High precision fluxgate AC/DC current transducer for galvanically isolated Features • 1414 A rms nominal current • 10 V output at 2000 A with BNC connector • Electric vehicle (EV) test bench • Power measurement and power analysis • Particle accelerators • MRI devices and medical scanners • Battery testing and evaluation systems • Current calibration purposes • Stable power supplies • Precision current sensing High precision DC current transducer (DCCT) measuring up to 2200 A currents and continuously measuring 2000 A currents with a linearity error less than 30 ppm. Based on the ultra stable Danisense closed loop flux gate technology, the DL2000UB-10V has very low offset and ultra low drift. With an integrated voltage output module (VOM) outputting 10V at 2000 A, the DL2000UB-10V makes high current measuring easy. It provides high resolution for precise monitoring, reliable and consistent performance, and a rugged, full aluminium design for durability. All information subject to change without notice
Open the catalog to page 12023-12-19 All information subject to change without notice www.danisense.com
Open the catalog to page 2Linearity error Linearity error is defined as the deviation from a straight line. The straight line is a linear regression trend line based on the least squares method of the measurement points from 0 to positive max current and another trendline is calculated from 0 to negative max current. The difference between each measured point and the linear trend line is the linearity error. The linearity error eL can be expressed as (1), where Ireading is the measurement result and Ifitted is the regression value. Reading is the actual value measured at a given time. Full scale is the rated nominal value...
Open the catalog to page 3Primary current (Arms) Figure 2: Maximum continuous primary current vs. frequency Figure 3: Frequency characteristics All information subject to change without notice
Open the catalog to page 4A When using REINFORCED insulated wire, all wiring must be insulated for the highest voltage used. When using BASIC insulated or uninsulated wire, follow the specified voltages in the table below: Connections: D-sub-9 power supply and BNC voltage reading All information subject to change without notice www.danisense.com
Open the catalog to page 5External devices: External devices connected to current transducers must comply with the standards IEC61010-1 and IEC62368-1 and be energy-limited circuitry The transducer should only be cleaned with a damp cloth. No detergent or chemicals should be used. When multiple primary turns are used or high primary currents are applied the temperature around the transducer will increase, please monitor to ensure that the maximum ratings are not exceeded. It is recommended to have minimum 1 November 2012 mm2 Dato: per15.ampere in the primary bus bar. Intended use The DL2000UB-10V is designed to measure...
Open the catalog to page 6Positive current direction Positive current direc Figure 4: Dimensions of transducer. 0.3 mm Tolerance Mounting Base plate mounting: Bottom direct mounting (Base plate removed): All information subject to change without notice
Open the catalog to page 7Figure 5: D-sub-9 connection pinout Dato: 15. November 2012 Status signal negative terminal 0 V connection for supply voltage Negative supply voltage Status signal positive terminal Positive supply voltage Positive current direction Is identified by an arrow on the back side isolation plastic insert. Figure 6: Status signal optocoupler NOTICE OF PROPRIETARY RIGHTS THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION. IT MAY NOT BE REPRODUCED OR TRANSFERRED TO DOCUMENTS OR DISCLOSED TO OTHERS OR USED FOR MANUFACTURING OR ANYOTHER PURPOSE WITHOUT PRIOR WRITTEN PERMISSION. When the sensor is operating...
Open the catalog to page 8Declaration of Conformity Danisense A/S Malervej 10 DK-2630 Taastrup Dato: 15. November 2012 Declares that under our sole responsibility that this product is in conformity with the provisions of the following EC Directives, including all amendments, and with national legislation implementing these directives: Directive 2014/30/EU Directive 2014/35/EU And that the following harmonized standards have been applied EEN 61010-1 (Third Edition):2010, EN 61010-1:2010/A1:2019 EN 61010-2-030:2021/A11:2021 EN 61326-1:2013 All DANISENSE products are manufactured in accordance with RoHS directive 2011/65/EU....
Open the catalog to page 99 Pages
9 Pages
9 Pages
9 Pages
8 Pages
8 Pages
8 Pages
7 Pages
7 Pages
9 Pages
7 Pages
10 Pages
10 Pages
9 Pages
5 Pages
1 Page
8 Pages
6 Pages
9 Pages
9 Pages
8 Pages
8 Pages
7 Pages
9 Pages
9 Pages
8 Pages
7 Pages
9 Pages
9 Pages
9 Pages
24 Pages
4 Pages
1 Page
6 Pages
6 Pages
4 Pages