1. Catalogs
  2. Delta OHM
  3. Dissolved oxygen HD 2109.1
video corpo

Dissolved oxygen HD 2109.1

Dissolved oxygen HD 2109.1

Dissolved oxygen HD 2109.1

Product catalog summary
Overview: The HD2109.1 and HD2109.2 are portable instruments designed for measuring dissolved oxygen concentration, saturation index, and temperature in liquids. They feature a large LCD display and utilize SICRAM combined probes with polarographic technology.
Specifications:
  • Dimensions: 185x90x40mm
  • Weight: 470g
  • Materials: ABS, rubber
  • Display: 2x4½ digits, visible area 52x42mm
  • Operating Conditions: -5 to 50°C, 0-90% RH
  • Protection: IP67
Power:
  • Batteries: 4 x 1.5V AA, 200 hours autonomy
  • Mains Adapter: 9Vdc / 250mA
Data Logging (HD2109.2):
  • Memory: 18,000 samples
  • Storage Interval: 1s to 3600s
  • Interfaces: RS232C and USB 2.0
Measurement Capabilities:
  • Dissolved Oxygen: 0.00 to 90.00 mg/l, accuracy ±0.03 mg/l
  • Saturation Index: 0.0 to 600.0%, accuracy ±0.3%
  • Barometric Pressure: 0.0 to 1100.0 mbar, accuracy ±2 mbar
  • Temperature (Probe): 0 to 45°C, accuracy ±0.25°C
  • Temperature (Pt100): -200 to 650°C, accuracy ±0.25°C
Probes and Modules:
  • Various Pt100 probes for immersion, penetration, and contact measurements with specific accuracy ranges.
  • Oxygen Probe: Polarographic type with replaceable membrane.
Additional Features:
  • Automatic compensation for barometric pressure and salinity.
  • Quick calibration function for dissolved oxygen probes.
  • Max, Min, Avg calculations, REL measurement, Auto-HOLD, and automatic shut-off.
Order Codes:
  • HD2109.1K and HD2109.2K kits include the instrument, calibrator, connection cables, batteries, manual, case, and software. Probes are ordered separately.
See more

Catalog excerpts

Dissolved oxygen HD 2109.1-2

The HD2109.1 and HD2109.2 are portable instruments with a large LCD display. They measure the concentration (in mg/l) of dissolved Oxygen in liquids, the saturation index (in %) and the temperature using SICRAM combined probes of polarographic type with two or three electrodes and integrated temperature sensor. Temperature only is measured by Pt100–SICRAM or direct 4 wire–immer- sion, penetration or contact probes. Thanks to an internal pressure sensor, the instruments automatically compensate for barometric pressure. The instrument anticipates automatic compensation of the Oxygen probe membrane...

 Open the catalog to page 2
Dissolved oxygen HD 2109.1-3

HD2101/USB S‘Print-BT DO9700 DO9701 Measurement of barometric pressure Measurement range 0.0…1100.0mbar Resolution 0.1mbar Accuracy ±2mbar±1digit between 18 and 25°C ±(2mbar+0.1mbar/°C) in the remaining range Setting the salinity Measurement range 0.0…70.0mg/l Resolution 0.1mg/l Temperature measurement with the sensor inside the dissolved Oxygen probe Measurement range 0…+45°C Resolution 0.1°C Accuracy ±0.25°C Drift after 1 year 0.1°C/year Temperature measurement by Instrument with Pt100 probe Pt100 measurement range -200…+650°C Resolution 0.1°C Accuracy ±0.25°C Drift after 1 year 0.1°C/year...

 Open the catalog to page 3
Dissolved oxygen HD 2109.1-4

AF209.60 DeltaLog9: Software for download and management of the data on PC using Windows 98 to XP operating systems. AF209.60: Stabilized power supply at 230Vac/9Vdc-300mA mains voltage. S’print-BT: On request, portable, serial input, 24 column thermal printer, 58mm paper width. DO9700: zero solution. DO9701: electrolyte solution. Combined dissolved Oxygen/temperature probes DO9709 SS: The kit includes: combined probe for measurement of O2 and tem- perature with replaceable membrane, three membranes, 50ml of zero solu- tion, 50ml of electrolyte solution. Cable length 2m. 12mm x 120mm. DO9709...

 Open the catalog to page 4

All Delta OHM catalogs and technical brochures

  1. HD9120

    4  Pages

  2. HD 32MT.1

    4  Pages

  3. HD2402

    4  Pages

  4. HD2817T...

    5  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.