DPR-2200

DPR-2200

DPR-2200

Product catalog summary
Product Overview
The DPR-2200ALW is a SMART Differential Pressure Transmitter designed for measuring differential pressure, level, and flow. It is suitable for corrosive environments and resistant to chemical attacks. Key features include high accuracy (±0.1%), 4-20mA analogue with digital communications, HART compatibility, and ATEX certification for intrinsic safety and flameproof environments.
Applications
This transmitter is used for measuring pressure differences in gases, vapors, and liquids, particularly where diaphragm seals are needed. Typical applications include hydrostatic level measurement in closed tanks, density, and phase boundary measurements.
Key Features
  • High accuracy of ±0.1%
  • 4-20mA analogue with digital communications
  • HART compatibility
  • Static pressure limit up to 420 bar
  • ATEX certified for intrinsic safety and flameproof environments
  • Fully welded sensor for long-term usage
  • Local configuration of measuring range
Configuration & Communication
The transmitter can be configured using a KAP-03 communicator or a PC with RS-HART converter and Delta D-Soft software. It allows changes in metrological parameters such as pressure units, range start and end points, and time constant.
Measuring Ranges
The nominal measuring range varies from -160 to 160 mbar up to -1.6 to 16 bar, with a static pressure limit of 40 bar. The transmitter can be configured to compensate for ambient temperature effects.
Recommendations
For hydrostatic measurements, it is recommended to use identical diaphragm seals at both upper and lower connection points. Optimal results are achieved using specific diaphragm seals like DN 80, DN 100, A 109, and S-Comp.
Ordering Information
Model codes and descriptions are provided for different configurations, including casing options, output signals, and static pressure limits. Special versions and certificates are available, such as EExia for intrinsic safety and EExd for explosion-proof environments.
Installation & Use
Instructions are provided for configuring the transmitter to measure liquid levels and densities. The process involves setting the transmitter in its working position, making electrical connections, and using HART communication for configuration.
Specifications and Calculations
The document provides a method to calculate the measurement range for a DPR E-2200 transmitter used to measure phase boundary height between liquids of different densities. An example involves calculating the start and end points of the measurement range for liquids with densities of 0.7 g/cm3 and 1.0 g/cm3, with a vertical seal spacing of 1600 mm. The sealing system uses DC-550 oil with a density of 1.068 g/cm3. The start of the measurement range is determined by calculating the pressure difference when the tank is filled with the lighter liquid, resulting in -588.8 mm H2O. The end-point is calculated by adding the pressure increase from a 1-meter column of the heavier liquid, resulting in -288.8 mm H2O.
Additional Remarks
Adjustments to the transmitter settings may be necessary based on laboratory density measurements, especially in pipeline segments with high flow velocity. Increasing the vertical spacing of diaphragm seals can widen the range and improve accuracy. The maximum vertical spacing depends on the transmitter's basic range and the density boundary values of the measured liquid.
Example Calculation
An example is provided for determining the maximum vertical spacing of seals for a DPRE-2200 transmitter measuring liquid density between 0.6 and 1.2 g/cm3, using AK-20 silicone oil with a density of 0.945 g/cm3. The lower boundary of the range is -10 kPa (-1020 mm H2O), and the upper boundary is +10 kPa (1020 mm H2O). The conditions are satisfied when the seal spacing is not more than 2957 mm.
Disclaimer
Delta Controls Ltd reserves the right to amend details without notice and accepts no legal liability for errors or omissions.
See more

Catalog excerpts

DPR-2200-1

Technical Datasheet D-Series SMART Differential Pressure Transmitter with Two Diaphragm Seals Models: DPR-2200 Key Features Ÿ Ÿ 4-20mA analogue with digital communications Ÿ Fully HART ® compatible Ÿ Static pressure limit up to 420 bar Ÿ ATEX certified (Intrinsic Safety, Flameproof) Ÿ Fully welded sensor guarantees tightness of oil systems for long term usage Product Applications The DPR-2200 is suitable for a wide range of applications for measuring: Ÿ Ability to locally configure measuring range Series Overview The D-Series pressure, differential pressure and temperature transmitters offer customers cost-effective and accurate solutions to their individual process requirements. Available with a wide range of process connections and easily configurable via the D-Soft software, the D-Series can be used for a variety of applications where pressure, differential pressure, temperature, level or flow measurements are needed. Other models in this series include: Ÿ Resistant to chemical attack DPT-2000 SMART Temperature Transmitter Ÿ Suitable for use in corrosive atmospheres DPC-2000 SMART Pressure Transmitter DPR-2000G SMART Differential Pressure Transmitter for low ranges The choice of models available DPR-2000 SMART Differential Pressure transmitter Delta Controls offers fast, efficient and knowledgeable support when and where you need it. Please visit our website at www.delta-controls.com to find your local support centre or call us on: +44 (0) 1252 729 140

