Overview: The document provides comprehensive technical specifications and operational guidelines for various humidity and temperature measurement instruments, including portable devices, data loggers, and probes. It focuses on models such as HD2101.1, HD2101.2, HD21AB, HD21AB17, HD9008TRR, and HD9009TRR, among others, detailing their features, applications, and calibration procedures.
Specifications: The HD2101.1 and HD2101.2 models are portable instruments designed to measure relative humidity and temperature using Pt100 sensors or thermocouple probes. They feature a large LCD display and can calculate parameters like absolute humidity, dew point, and enthalpy. The HD2101.2 includes data logging capabilities, storing up to 38,000 samples, and supports data transfer via RS232C and USB ports. Other models, such as HD9008TRR and HD9009TRR, offer different output configurations and operate within specific temperature ranges.
Technical Characteristics: The instruments generally operate within a temperature range of -5 to 50°C and a humidity range of 0 to 90%RH, with some models supporting extended ranges. They are powered by batteries or external power supplies and have varying protection ratings, such as IP66.
Measurement Capabilities: These instruments measure relative humidity and temperature with high accuracy and resolution. They support various probes, including immersion, penetration, and contact probes, each with specific accuracy and application ranges.
Probes and Modules: The document lists several models of temperature and humidity probes, such as HP472ACR and HP572ACR, and describes the SICRAM module for automatic detection and calibration of probes.
Accessories and Calibration: Accessories include saturated salt solutions for calibration verification and protective covers for probes. Calibration is emphasized for accurate measurements, with detailed procedures provided.
Applications: The instruments are suitable for indoor air quality monitoring, weather stations, and compressed air systems. Specific probes are designed for measuring dew points in compressed air systems to prevent issues like corrosion and ice formation.
Order Codes and Accessories: The document provides order codes for different models and accessories, facilitating easy procurement and setup.
Calibration Procedure for Relative Humidity Sensors: Detailed steps are provided for calibrating humidity sensors using saturated salt solutions, emphasizing temperature equilibrium and proper handling to ensure accuracy.
Instrument Features: Models like HD37AB17D and HD37B17D measure and log relative humidity, temperature, CO, and CO2, with applications in indoor air quality monitoring. They use capacitive sensors for RH and NTC sensors for temperature, with data logging capabilities and USB connectivity.
Connection and Memory: Devices use MiniUSB B-Type connections and have significant memory capacity, with logging intervals affecting storage duration.
Transmitters and Regulators: The HD45 and HD46 series measure and control relative humidity, temperature, and CO2, suitable for indoor air quality monitoring. They offer various output options and are configurable via PC or onboard keyboard.
Installation and Sensor Protection: Installation guidelines emphasize proper sensor positioning to minimize dust accumulation and ensure accurate measurements.
Programming and Serial Communication: Instruments support RS232C and RS485 serial communication, allowing network formation with multiple devices. Programming procedures include setting alarm values and configuring measurement parameters.
Conclusion: The document serves as a comprehensive guide for selecting and using humidity and temperature measurement instruments, highlighting their technical specifications, applications, and calibration procedures.
See more

Catalog excerpts

Relative Humidity & Dew Point-1

- Combined humidity and temperature probes - Saturated salt solutions for calibration verification. Sensor protection. How to verify the calibration. Portable instruments for combined humidity and temperature probes • Accredited ACCREDIA LAT calibration laboratory N° 124 for Humidity Data logger for temperature, humidity, CO, CO2 Indoor air quality monitor Indoor air quality monitor Transmitter and regulator wall mounting series for air quality: temperature, humidity, CO2. With relays, analog and RS485 MODBUS-RTU digital outputs. Series of temperature and humidity transmitters with analogue or digital output (RS485 MODBUS-RTU). With or without display and relay. Series of temperature and humidity transmitters with interchangeable probe. With or without display and relay, analogue and digital outputs. Series of temperature and humidity transmitters with interchangeable probe. Graphic display, relays, analogue and digital outputs. Temperature, humidity, barometric pressure and air speed transmitters. With display, analogue and digital outputs. Temperature and humidity transmitter with analogue output, for weather stations. Solar radiation shield. Temperature and humidity transmitters with analogue and digital outputs RS232, USB, RS485 MODBUS-RTU Absolute humidity transmitters with analogue output. Led displays for 4÷20mA passive transmitter with one or two inputs. Indicators, regulators The qualitative level of our instruments is the result of a continuous evolving of the product itself. This may bring to slight differences between what written in the following manual and the instrument you bought. We cannot completely exclude the presence of errors inside the manual, which we apologise for. Data, images and descriptions included in this manual cannot be enforced legally. We reserve the right to perform modifications and corrections at any time without notice.

