Introduction
The CRY342 is a pre-polarized 1/4 inch free field condenser microphone designed for high-level and high-frequency acoustic measurements. It is capable of measuring sound pressure levels up to 168dB and is suitable for use in harsh environments due to its titanium membrane.

Compatibility
The microphone is compatible with CRY547 and CRY543 preamplifiers and can also be used with other brands such as B&K and G.R.A.S.

Applications
It is applicable for use with acoustic testers, multi-channel noise analyzers, and other related instruments.

Standards Compliance
The CRY342 meets the IEC 61094-4:1995 and GBT 20441.4-2006 standards for working standard microphones.

Technical Specifications
  • Type: Pressure field
  • Sensitivity: 1.6mV/Pa (-56dB) ±2dB
  • Frequency Response: 4Hz-40kHz ±2dB, 10Hz-10kHz ±1dB
  • Polarity Voltage: 0V
  • Linearity Range: 40-160dB (250Hz sensitivity changes 0.2dB/10dB)
  • Dynamic Range Limit: ≥168dB (250Hz THD<3%)
  • Equivalent Noise: ≤30dB (250Hz)
  • Working Temperature: -20℃ to 60℃
  • Temperature Coefficient: -0.03dB/℃ (-10℃ to 50℃)
  • Static Pressure Coefficient: ±0.04dB/kPa
  • Relative Humidity Range: 10-90% no condensation
  • Relative Humidity Coefficient: <0.05dB (10-90% no condensation)
  • Long Period Stability: <0.03dB/a (15℃ to 25℃ at 250Hz)
  • Short Period Stability: <0.03dB (15℃ to 25℃ at 250Hz)
  • IEC 61094-4 Name: WS3P

Conclusion
The CRY342 microphone is a robust and versatile tool for precise acoustic measurements, meeting international standards and offering compatibility with various preamplifiers and measurement systems.
See more

Catalog excerpts

CRY342 Measurement Microphone-1

CRY342 Measurement Microphone CRY342 measurement microphone is a pre-polarized 1/4 inch free field condenser microphone. It mainly used for high-level and high-frequency measurements since the sensitivity is low. It can measure high sound pressure levels to 168dB. The Titanium membrane can achieve high stability and accuracy in harsh environment. It can be compatible with CRY547 and CRY543 preamplifier, and it also can be compatible with other brands like B&K, G.R.A.S, etc. It is applicable to the acoustic tester, multi-channel noise analyzer and other instruments. Meet IEC 61094-4:1995 Measurement microphones―Part 4: Specifications for working standard microphones Meet GBT 20441.4-2006 Measurement microphones―Part 4: Specifications for working standard microphones Technique Specification Specification Type Field Type Pressure field Sensitivity mV/Pa (dBV/Pa) Frequency Response(dB) Polarity Voltage(V) Dynamic range limit(ref. 20uPa) ≥168dB (250Hz THD<3%) Equivalent noise(ref. 20uPa) Working Temperature Temperature Coefficient Static Pressure Coefficient Relative Humidity Range Relative Humidity Coefficient Long Period Stability Short Period Stability Add: No.31-1, Xianxing Road, Xianlin Industrial Park, Yuhang District, Hangzhou, Zhejiang

 Open the catalog to page 1
CRY342 Measurement Microphone-2

CRY342 Pressure field measurement microphone typical response curve Add: No.31-1, Xianxing Road, Xianlin Industrial Park, Yuhang District, Hangzhou, Zhejiang

 Open the catalog to page 2

All CRYSOUND catalogs and technical brochures

  1. CRY721

    2  Pages

  2. CRY720

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.