Personnel in chemical plants use personal gas detectors to protect against gas exposure. These detectors typically display "0" either because the concentration is zero or the instrument is defective. A daily check is mandatory to ensure functionality, as full calibration is only required every four months.

Daily Functional Test
The daily test involves exposing the sensor to a short puff of gas to trigger an alarm, ensuring the detector is not defective. The test checks for defective electronics, alarm buzzer or lamp, sensor defects, slow response, or insufficient sensitivity. The gas generator stops after 10 seconds, and failure to alarm within this time indicates a problem.

New Gas Generator Features
Compur Monitors introduces a new gas generator model with features including:
  • Flow adjusted to simulate gas detector operating conditions.
  • Electrochemical cell with electrolyte paste that generates gas only when activated.
  • Small gas production, allowing unrestricted transportation and storage.
  • Automatic gas production stop after 10 seconds or detector removal.
The generator is compatible with Monitox, Monitox plus Minitox, Dositox, and Tracer models, replacing the model 4100.

Usage and Safety
The gas detector must be placed in the red receptacle, with red and green LEDs indicating operation status. Compur Monitors emphasizes the importance of testing products for suitability and disclaims liability for product use.

Contact Information
Compur Monitors GmbH & Co. KG
Weißenseestraße 101, D-81539 München
Tel.: ++49 89 62038-0
Fax: ++49 89 62038-184
Email: [email protected]
Website: www.compur.com

Disclaimer
Specifications are subject to change, and users must test products for suitability. Compur Monitors disclaims liability for product use and does not authorize any statements not contained in the document.
See more

Catalog excerpts

Gas Generator-1

New Gas Generator to check personal Gas Detectors A 100% Test in just 10 Seconds Personnel working in chemical plants protect themselves with personal gas detectors from gas exposure. Gas leaks are very unlikely to occur, so the gas detector will always display „0“. An instrument displaying zero can do so for two reasons: a) The concentration is zero. b) The instrument is defective. For good reasons a daily check of the instrument is mandatory. It would be too time consuming to perform a complete calibration every day. Regulations allow calibration intervals of up to four months. The daily functional test with the gas generator makes sure that no instrument failure caused by abuse or wear and tear will go unnoticed. To perform a test it is sufficient to expose the sensor to a short puff of gas and see if the instrument gives an alarm. To make this test as fast, easy and cheap as possible, no accurate gas concentration is used, but approximately twice the value of the alarm threshold. The quality criterion for the detector is not reaching a certain value, but the response time: A good sensor is also a fast sensor. So the gas generator stops gas generation after 10 seconds automatically. If the detector fails to alarm within this time interval it is not safe to be used. Thus the generator test reliably indicates the following potential errors: - Defective electronics Alarm buzzer or lamp defective Sensor defective Sensor too slow Sensor not sensitive enoug

 Open the catalog to page 1
Gas Generator-2

Compur Monitors introduces a new gas generator model with the following features: The flow of the built- in pump is adjusted to perfectly simulate the operating conditions of the gas detector. The gas generator itself consists of an electrochemical cell filled with a electrolyte paste. This cell will only generate gas when it is activated by putting a gas detector on the receptacle. The amount of gas produced when activated is very small. Therefore no restrictions regarding transportation and storage of the gas generator apply. The gas generator is very easy to use. The gas detector operates...

 Open the catalog to page 2

All COMPUR MONITORS catalogs and technical brochures

  1. Statox 560

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.