1. Catalogs
  2. Cubic Instruments (Wuhan) Ltd.
  3. Ruiyi Online Infrared Syngas Analyzer Gasboard 3100

Ruiyi Online Infrared Syngas Analyzer Gasboard 3100

Ruiyi Online Infrared Syngas Analyzer Gasboard 3100

Ruiyi Online Infrared Syngas Analyzer Gasboard 3100

Product catalog summary
Overview: The Gasboard-3100 is an online infrared syngas analyzer designed for continuous measurement of various gas concentrations and heating value calculations. It is suitable for integration into industrial systems, particularly in sectors like iron & steel making, chemical processing, and energy production.
Key Features:
  • International patented technology utilizing non-dispersive infrared (NDIR) and thermal conductivity detector (TCD) technologies.
  • Measures concentrations of CO, CO2, CH4, H2, O2, CnHm, C2H2, C2H4, and calculates heating value and N2 concentration.
  • Modular design with independent sensors for each gas and constant temperature design.
  • Auto-zeroing with air and real-time measurement capabilities.
  • Remote diagnosis and smart communication features.
Specifications:
  • Measurement Ranges: CO, CO2, CH4, H2, O2, CnHm, C2H2, C2H4 with customizable ranges.
  • Accuracy: ≤2% FS for NDIR, ≤3% FS for TCD and ECD.
  • Repeatability: ≤1%.
  • Response Time: ≤10 seconds (NDIR).
  • Operating Conditions: Temperature 0~50°C, Pressure 86 to 108 kPa, Humidity 0-95% non-condensing.
  • Communication Interface: RS232 with real-time monitoring and data download capabilities.
  • Power Supply: 220V/110V±44VAC, 50Hz/60Hz±1Hz.
  • Dimensions: 483mm x 373mm x 140mm, Weight: 12Kg.
Applications: The analyzer is applicable in various processes such as gasification, pyrolysis, waste to energy, plasma gasification, chemical conversion, synthetic ammonia production, hydrogen reforming, and more.
Contact Information: Located in Wuhan, China, with contact details for further inquiries.
See more

Catalog excerpts

Ruiyi Online Infrared Syngas Analyzer Gasboard 3100-1

Gasboard-3100 Online infrared syngas analyzer CO%+CO2%+CH4%+CnHm%+H2%+O2% +C2H2%+C2H4%+Heating Value calculation +N2 calculation Modular Design is a stationary analyzer with standard 19’’ 3U rack enclosure, easy to be integrated inside industrial system. It is designed with international patented technology with non-dispersive infrared (NDIR) and thermal conductivity detector (TCD) technology on the basis of MEMS, able to online continuously measure volume concentration of CO, CO2, CH4, H2 , O2 , CnHm, C2H2 , C2H4 simultaneously, and calculate gas heating value of syngas, is an ideal solution for online gas monitoring in fields of iron & steel making, chemical, coal gasification, biomass gasification, energy-saving monitoring, energy efficiency test, etc. Applications Energy production Gasification /Pyrolysis Waste to Energy Plasma gasification Chemical conversion Synthetic Ammonia/Carbamide Hydrogen reforming Coal to Gasoline/ Methanol/ DME/ Ethylene Glycol… Calcium carbide furnace tail gas Steel making Blast furnace Converter Coking Direct iron ore smelting reduction processes Heat treatment Endogas/Exogas www.gas-analyzers.com www.gasanalyzer.c

 Open the catalog to page 1
Ruiyi Online Infrared Syngas Analyzer Gasboard 3100-2

General Features International Patent 8 Gas Measurement Modular Design Independent Sensor for Each Gas Constant Temperature Design Auto-Zeroing with Air Precise CH4 measurement Reliable CnHm, C2H2, C2H4 reading Linear Correction on TCD Sensor Online Real Time Measurement Easy Operation, No Consumables Remote Diagnosis Modular Design & Smart Communication www.gas-analyzers.com www.gasanalyzer.com.cn

 Open the catalog to page 2
Ruiyi Online Infrared Syngas Analyzer Gasboard 3100-3

Specifications Items High Heating Value or Low Heating Value in MJ/m³ or kcal/m³ N2 calculation by balance (Optional) CO, CO2, CH4, CnHm, C2H2, C2H4 : proprietary dual beam NDIR detectors O2 : industrial use long lifetime electrochemical cell (ECD) H2 : proprietary thermal conductivity detector (TCD) Auto-zeroing function with fresh air Gas conditions at analyzer inlet Operating conditions Tamb : 0~50°C / Pamb : 86 to 108 kPa / RH : 0-95% non condensing Warm-up time 800 seconds (auto-zeroing cycle during the last 120 seconds) Communication interface RS232 ( real time monitor and memory data download...