149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet Bently Nevada Machinery Condition Monitoring Description The 149744 Trendmaster Dynamic Scanning Module (DSM) is a compact rack-based data acquisition system that is fully integrated with System 1 software. The DSM rack has a total of 5 card slots. The first slot is dedicated for communications and will accept either the copper or fiber Ethernet card. The second slot is reserved. The 3rd to 5th slots are TIM Input card or reserved. The TIM input card connects to Bently Nevada TIM and proTIM modules. Each TIM input card provides 2 TIM lines, and each TIM line supports up to 254 channels. Input cards offer highresolution sampling with onboard real-time processing. Onboard processing is the key to the powerful and efficient features available with the DSM platform. Because each input card can process data locally, the DSM can return postprocessed variables to the host computer and reduce the required network bandwidth. If the host computer requires raw data, the DSM can also return waveforms and spectrums. Modbus Communications Capability The introduction of a Modbus digital interface now permits DSMs to communicate directly with process control and automation systems without the need for additional hardware. This capability provides a low-cost entry-level alternative to System 1 that uses the basic trending and alarming functionality that is integral to existing process control systems. All DSMs now include Modbus over TCP/IP capability and require only the DSM Modbus Exporter software to configure all the DSM inputs and define the Modbus interface. The DSM requires the Modbus Serial to Ethernet Bridge for RS232/485 Modbus communication. See the accessory section of this datasheet for the Modbus Exporter software and Serial to Ethernet Bridge part

Open the catalog to page 1
149744 Trendmaster Dynamic Scanning Module (DSM) - 2

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet DSM Features l Fully integrated with System 1 and Decision Support l Ethernet Modbus server with or without System 1 l Up to 150 DSMs per single data acquisition computer l Small package, 21cm x 13 cm x 11cm (8.3 in x 5.1 in x 4.3 in) l Choice of copper or fiber Ethernet l Synchronous and asynchronous processing l Automatic self-checking for DSM and Input cards, and transducers l Up to 24 kHz high bandwidth inputs l Up to 16-bit high resolution sampling l Up to 16x auto gain l TIM (Transducer Interface Module) support for: -...

Open the catalog to page 2
149744 Trendmaster Dynamic Scanning Module (DSM) - 3

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet 149831 Rev. AC Operation Protocol Baud Rate DSM Rack (149744) Power connector located on Input Voltage communications card. 20 to 30 Vdc Input Power Input Lines Tri-color LED indicates status of DSM and input modules with combinations of colors and flash rates Both lines sampled simultaneously 2 lines per card 127 pro-TIMs per line Supported TIMs 10/100 TX Copper Ethernet Card (149776-01) Status LED 400 meters (1312 feet) multimode fiber optic cable (half duplex) 2000 meters (6562 feet) multimode fiber optic cable (full duplex) Fuse...

Open the catalog to page 3
149744 Trendmaster Dynamic Scanning Module (DSM) - 4

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet Asynchronous Waveforms (Software configurable) Frequency Spans Weight DSM with no input cards Mounting DIN rail option 35mm DIN rail. Requires 26.7 cm (10.5 in) rail length. Weatherproof 35mm DIN rail. Requires 26.7 cm housing (10.5in.) rail length. Bulkhead option Requires 4 #8 screws required. Environmental Limits Operating Temperature 95%, non-condensing 100% condensing when installed in weatherproof housing with power applied. Operating or Storage Humidity None, flat-top, or hanning Storage Temperature Spectrum averages Input...

Open the catalog to page 4
149744 Trendmaster Dynamic Scanning Module (DSM) - 5

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet CSA/NRTL/C (Approval Options 05) Ex ia IIC T4 Ga Installed with intrinsically safe zener barriers per drawing 112M7732 FCC This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: Class I Zone 0: AEx ia IIC T4 Ga Class I, Div 1 Groups A, B, C & D Class II, Groups E, F & G Class III T4 @ -40° C[ Ta [+ 100° C (-40° F [ Ta [+212° F) Ex nA IIC T4 Gc Installed without Class I Zone 2: AEx nA IIC T4 Gc barriers per Class I, Div 2 Groups A, B, C & D drawing 112M7732 T4 @ -40° C [ Ta[+ 100° C...

Open the catalog to page 5
149744 Trendmaster Dynamic Scanning Module (DSM) - 6

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet Ordering Information Accessories Use the part numbers listed in this section to order spare parts or additional components for your Trendmaster DSM system. For the detailed listing of country and product specific approvals, refer to the Approvals Quick Reference Guide (108M1756) available from Bently.com. DSM Modbus Exporter Software Spare 10/100 Base T Ethernet Communication Card. Spare Fiber Optic Ethernet Communication Card. Spare TIM Line Input Card. Blank Slot Cover. Power Supply to DSM Wiring Harness. Weatherproof Housing. TIM...

Open the catalog to page 6
149744 Trendmaster Dynamic Scanning Module (DSM) - 7

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet Graphs and Figures Note: All dimensions shown in millimetres (inches) except as noted. Figure 1: Weatherproof Housing Dimensions

Open the catalog to page 7
149744 Trendmaster Dynamic Scanning Module (DSM) - 8

149744 Trendmaster Dynamic Scanning Module (DSM) Datasheet Copyright 2020 Baker Hughes Company. All rights reserved. Bently Nevada, Orbit Logo and Trendmaster are registered trademarks of Bently Nevada, a Baker Hughes Business, in the United States and other countries. The Baker Hughes logo is a trademark of Baker Hughes Company. All other product and company names are trademarks of their respective holders. Use of the trademarks does not imply any affiliation with or endorsement by the respective holders. Baker Hughes provides this information on an “as is” basis for general information...