 Open the catalog to page 1
DPR-2200-2

Level Measurement: Example Pressure Tank Transmitter with two types of diaphragm seal: direct diaphragm seal; remote diaphragm seal Upper remote diaphragm seal Capillary outlet at side the diaphragm seal Capillary fastened to guide Coiled excess of capillary Differential pressure transmitter Lower direct seal The transmitter with a direct diaphragm seal (connected to the positive measurement chamber) and a remote diaphragm seal (connected to the negative chamber) is recommended for hydrostatic measurements of: level, density, phase boundaries and pressure differences (with differentiated height...

 Open the catalog to page 2
DPR-2200-3

Versions Aluminium casing with M20×1.5 packing gland Degree of protection IP 66 S-CompK diaphragm seal M20×1.5 or Æ51 S-DIN K diaphragm seal Æ50, Æ65, Æ80 Transmitter with two types of diaphragm seal: one – direct diaphragm seal and the other – remote diaphragm seal. The example with S-T DN80 diaphragm seal. S-ChK diaphragm seal transmitter with two remote diaphragm seals. Example with S-PK Note: The appropriate configuration of the complete set of pressure transmitter, diaphragm seals and capillaries, as well as the proper selection of manometric fluid depends on several factors such as: the physical...

 Open the catalog to page 3
DPR-2200-4

Application & Construction The differential pressure transmitter is applicable to the measurement of pressure differences of: gas, vapour and liquid, in cases where it is necessary to use seals and the pressure pulse source points may be several metres apart. Typical applications include the hydrostatic measurement of: level in closed tanks; density and phase boundaries; filter loss pressure differences between media in pasteurisers etc. The available range of the diaphragm seals allow measurement at great majority of media. The active element is a piezoresistant silicon sensor separated from...

 Open the catalog to page 4
DPR-2200-5

Electrical Diagrams Transmitters with HART protocol DPR-2200ALW Version: DPR-2200ALE with 0…5 or 0…20mA output signal RS-HART converter or communicator Power supply Power supply SIGNAL Transmitter RS-HART converter or communicator Versions, Certificates∗ ∗ more than one option is ) available Stainless steel housing, IP66/IP67, with display, output 4-20mA + Hart Ex I M1 Ex ia I Ma (steel enclosure only) Ex II 1/2G Ex ia IIC T4/T5 Ga/Gb Ex II 1/2G Ex ia IIB T4/T5 Ga/Gb (version with Teflon shielded cable) Ex II 1D Ex ia IIIC T105°C Da Ex I M2 Ex d ia I Mb (steel enclosure only) Ex II 1/2G Ex ia/d...

 Open the catalog to page 5
DPR-2200-6

To simplify the mathematical operations, we introduce the density coefficient of the medium, Xr . rmedium [g/cm] 3 r water at 4 °C [g/cm 3 ] 3 Since the density of water at 4°C is 1 g/cm , the density coefficient Xr is numerically equal to the density of the medium expressed in g/cm3 . To determine the hydrostatic pressure of a column of liquid in mm H2O, it is sufficient to multiply the height of the column h [mm] by the density coefficient of the liquid Xr. Since it is easy to determine the hydrostatic pressure in mm H2 O and the transmitter can be configured in those units, in the descriptions...

 Open the catalog to page 6
DPR-2200-7

Directions for Use (cont.) Additional remarks 6. Connect the KAP-03 communicator, identify the transmitter and select the “configuration” function. 7. On the configuration menu select “Reranging” proc edure. 8. On the “Reranging” menu: a) change the measurement units to mm H2 O at 4°C; b) enter the calculated values for the start (–1404) and end (396) of the range. The settings of the transmitter can be adjusted with re ference to laboratory results from density measurements carried out on samples of the liquid being measured. This is most often necessary when the measurement takes place in a pipeline...

 Open the catalog to page 7
DPR-2200-8

In the interest of development and improvement Delta Controls Ltd, reserves the right to amend, without notice, details contained in this publication. No legal liability will be accepted by Delta Controls Ltd for any errors, omissions or amendments. Delta Controls Limited Riverside Business Park, Dogflud Way, Farnham, Surrey GU9 7SS, UK. T+44 (0)1252 729 140 F+44 (0)1252 729 168 E [email protected] W www.delta-controls.com

 Open the catalog to page 8
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.