SYSTEM Prep and Load Platform Specifications for PAL Systems COMBI | GC | LC HTC | HTS | LHX | HTX Not Valid for PAL-xt Systems PAL is a registered trademark of CTC Analytics AG | Switzerland

Open the catalog to page 1
PAL Systems Specifications - 2

Gas Chromatographic Test: High Split Ratio | Retention Gap | Solvent iso-Octane 4 Gas Chromatographic Test: Split Mode | Analytical Column | Alkene Test Mix 5 Gas Chromatographic Headspace Test: High Split Ratio | Retention Gap | Solvent iso-Octane 6 HPLC Chromatographic Test: Loop Overfill and Partial Loop Filling 7 Formulas used to calculate the Specification Values 8

Open the catalog to page 2
PAL Systems Specifications - 3

Syringe Volume Specification Test Conditions 10 pi Repeatability 15 times 8 pi Water transferred into Relative Standard Deviation target vial. s-rel :_0.20 % 100 pL Repeatability 100 pL Water transferred in a target vial. Relative Standard Deviation s-rel :_0.15 % Linearity 6 levels and 3 replicates of each level. R: >0.9999 1000 pL Repeatability 500 pL Water transferred in a target vial. Relative Standard Deviation s-rel : 0.10 % Linearity 6 levels and 3 replicates of each level. R: > 0.9999 2500 pL Repeatability 1000 pL Water transferred in a target...

Open the catalog to page 3
PAL Systems Specifications - 4

Specification: Repeatability: Relative Standard Deviation s-rel: 0.30 % Carry-over: not applicable General Remarks: Number of specimens for Repeatability Tests: n = 20 Carry-over: Test: Remaining solvent in syringe can not be eliminated from injection to injection. Test Conditions: Test Sample: iso-Octane puriss., 2 mL vials filled to 50 % of nominal content Syringe: Hamilton Art. No.: 203205 | CTC Art. No.: SyrC L10-26S-AS PAL Parameter Settings: CYCLE GC-Inj Fill Strokes 5 SYRINGE10 |jlPullup Del5 s Sample Volume 1.0 p.L Inject to GC Inj1 Fill Speed 2 |jl/s...

Open the catalog to page 4
PAL Systems Specifications - 5

Specification: Repeatability: Relative Standard Deviation s-rel: 0.80 % General Remarks: Number of specimens for Repeatability Tests: n = 8 Repeatability determination based on C14 Peak Carry-over Test: Peak < 1 area count, not detectable. Test Conditions: Test Sample: Alkane Test Mix C14,G5,G6 | 0.3 p.g dissolved in iso-Octane. 2 mL vials filled to 50 % of nominal content Syringe: Hamilton PNo.: 203205 | CTC Analytics PNo.: SyrC L10-26S-AS PAL Parameter Settings: CYCLE GC-Inj Fill Strokes 5 Sample Volume 1.0 |jl Inject to GC Inj1 Air Volume 0.5 |jl Inject...

Open the catalog to page 5
PAL Systems Specifications - 6

Specification: Repeatability: Relative Standard Deviation s-rel: < 1.00 % General Remarks: Number of specimens for Repeatability Tests: n = 7 10 |jl are measured by using a glass capillary, Drummond Microcaps (Supplier Supelco, PNo. P-1924) Test Conditions: Test Sample: Iso-Octane, puriss., Microcaps Glass Capillary hold with Tweezers, filled and put into 20 mL Headspace vial, sealed immediately. Syringe: Hamilton PNo.: 203084 | CTC Analytics PNo.: SyrC HS 2.5 -23-5 GC Agilent 6890: Injector: split/splitless, Liner: Single Taper 4 mm ID, Needle Penetration 40 mm Silanized glass wool...

Open the catalog to page 6
PAL Systems Specifications - 7

Specification: Syringe VolumeSpecificationTest Conditions 100 pL Repeatability Loop Overfill Loop size 20 pi Relative Standard Deviation Loop fill volume: 80 pL 100 pL Repeatability Partial Loop Filling Loop size 20 pL Relative Standard Deviation Loop fill volume: 10 pL s-rel :_0.20 % 100 pL Linearity 5 levels and 3 replicates of each level. General Remarks: Number of specimens for Repeatability Tests: n = 9 Carry-over: Test: Peak < 1 area count, not detectable. Test Conditions: Test Sample: Benzophenone, Supplier Fluka PNo. 12750, 50 pg/mL dissolved...

Open the catalog to page 7
PAL Systems Specifications - 8

Mean Value (average). Average of measured area counts. x = Mean Value of area counts. Xt = Peak area counts, i th observation of n n = Number of injections, calibration points. Standard Deviation O n - 1 = Standard Deviation Relative Standard Deviation x RSD = Relative Standard Deviation

Open the catalog to page 8