G1 Control
6Pages

{{requestButtons}}

Catalog excerpts

G1 Control - 2

Comments: - The timebase of the programme is hours. This is because in the example only programmesteps with whole hours are used. - Always set a starting value for the programme (first line with time t=0). In other case the controller starts the programme with the last used setvalues - We suggest to create the testcycles via CID-Software, because programme generation with CID is very easy. The programme should be saved as Prog.No. 12. It’s not neccessary to switch off the chamber to edit a new programme +/- Temp if there is a programme saved under Prog.No 12, the number of used Lines of...

Open the catalog to page 2
G1 Control - 3

+/- Humidity Set startvalue for humidity H= 60 % r.H. P time of the first line is 0, therefore continue with S instead of +/- Temp to change time Finish first line. Last value pair of temp./hum. is used as default for the next line. Because we want to use this values continue with S rather than +/- Temp to set new temp. value Set duration of 1. Line stay at +20 °C / 60 % r.H. for one hour Last valuepair is used as default Start of inactive programme line 1 2 +/- Humidity Number of loops to do End of inactive line Last valuepair is used as default 80.0 60.0 +/- Humidity Put in new Humidity...

Open the catalog to page 3
G1 Control - 5

Use last value pair as default Start of inactive programme line Set WAIT-Function on temperature channel Hold 0 lo 0.0 +/- Humidity Set Tolerance for Wait-function to ± 1 Kelvin End of inactive line Hold 0 lo 1.0 S P Use last value pair as default Use last value pair as default Set duration of line 8 Stay at -40 °C for 2 hours Duration of line 9 Change in 1 hour to +20 °C Use last value pair as default

Open the catalog to page 5
G1 Control - 6

+/- Humidity Switch on Humidity. New value is 60%r.H. P Duration of line 10 is 0 hours Use last value pair as default Start of inactive programme line Set end of loop (SE) End of inactive line Use last value pair as default Duration of line 12 Stay at +20 °C for 1 hour Finish and save last programme line Save the whole programme

Open the catalog to page 6

All CTS Clima Temperatur Systeme GmbH catalogs and technical brochures

  1. CDI V4.02

    174 Pages

  2. CID V300

    88 Pages

  3. CID V2.01

    88 Pages