General Specifications:
X-Scan f3-series offers a range of X-ray linear array detectors with varying pixel pitches and active area lengths. The X-ray tube kVp range is 20-160, and the scintillator used is GOS. The detectors have a pixel pitch ranging from 0.2 mm to 1.5 mm, with corresponding pixel heights and widths. The maximum scanning speed varies from 40 cm/s to 625 cm/s, depending on the model, with a minimum integration time of 0.24 ms for the fast version. The A/D resolution is 14-bits, and the dynamic range is greater than 2000. The operational voltage is +12 V DC, with a power consumption of 60 W max. The ambient temperature range is 0 - 40ºC, and the relative humidity is 30 - 80%.
Enclosure Models:
The X-Scan f3-series includes various enclosure models with active lengths ranging from 102 mm to 1638 mm. Each model has a consistent width and height of 160 mm and 50 mm, respectively, with weights varying from 5 kg to 20 kg.
Interfaces and Features:
The series supports three different interfaces: RS422 and RS232, USB 2.0, and an optional Ethernet interface. It offers faster scanning speeds with a wide active area and supports pixel averaging from 1x to 256x. Advanced calibration functions and internal temperature monitoring are also included.
Notes:
Note 1: USB 2.0 interface limits active area length to 102-614 mm.
Note 2: USB 2.0 interface limits the number of pixels to 512-3072.
Note 3: Fast version specifications apply to active lengths up to 2048 pixels.
Note 4: Dynamic range is defined as saturation signal/RMS noise.
Note 5: Digital interface default data format is 12-bits.
Note 6: Non-linearity is defined as max deviation from the ideal response line.
Note 7: External power supply for 100-240 AC is available upon request.
Note 8: Relative humidity is non-condensing.
See more

Catalog excerpts

X-ray detector for industrial inspection machine 20 - 160 KeV | X-Scan f3 series -1

general characteristics X-Scan 0.2f3 X-Scan 0.4f3 X-Scan 0.8f3 X-Scan 1.5f3 Notes X-ray Tube kVp Range 20-160 Scintillator GOS Active Area Lenghts 102-820 mm 102-1638 mm 102-1638 mm 96-1536 mm See Note 1) Number of Pixels 512-4096 256-4096 128-2048 128-2048 See Note 2) Pixel Pitch 0.2 mm 0.4 mm 0.8 mm 1.5 mm Pixel Height 0.3 mm 0.6 mm 0.8 mm 3.2 mm Pixel Width 0.1 mm 0.3 mm 0.7 mm 1.4 mm Max Scanning Speed Fast Version 40 cm/s 67 cm/s 80 cm/s 167 cm/s 333 cm/s 625 cm/s See Note 3) Min. Integration Time Fast Version 0.5 ms 0.3 ms 0.5 ms 0.24 ms 0.24 ms 0.24 ms See Note 3) Max Integration Time 128 ms A/D Resolution 14-bits Dynamic Range > 2000 See Note 4) Data Digital Interface 16-bits See Note 5) Linearity (Non-Linearity) < 1% See Note 6) Operational Voltage +12 V DC See Note 7) Power Consumption 60 W max Ambient Temperature 0 - 40ºC Relative Humidity 30 - 80% See Note 8) Storage Temperature -10 - 50ºC Detection Technology, Inc. | Piisilta 1, 91100 Ii, FINLAND | Tel. +358 8 553 6600 | Fax +358 8 553 6611 | http://www.deetee.com | [email protected] Note 1) When the USB 2.0 interface is used, the active area length will be 102-614 mm Note 2) When the USB 2.0 interface is used, the number of pixels will be 512-3072 Note 3) Max. scanning speed (min. integration time) for “fast” version is specified for active length up to 2048 pixels. Note 4) Dynamic range is defined as saturation signal/RMS noise, normal detector sensitivity. Note 5) Digital interface has 16-bit as data format, but user can select actual data to be 14, 12, 10 or 8-bits. The default value is 12-bits. Note 6) Non-linearity is defined as max. deviation from ideal response line defined by zero offset and signal at 80% of dynamic range. Note 7) Operational voltage is supplied to detector. External power supply for 100-240 AC is provided if requested. Note 8) Non condensing. X-Scan f3-series X-ray linear array detector DS0000006G

 Open the catalog to page 1
X-ray detector for industrial inspection machine 20 - 160 KeV | X-Scan f3 series -2

enclosures of the x-scan f3-series Enclosure Model Active Length Length Width Height Max Weight X-Scan 0.xf3-307 102, 205, 256 or 307 mm (4”, 8.1”, 10.1” or 12.1”) 348 mm (13.7”) 160 mm (6.3”) 50 mm (2”) 5 kg (11 lbs) X-Scan 0.xf3-410 410 mm (16.1”) 450 mm (17.7”) 160 mm (6.3”) 50 mm (2”) 6 kg (13 lbs) X-Scan 0.xf3-512 512 mm (20.2”) 552 mm (21.7”) 160 mm (6.3”) 50 mm (2”) 10 kg (22 lbs) X-Scan 0.xf3-614 614 mm (24.2”) 655 mm (25.8”) 160 mm (6.3”) 50 mm (2”) 12 kg (26 lbs) X-Scan 0.xf3-820 820 mm (32.3”) 860 mm (33.9”) 160 mm (6.3”) 50 mm (2”) 15 kg (33 lbs) X-Scan 0.xf3-922 922 mm (36.3”) 962...

 Open the catalog to page 2

All Detection Technology catalogs and technical brochures

  1. X-panel 2301

    2  Pages

  2. X-panel 1214i

    4  Pages

  3. X-GCU

    2  Pages

  4. X-Scan C01