VOC emission test chamber EK1000 Highly accurate and affordable VOC chamber testing in the automotive industry
Open the catalog to page 1Olfasense and CTS – A joint venture that makes sense Dear Madam, dear Sir, Thank you very much for your interest in our EK1000 VOC emission test chamber. Olfasense is the world-leading supplier of odour laboratory technology, which is used by more than 400 laboratories, research institutes, universities and companies worldwide. In our daily work, Olfasense also offers specialist odour consultancy services to evaluate products and materials. Besides using sensory evaluation techniques, Olfasense helps clients to understand the chemical composition of air samples in terms of specific VOCs....
Open the catalog to page 2VOC EMISSION TEST CHAMBER EK1000 FULLY COMPLIANT WITH : ISO 12219-4 (Determination of VOCs from vehicle interior parts and materials ) VDA 276 (Determination of Organic substances as emitted from automotive interior products using a 1 m3 test chamber) GS97014-3 BMW Method (Determination of VOCs from components, semifinished products and materials) Highly accurate and affordable VOC chamber testing in the automotive industry PN 780 Porsche Method (Interieur – Emission behavior) The Olfasense VOC emission test chamber EK1000 allows characterization of the emissions of volatile organic...
Open the catalog to page 3VOC EMISSION TEST CHAMBER EK1000 Standard features » External dimensions: 2790 x 1263 x 1980 mm (LWH) » Temperature range: room temperature plus 5° C to 210° C inside the VOC emission test chamber » Adjustable fan speed » Automated emergency stop by temperature selection limiter class 2 (DIN 12880) » Left hinged door » Electro certificate of acceptance If you have individual requirements, please ask for your tailormade product solution.
Open the catalog to page 4VOC EMISSION TEST CHAMBER EK1000 Emission test space » VOC emission test chamber made of polished stainless steel, usable space: 1735 x 745 x 745 mm (LWH) » Precise compliance with all method and measurement parameters » Minimized adhesion » User-friendly pressure chamber door locks » Rotor with drive via non-contact magnetic coupling and emission-free ceramic bearings (speed adjustable) - patent pending » Air switch: e.g. from regulated and humidified 6.67l/min to dry 50l/
Open the catalog to page 5VOC EMISSION TEST CHAMBER EK1000 Standard features » Setting, control and monitoring of parameters: temperature, humidity, mass flow » Integration of additional data channels (e.g. FID) » Multi-chamber edition » Interfaces: USB, RS232, RS422, RS485 (usable e.g. for FID data connection); an FID is not included » Multi-language e.g. German, English, Spanish Data logger softwa
Open the catalog to page 6VOC EMISSION TEST CHAMBER EK1000 Sampling connections » Sampling connections on the right » Process-monitoring sensor (temperature and humidity in the chamber) » Functional reliability, mass flow control at chamber inlet » Heated sampling port, 4 ports (e.g. Swagelok, others on request) If you have individual requirements, please ask for your tailormade product solution
Open the catalog to page 7VOC EMISSION TEST CHAMBER EK1000 Standard features » Easy to maintain » Exchangeable by plug and play technology » Can be exchanged by the client within a few minutes » Comes in a robust transport box for easy maintenance
Open the catalog to page 8VOC EMISSION TEST CHAMBER EK1000 Other standard features » Viton door seal, thermally loadable up to 160° C; easy to replace » Lab water container » Separate connections for compressed air and for purging air Optional features » Lab water connection » Fogging-device (without cooling thermostat device) VDA 276, VCS 1027, 2769 » Nose mask mounted directly on the test chamber VCS 1027, 2769 » Tedlar bag for volume compensation - GS-97014-2 » Air cooled refrigerant compressor - GS-9
Open the catalog to page 9VOC EMISSION TEST CHAMBER EK1000 Excerpt of standard compliance test Background concentration Olfasense has conducted validation tests for the VOC emission test chamber EK1000, including, but not limited to spatial temperature distribution, leak tightness, intermixing and background concentration. The chamber showed compliance with the normative requirements. Additional documentation is available on request. Analysis of TVOC and single target VOC on background concentration, supplied air quality and the water used for humidification showed compliance with the normative requirements. Leak...
Open the catalog to page 10VOC EMISSION TEST CHAMBER EK1000 Technical data Description Spatial temperature distribution Emission test space volume (usable) To test the spatial temperature distribution, eight PT100 temperature sensors were positioned in a wire frame in the middle of the emission test chamber. The uniformity of the distribution was better than the resolution of the temperature sensors (±0.2 K). Air jacket temperature conditioning Temperature range Temperature deviation over time Temperature homogeneity spatially Humidity range Deviation in humidity over time 1 to 12 l/min (fixed purging air) approx....
Open the catalog to page 11Olfasense GmbH Fraunhoferstr.13 24118 Kiel Germany Tel.: +49 (0)431-22012-0 Fax: +49 (0)431-22012-17 sales@olfasense.com
Open the catalog to page 12All CTS Clima Temperatur Systeme GmbH catalogs and technical brochures
-
G1 Control
6 Pages
-
CDI V4.02
174 Pages
-
CID V300
88 Pages
-
CID V2.01
88 Pages
-
Environmental Simulation
7 Pages
-
CTS Climatic Test Cabinets
2 Pages
-
CTS Shocktest cabinet
2 Pages
-
Customised Solutions
8 Pages