Overview: The pressure transmitter with a ceramic measuring membrane is designed for universal use, offering high chemical resistance, stability, and a favorable price. It is suitable for measuring both positive and negative pressures in liquids and gases, including those that are partially chemically aggressive. The transmitter is applicable in various fields such as industrial automation, power measurements, heating technology, and agriculture.
Specifications:
  • Nominal pressure range: 0 to 10 kPa up to 0 to 60 MPa
  • Overpressure capacity: 200% (400%) of nominal range
  • Error margin: Maximum 1% (0.5%)
  • Temperature error: Zero temperature error typical 0.1%, maximum 0.3% per 10°C; Span temperature error typical 0.1%, maximum 0.3% per 10°C
  • Compensated temperature range: 0 to +85°C
  • Operating temperature range: -15 to +85°C (non-condensing)
  • Storage temperature: -25 to 100°C
  • Supply voltage: 12 to 36 V DC
  • Output options: 4 to 20 mA two-wire, 0 to 20 mA three-wire, 0 to 10 V three-wire
  • Protection: Minimum IP 54
  • Weight: Approximately 200g
  • EMC compliance: ČSN EN 61326-1
Design and Features: The transducer case is made from stainless steel, providing high chemical resistance. It features a connecting thread dimension of G1/2” or M20x1.5, with a special design for media with high viscosity and contamination. The piezoelectric sensors on a ceramic membrane are used as the pressure-sensing element, with a measuring system created by thick-film technology. The transmitter includes a protection circuit against electromagnetic and electrostatic interference.
Instructions for Use: Before connecting the transducer, verify that the pressure corresponds to the nominal range. The device requires no maintenance during operation, and repairs are performed by the manufacturer. It is recommended to connect the earthing pin to a fixed potential to ensure immunity against electromagnetic interference.
Ordering Information: The ordering code includes specifications for pressure measurement type, pressure range, membrane material, case material, electrical connection, and output type. The basic version includes relative overpressure, stainless case with thread G1/2“, connector according to DIN 43650, and output 4 to 20 mA.
See more

Catalog excerpts

Cressto A series - datasheet - pressure transmitter with ceramic membrane-1

pressure transmitter with ceramic measuring membrane The transducers series A are designed for a general use. They are capable to measure both the positive and negative pressure in the liquids and gases that can be partially chemically aggressive. These transducers series A will find a wide use in a full scale of the industrial automation, power measurements, heating technology, agricultural applications. The transducer case is made from stainless steel with high chemical resistance. Connecting thread proper has a dimension G1/2” or M20x1.5. Special design connection with open membrane is suitable for media with high viscosity and contamination. This solution has advantage in cleaning this membrane from residues of media. The second advantage is faster dynamic response and it’s recommended for measuring fast pressure shocks. As a pressure-sensing element in this series of transducers, there are used piezoelectric sensors on ceramic membrane. Measuring system is created by thickfilm technology and is adjusted during production by laser beam. With proper supply is output signal equal to input pressure. This sensor is sealed in stainless steel by “O” ring from Viton. high chemical resistance simple application universal usage high stability favourable price Small signal from sensor is amplified, compensated and calibrated by electronics, which is based inside the transmitter case. This transmitters are produced with twowire current output 4 ÷ 20 mA, three-wire current output 0 ÷ 20 mA and three-wire voltage output 0 ÷ 10 V. Nominal supply voltage is 24 V DC, but they can operate from 12 to 36 V DC without influence on accuracy. The transmitter has a protection from changing polarity of supply voltage. Calibration is provided through lasertrimmed-thick-film resistors and for slight change of transfer characteristics endpoints are used two small multi-rotation trim pots under connector cover. Correction range is ±2 % form nominal range, approximately. Trim pot for zero-correction is signed by red colour. The transducer design uses new modern elements, surface mount technology, etc. A special care is taken of the resistance against the outer electromagnetic and electrostatic interference. On bottom side of connector base is placed special filter and protection circuit. The case body is not electrically connected with the system but it is connected to the earthing pin of the connector. We recommend connection of this pin always to the fixed potential, especially with a view to achieving satisfactory immunity against the outer electromagnetic interference. Electric connection is realized by means of the sealed arrested connector (ISO 4400/6952 – DIN 43650) with a cable bushing PG9, which enables to connect the transducer by means of the cable with a diameter 6 – 9 mm. If agreed, it is possible to set different input pressure values and /or output electrical values, namely for the

 Open the catalog to page 1
Cressto A series - datasheet - pressure transmitter with ceramic membrane-2

ORDERING CODE TECHNICAL INFORMATION Nominal pressure range At the customer's wish it is possible to secure a metrological verification of the transmitters at an accredited Calibration Service Centre. type of press. measure. relative overpressure relative underpressure V exponent of pressure range multiplicand of pressure range membrane material case material stainless, G1/2'', open membrane B electrical connection • Before connection of the transducer into the pressure circuit, it is necessary to verify that the pressure being measured corresponds to the nominal range of this transducer. Even...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.