Rev. 0709 DICKSON PR125/PR325/PR525 Pressure Logger Operation Contents: Product Applications and Useful Features Product Specifications Operating Instructions / Getting Started DicksonWare Software Specifications Product Accessories Calibrations Troubleshooting & Battery Warranty / Factory Service & Returns Applications, Useful Features, & Specifications DICKSON Operating Instructions / Getting Started Software Specifications Accessories & Calibrations Troubleshooting & Warranty/ Returns

Open the catalog to page 1
PR125/PR325/PR525 Pressure Logger Operation - 2

Rev. 07/09 Product Specifications Pressure Range: PR125: 0 - 100 PSI , PR325: 0 - 300 PSI, PR525: 0 - 500 PSI Accuracy: ±1% of full scale Ambient Operating Conditions: -2222 to +176¢ªF (-30 to +80¢ªC) 0 to 100% RH IP Rating IP68 Units of Measure: Via DicksonWare: PSI, ATM, Bar, mBar, InH2O, FtH2O, mmH2O, mmHG, InHg, Pascal, kP, MCA; Via Display: PSI, ATM, mBar, InH2O, kP, MCA, BAR Approvals: CE Battery Type: 1 DL12123 Lithium Battery (User Replaceable) Battery Life: 6 months at 5 Minute Sample Rate Fitting: 1 /4” NPT Thread, 5/8” Hex Sample Point Storage: 60,000 points Sample Interval: User...

Open the catalog to page 2
PR125/PR325/PR525 Pressure Logger Operation - 3

Rev. 07/09 Applications, Useful Features, & Specifications DICKSON Getting Started Install DicksonWare version 13.0 or greater on your PC (compatible with windows 98 or higher) Install battery a. Carefully remove case lid b. Remove battery compartment lid carefully by twisting cap clockwise c. Insert battery positive side up d. Replace battery cap e. It may take up to 5 seconds for unit to power up Open DicksonWare software via the icon on your desktop. Connect the USB cable to the logger (port located to the left of the display) and to an available working USB port on your PC. Your...

Open the catalog to page 3
PR125/PR325/PR525 Pressure Logger Operation - 4

Rev. 07/09 Applications, Useful Features, & Specifications DICKSON Operating Instructions / Getting Started Useful Features Real Time Monitoring Monitor and graph your process as it happens and print and save from real time screen. (This feature should be used sparingly as it causes the battery to drain very quickly.) Customize Graphed Data DicksonWare™ calculates Min, Max and Average of all data collected. Customize data by eliminating unnecessary data points and customize Min, Max and Average to show only desired information. Export Data A snapshot of your graph or a real-time graph of...

Open the catalog to page 4
PR125/PR325/PR525 Pressure Logger Operation - 5

Rev. 07/09 Applications, Useful Features, & Specifications DICKSON Operating Instructions / Getting Started Water Hammer Mode Indicator Logger Full Indicator (Logger has stopped logging) Min/Max Readings IP Rating: This logger has an IP rating of 68 This logger has maintained the IP68 rating under 50 meters of water. A silicone lubricant must be applied to the cap O-ring in order to achieve these results. Installing a Flash Memory Card Reader Follow the instructions included with the flash card reader. Additional information and troubleshooting 3. 4. 5. . . DicksonWare™ Software...

Open the catalog to page 5
PR125/PR325/PR525 Pressure Logger Operation - 6

Applications, Useful Features, & Specifications DICKSON Operating Instructions / Getting Started Calibration Services - New Units N100 - NIST Traceable Calibration 1-Point: Includes documentation to one Dickson pre-selected point on new units only. N300 - NIST Traceable Calibration 3-Point: Includes documentation of three Dickson pre-selected points (a high, medium, and low) on new units only. N400 - Deluxe A2LA Accredited NIST Traceable Calibration 3-Point: ISO Guide 25/A2LA Documentation of 3 pre-selected points of as found data before and after calibration for Dickson temperature and/or...

Open the catalog to page 6
PR125/PR325/PR525 Pressure Logger Operation - 7

Applications, Useful Features, & Specifications DICKSON Troubleshooting For troubleshooting information, click here for the technical support page. Warranty Dickson warrants that the products it sells will be free from defects in material and workmanship under normal use and service for a period of twelve months after delivery. In the event of a claim under this warranty, the product or part must be returned to the factory for repair or replacement (shipping pre-paid) with a Return Authorization Number (see Return Information above). It will be repaired at Dickson’s option without charge....