Data sheet DS 400 dew point monitoring
6Pages

{{requestButtons}}

Catalog excerpts

Data sheet DS 400 dew point monitoring - 1

Dew point monitoring DS 400 for stationary dew point monitoring of refrigeration or desiccant driers. The touch screen graphic display enables an intuitive operation and shows the progress of the measured values. 2 alarm relays are available for monitoring of threshold values. Available either with a classic analogue output 4...20 mA or optionally with digital interfaces like Ethernet and RS 485 (Modbus protocol). As a stand-alone solution the measured data stored in the optional data logger can be read-out via USB stick and evaluated by means of the software CS Soft Basic. Dew point monitoring DS 400 consisting of: Special features: • 3.5“ graphic display - easy operation with touch screen • System ready for plug-in: Everything completely wired • 2 alarm contacts (230 VAC, 3 A) pre-and main alarm freely adjustable • NEW: An alarm delay can be set for each alarm relay • 4...20 mA analogue output • Option: Ethernet and RS 485 interface (Modbus protocol) • Option: webserver - Option alarm unit (buzzer and continuous red light) - Standard measuring chamber - 2nd sensor input for dew point or RS 485 interface flow sensor Option: Integrated data logger • Recording of the dew point progression of up to 100 million measuring values • CS Soft Basic for evaluation in graphic and table form. Read-out of the data either via USB stick or via Ethernet Technical data DS 400 Dimensions: 118 x 115 x 98 mm IP 54 (wall housing) 92 x 92 x 75 mm (panel mounting) Inputs: 2 digital inputs for FA sensors Interface: USB Alarm outputs: 2 relays, (pot.-free) Options: Data logger: 100 million measuring values start/stop time, measuring rate freely adjustable 2 additional sen- for connection of sor inputs: pressure sensors, temperature sensors, clamp-on ammeters, third-party sensors with 4...20 mA 0 to 10 V, Pt 100, Pt 1000 Technical data FA 510 Measuring range: -80...20 °Ctd resp. -20...50 °Ctd Accuracy: ± 1 °C at 20...-20 °Ctd ± 2 °C at -20...-50 °Ctd ± 3 °C at -50...-80 °Ctd Pressure range: -1...50 bar, special ver

Open the catalog to page 1
Data sheet DS 400 dew point monitoring - 2

Easy operation via touch screen Actual measured values All measured values can be seen at a glance. Threshold exceeding are indicated in red color. A ..measuring site name" can be allocated to each sensor. Graphic view In the graphic view all measured values are indicated as curves. It is possible to brows back on the time axis by a slide of the finger (without data logger maximum 24 h, with data logger back to the start of the measurement). Data logger Measured values are stored in DS 400 by means of the option .integrated data logger". The time interval can be freely set. Furthermore...

Open the catalog to page 2
Data sheet DS 400 dew point monitoring - 3

DS 400 - Chart recorder for all relevant parameters of compressed air Software options: • Integrated Webserver • Mathematics calculation function • Totalizer function Hardware options: • Integrated data logger • Ethernet / RS 485 interface • additional sensor inputs (digital or analogue) selectable The 2 sensor inputs board 1 and 2 can be selected according to the required sensors: Standard equipment: • USB interface • 3.5“ graphic display with touch screen • Integrated mains unit for supply of the sensors • 4...20 mA output of all connected active sensors • Pulse output (for...

Open the catalog to page 3
Data sheet DS 400 dew point monitoring - 4

VA 500 flow sensor in basic version: Standard (92.7 m/s), sensor length 220 mm, without display Options for VA 500: (see page 81) Flow meter VA 520 with integrated measuring section, (R 1/4“ DN 8) Flow meter VA 520 with integrated measuring section, (R 1/2“ DN 15) Flow meter VA 520 with integrated measuring section, (R 3/4“ DN 20) Flow meter VA 520 with integrated measuring section, (R 1“ DN 25) Flow meter VA 520 with integrated measuring section, (R 1 1/4“ DN 32) Flow meter VA 520 with integrated measuring section, (R...

Open the catalog to page 4
Data sheet DS 400 dew point monitoring - 5

CS Soft Basic - evaluation of measured data for single computers Ethernet The measured data stored in the data logger integrated in DS 400 can be read-out via USB stick. If DS 400 has the optional Ethernet interface the measured data can also be read-out over big distances via the computer network Graphic evaluation All measurement curves are indicated in different colors. All necessary functions like free zoom, selection/deselection of single measured curves, free selection of time periods, scaling of the axis, selection of colors and so on are integrated: This view can be stored as a pdf...

Open the catalog to page 5
Data sheet DS 400 dew point monitoring - 6

By means of the CS Soft Network an optional number of DS 500/ DS 400 instruments can be evaluated via Ethernet. The software stores the measured data of all DS 500 / DS 400 cyclically (cycle freely selectable) in a SQL database on the server. In case of an exceeding of the stored alarm values the software automatically sends an SMS or an e-mail. Furthermore, different user levels can be defined in the server software so that single staff members only can access the measured data of certain DS 500 / DS 400. The evaluation of the measured data can be carried out by means of the client...

Open the catalog to page 6