 Open the catalog to page 1
Relative Humidity & Dew Point-2

INSTRUMENT TECHNICAL CHARACTERISTICS Instrument Dimensions (Length x Width x Height) 185x90x40mm Weight 470g (complete with batteries) Materials ABS, rubber Display 2x4½ digits plus symbols Visible area: 52x42mm Operating conditions Operating temperature Storage temperature Working relative humidity Protection degree Power Batteries 4 1.5V type AA batteries Autonomy 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20μA Mains Output mains adapter 12Vdc / 1000mA Measuring unit Security of stored data Time Date and time Accuracy Unlimited, independent of battery charge...

 Open the catalog to page 2
Relative Humidity & Dew Point-3

TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT Temperature probes Pt100 sensor with SICRAM module Measurement of temperature by Instrument Pt100 measurement range -200…+650°C Pt1000 measurement range -200…+650°C Resolution 0.1°C Accuracy ±0.1°C Drift after 1 year 0.1°C/year Common characteristics Relative humidity Sensor Measuring range Temperature drift @ 20°C Response time %RH at constant temperature Temperature with sensor Pt100 Temperature drift @ 20°C Class A - Thin film TP49AC.0 Class A - Thin film Class A - Thin film HP480 Measuring range Dew point Working temperature Working...

 Open the catalog to page 3
Relative Humidity & Dew Point-4

Relative humidity and temperature probes complete with sicram module HP472ACR: %RH and temperature combined probe, dimensions Ø 26x170 mm. 2 m connecting cable. HP572ACR: %RH and temperature combined probe, K thermocouple sensor. Dimensions Ø 26x170 mm. 2 m connecting cable. HP473ACR: %RH and temperature combined probe. Dimensions: handle Ø 26x130 mm, probe Ø 14x120 mm. 2m connecting cable. HP474ACR: %RH and temperature combined probe. Dimensions: handle Ø 26x130 mm, probe Ø 14x215 mm. 2m connecting cable. HP475ACR: %RH and temperature combined probe. 2 m connecting cable. Handle Ø 26x110 mm....

 Open the catalog to page 4
Relative Humidity & Dew Point-5

Protection for humidity probes Ø 26 M24x1,5 P1: Stainless steel 200μ grid protection for probes Ø 26 mm. P2: 20μ sintered polyethylene PE protection for probes Ø 26 mm. P3: 20μ sintered bronze protection for probes Ø 26 mm. P4: 20μ sintered PE complete cap for probes Ø 26 mm. Protection for humidity probes Ø 14 M12x1 P6: 10μm sintered complete protection made of AISI 316 stainless steel for probes Ø 14 mm. P7: 20μm sintered complete protection made of PTFE for probes Ø 14 mm. P8: 20μm stainless steel grid and Pocan protection for probes Ø 14 mm. HP480 - T480.1 - S.TC2.480.2 probes for the measurement...

 Open the catalog to page 5
Relative Humidity & Dew Point-6

INSTALLATION The probe can be connected to the compressed air in three ways: A. by using the measuring chamber with a quick coupling; B. by using the measuring chamber with a threaded G 1/4” connection; C. directly (without measuring chamber) with a threaded G 1/2” connection. DIMENSIONS Dimensions (mm) of the probe without measuring chamber: Connection with measuring chamber and quick coupling: Dimensions (mm) of the probe with measuring chamber, without quick coupling: To connect with quick coupling, you can use one of the standard 1/4” couplings provided. Other couplings than those supplied...

 Open the catalog to page 6
Relative Humidity & Dew Point-7

Measurement of relative humidity by Instrument Measurement range 0…100%RH Resolution 0.1%RH Accuracy ±0.1%RH Drift after 1 year 0.1%RH/year Measurement of temperature by Instrument Pt100 measurement range -200…+650°C Pt1000 measurement range -200…+650°C Resolution 0.1°C Accuracy ±0.1°C Drift after 1 year 0.1°C/year Relative humidity and temperature probes using SICRAM module Model HP472ACR HP572ACR HP473ACR HP474ACR HP475ACR HP475AC1R HP477DCR HP478ACR Working range Accuracy %RH Temperature %RH 0…100%RH -20°C…+80°C ±1,5%RH (10…90%RH) 0…100%RH -20°C…+80°C ±2,0%RH (in the remaining 0…100%RH -20°C…+80°C...

 Open the catalog to page 7
Relative Humidity & Dew Point-8

Common characteristics Temperature drift @ 20°C 4 wire Pt100 and 2 wire Pt1000 Probes Model TP47.100.0 Application range Common characteristics Temperature drift @ 20°C Pt100 Pt1000 ORDER CODES HD2301.0: The kit is composed of the instrument HD2301.0, 3 1.5V alkaline batteries, operating manual, case. Probes and cable must be ordered separately. Relative humidity and temperature probes complete with sicram module HP472ACR: %RH and temperature combined probe, dimensions Ø 26x170 mm. 2 m connecting cable. HP572ACR: %RH and temperature combined probe, K thermocouple sensor. Dimensions Ø 26x170 mm....