacity can be set to wrap data or stop when full • Allow for real time monitoring and graphing. • Effortless exporting of data and graphs to other software • Data can be viewed in tabular...

Open the catalog to page 6
SP325 & TP325 Logger Operation - 7

Rev. 05/09 Product Applications & Useful Features Product DICKSON Specifications Accessories (for current pricing go to www.dicksonweb.com or call 1-800-323-2448) Software Order # DicksonWare™ Software and Serial Download Cable A016 Logger Calibration Software SW400 Calibrations NIST Traceable Calibration 3-pt (new unit) N300 NIST Traceable Calibration 1-pt (new unit) N100 A2LA Accredited Calibration 3-pt. (new unit) N400 Certificate of Validation N520 Frequently Asked Questions Sample Interval: The Sample Interval determines how frequently the data logger will save a reading. Using...

Open the catalog to page 7
SP325 & TP325 Logger Operation - 8

Rev. 05/09 Product Applications & Useful Features Product DICKSON Specifications Calibration Services - New Units N100 - NIST Traceable Calibration 1-Point: Includes documentation to one Dickson pre-selected point on new units only. N300 - NIST Traceable Calibration 3-Point: Includes documentation of three Dickson pre-selected points (a high, medium, and low) on new units only. N400 - Deluxe A2LA Accredited NIST Traceable Calibration 3-Point: ISO Guide 25/A2LA Documentation of 3 pre-selected points of as found data before and after calibration for Dickson temperature and/or humidity...

Open the catalog to page 8
SP325 & TP325 Logger Operation - 9

Rev. 05/09 Product Applications & Useful Features Product DICKSON Specifications Warranty Dickson warrants that the products it sells will be free from defects in material and workmanship under normal use and service for a period of twelve months after delivery. In the event of a claim under this warranty, the product or part must be returned to the factory for repair or replacement (shipping pre-paid) with a Return Authorization Number (see Return Information above). It will be repaired at Dickson’s option without charge. This warranty DOES NOT cover routine calibration, pen, chart and...

Open the catalog to page 9