DL2000UB-1V

DL2000UB-1V

DL2000UB-1V

Product catalog summary
Overview: The DL2000UB-1V is a high precision fluxgate AC/DC current transducer designed for galvanically isolated measurement up to 3000 A. It features a 1.5 V output at 3000 A with a BNC connector and a 68 mm aperture. The device boasts a total accuracy of 106 ppm and an offset of 6 ppm, with a status signal and LED indicator.
Specifications: The transducer operates with a nominal primary AC current of 2000 Arms and a DC current range of -3000 to 3000 A. It has a transfer ratio of 2000 A/V and an output resistance of 50 Ω. The linearity error is within ±80 ppm, and the device maintains a bandwidth of 300 kHz. It requires a power supply voltage of ±14.25 to ±15.75 V and consumes ±180 mA at idle.
Applications: Suitable for use in electric vehicle test benches, power measurement and analysis, particle accelerators, MRI devices, battery testing, current calibration, stable power supplies, and precision current sensing.
Isolation and Safety: The device complies with IEC 61010-1 standards, offering reinforced insulation with a clearance and creepage distance of 22 mm. It supports continuous working voltages up to 4000 V for basic insulated wire and transient voltages up to 12500 V for CAT III.
Environmental and Mechanical Characteristics: Designed for indoor use, the transducer operates within a temperature range of -40 to 65°C and a relative humidity of 20-80%. It has an ingress protection rating of IP20 and weighs 5.7 kg.
Installation and Use: The device should be mounted using M5 or M4 screws, depending on the configuration. It requires careful connection of power and measurement cables, with specific instructions to avoid electrical shock and ensure proper operation.
Compliance and Standards: The DL2000UB-1V complies with EC Directives 2014/30/EU and 2014/35/EU, and is manufactured in accordance with the RoHS directive 2011/65/EU. It adheres to standards EN 61010-1, EN 61010-2-030, and EN 61326-1.
See more

Catalog excerpts

DL2000UB-1V-1

High precision fluxgate AC/DC current transducer for galvanically isolated Features • 2000 A rms nominal current • 1.5 V output at 3000 A with BNC connector • Electric vehicle (EV) test bench • Power measurement and power analysis • Particle accelerators • MRI devices and medical scanners • Battery testing and evaluation systems • Current calibration purposes • Stable power supplies • Precision current sensing High precision DC current transducer (DCCT) measuring up to 3000 A currents and continuously measuring 3000 A currents with a linearity error less than 80 ppm. Based on the ultra stable Danisense closed loop flux gate technology, the DL2000UB-1V has very low offset and ultra low drift. With an integrated high precision burden resistor outputting 1.5 V at 3000 A, the DL2000UB-1V makes high current measuring easy. It provides high resolution for precise monitoring, reliable and consistent performance, and a rugged, full aluminium design for durability. All information subject to change without notice

 Open the catalog to page 1
DL2000UB-1V-2

2023-12-19 All information subject to change without notice www.danisense.com

 Open the catalog to page 2
DL2000UB-1V-3

Linearity error Linearity error is defined as the deviation from a straight line. The straight line is a linear regression trend line based on the least squares method of the measurement points from 0 to positive max current and another trendline is calculated from 0 to negative max current. The difference between each measured point and the linear trend line is the linearity error. The linearity error eL can be expressed as (1), where Ireading is the measurement result and Ifitted is the regression value. Reading is the actual value measured at a given time. Full scale is the rated nominal value...

 Open the catalog to page 3
DL2000UB-1V-4

Figure 3: Frequency characteristics 9111100008 rev. 2 All information subject to change without notice www.danisense.com

 Open the catalog to page 4
DL2000UB-1V-5

A When using REINFORCED insulated wire, all wiring must be insulated for the highest voltage used. When using BASIC insulated or uninsulated wire, follow the specified voltages in the table below: Connections: D-sub-9 power supply and BNC voltage reading All information subject to change without notice www.danisense.com

 Open the catalog to page 5
DL2000UB-1V-6

External devices: External devices connected to current transducers must comply with the standards IEC61010-1 and IEC62368-1 and be energy-limited circuitry The transducer should only be cleaned with a damp cloth. No detergent or chemicals should be used. When multiple primary turns are used or high primary currents are applied the temperature around the transducer will increase, please monitor to ensure that the maximum ratings are not exceeded. It is recommended to have minimum 1 November 2012 mm2 Dato: per15.ampere in the primary bus bar. Intended use The DL2000UB-1V is designed to measure...

 Open the catalog to page 6
DL2000UB-1V-7

Positive current direction Positive current direc Figure 4: Dimensions of transducer. 0.3 mm Tolerance Mounting Base plate mounting: Bottom direct mounting (Base plate removed): All information subject to change without notice

 Open the catalog to page 7
DL2000UB-1V-8

Figure 5: D-sub-9 connection pinout Dato: 15. November 2012 Status signal negative terminal 0 V connection for supply voltage Negative supply voltage Status signal positive terminal Positive supply voltage Positive current direction Is identified by an arrow on the back side isolation plastic insert. Figure 6: Status signal optocoupler NOTICE OF PROPRIETARY RIGHTS THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION. IT MAY NOT BE REPRODUCED OR TRANSFERRED TO DOCUMENTS OR DISCLOSED TO OTHERS OR USED FOR MANUFACTURING OR ANYOTHER PURPOSE WITHOUT PRIOR WRITTEN PERMISSION. When the sensor is operating...

 Open the catalog to page 8
DL2000UB-1V-9

Declaration of Conformity Danisense A/S Malervej 10 DK-2630 Taastrup Dato: 15. November 2012 Declares that under our sole responsibility that this product is in conformity with the provisions of the following EC Directives, including all amendments, and with national legislation implementing these directives: Directive 2014/30/EU Directive 2014/35/EU And that the following harmonized standards have been applied EEN 61010-1 (Third Edition):2010, EN 61010-1:2010/A1:2019 EN 61010-2-030:2021/A11:2021 EN 61326-1:2013 All DANISENSE products are manufactured in accordance with RoHS directive 2011/65/EU....

 Open the catalog to page 9

All Danisense catalogs and technical brochures

  1. DS600ID

    9  Pages

  2. DS400ID

    9  Pages

  3. DS300ID

    9  Pages

  4. DS200ID

    9  Pages

  5. DT200ID

    8  Pages

  6. DT100ID

    8  Pages

  7. DT50ID

    8  Pages

  8. DQ640ID-B

    7  Pages

  9. DQ600ID-P1300

    7  Pages

  10. DQ600ID

    9  Pages

  11. DQ500ID

    7  Pages

  12. DL2000ID

    10  Pages

  13. RCM Smart

    10  Pages

  14. RCM Standard

    9  Pages

  15. DSSIU-1

    5  Pages

  16. DSUB2-20

    1  Page

  17. DSSIU-6-1U-V

    8  Pages

  18. DSSIU-4-1U

    6  Pages

  19. DS50ID

    9  Pages

  20. DL2000UB-10V

    9  Pages

  21. DS300UB-1V

    8  Pages

  22. DS300UB-10V

    8  Pages

  23. DN1000ID

    9  Pages

  24. DW500UB-2V

    9  Pages

  25. DM1200ID

    8  Pages

  26. DSSIU-6-1U

    7  Pages

  27. DR10000IM

    9  Pages

  28. DR5000IM

    9  Pages

  29. DP50IP-B

    4  Pages

  30. VOM1333-10

    1  Page

  31. DS400UB-10V

    6  Pages

  32. DC200IF

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.