Catalog excerpts
XRFX-Ray Fluorescent mental Analyzer http://en.dfmc.cc
Open the catalog to page 1X-Ray Fluorescent Elemental Analyzer (XRF) I. Product Overview XRF is one kind of inline multi-stream and multi-element detecting and analytical instrument. It adopts energy dispersion method and motivates various elements in materials, so that these elements generate complex X-ray mixed energy spectrum. XRF analyzes the mixed energy spectrum and thus to calculates types and contents of elements in materials. XRF is an inline instrument for real-time analysis of element contents in industrial production processes. The instrument directly analyzes and measures slurry and rapidly provides...
Open the catalog to page 2therefore vacancies appear. Such vacancies are filled by outer-shell electrons at high levels rapidly (within 10-15s) so that energy level transition occurs. Energy difference between energy levels is released in the form of characteristic X-rays. When pulps contain several elements, each element will emit several types of characteristic X-rays. Such X-rays received by detector and small electronic pulses will be generated. The peak amplitude of these pulses is proportional to the energy of incident X-rays. The intense of X-rays is proportional to the content of elements in slurry....
Open the catalog to page 32. Technical Parameters • Analysis method : X ray fluorescence spectrum analysis technology • One XRF can measure up to 12 streams • Type of analyzable elements: various elements having atomic number over 20th • Range of analyzable element content: 100% ~ 0.001% • Precision: The precision of the sample is influenced by the composition of the mineral matrix, the nature of the mineral, the grain size and so on. Normally, relative error is 1~4% under high content, 3~6% under low content, 10~20% at extremely low content. • Stability: Under certain conditions, the stability is no more than 0.3%,...
Open the catalog to page 4Flushing water consumption: 10~20 L/h depending on spraying intensity and number of water nozzles in the analytic cell. · Requirements for power supply: Power supply for the analyzer: 380/460VAC, 3 phase. Normal operation power: below 4KW, max. starting peak power: below 8KW, earth resistance of grounding: <4Ω. · Requirements for compressed air: Supply pressure: 0.6MPa ~0.8MPa. Clean and dry compressed air, handled by 0.1 micron filter, with a dew point of less than 2 ℃ (dew point refers to the temperature at which air becomes saturated and produces dew). Compressed air consumption: 25L/h...
Open the catalog to page 5Add.: No.136 Binjiang Middle Road, Yanjiang Development Zone, Dandong,Liaoning, China Tel.: +86 415 3862214 Fax: +86 415 3862272 E-mail: intersales@dfmc.cc http://en.dfmc.cc/
Open the catalog to page 6All DFMC catalogs and technical brochures
-
Ash Analyzer (AA-D)
5 Pages
-
Moisture Analyzer (MA)
8 Pages
-
Particle Size Analyzer (PSI)
7 Pages
-
Ash Analyzer (AA-X)
2 Pages
-
Elemental Analyzer
10 Pages
-
Laser Elemental Analyzer
7 Pages
-
LM
9 Pages
-
PSM
7 Pages
-
EA-Coal
7 Pages
-
Ash Moisture Analyzer
2 Pages
-
Analyzer product of DFMC
17 Pages