TGM226 Automatic Moisture Analyzer

TGM226 Automatic Moisture Analyzer

TGM226 Automatic Moisture Analyzer

Product catalog summary
Working Principle
The TGM226 Automatic Moisture Analyzer uses a controllable direct drying method to measure the change in sample mass by setting the temperature and atmosphere. It consists of a temperature control system, gas circuit module, and weighing module. The sample is placed in a furnace with precise temperature and atmosphere control, and the weighing module provides real-time mass data to determine constant weight conditions and calculate results.
Features
The analyzer can test up to 26 samples per batch with a large capacity crucible (35ml). It features a sample loader of appropriate height and a large touch screen for easy operation. The process is automated, allowing operators to manage multiple devices simultaneously. It includes a built-in balance for continuous real-time weighing and offers flexible settings for temperature, atmosphere, and report content. A thermal weight curve can be displayed in real-time, ensuring accurate and precise results.
Specifications
  • Model: TGM226
  • Max. Sample Loading: 26
  • Sample Mass: 0-30g
  • Balance Resolution: 0.0001g
  • Temperature Range: 0-200°C
  • Temperature Control Precision: ±2°C
  • Temperature Ramp Rate: 0-30°C/min
  • Temperature Control Stage: 20
  • Gas Supplied: Air/Nitrogen
  • Gas Flow: 0-20L/min
  • Overheating Protection: Double thermocouples
  • Power Supply: Single phase, AC220±10%, 50/60Hz, ≤1kW
  • Net Weight: 50kg
  • Dimensions: 700×600×450mm
Sample Application
The analyzer is suitable for various samples, including foods, feed, herb medicine, coal, and soil. Each sample type has specific parameters for sample weight, heating time, constant temperature, incubation time, and constant weight conditions to ensure accurate moisture analysis.
See more

Catalog excerpts

TGM226 Automatic Moisture Analyzer-1

5E Series Proximate Analyzer Automatic Moisture Analyzer Working principle TGM226 Automatic Moisture Analyzer adopts the controllable direct drying method, and measures sample mass change under this environment by setting the temperature and atmosphere, and obtains the test result of the sample moisture. It is mainly composed of temperature control system, gas circuit module and weighing module. The test sample is placed in the furnace with accurate temperature control and atmosphere provided by temperature control system and gas circuit module. The weighing module provides the real-time mass data during drying process and automatically determine constant weight conditions and calculate the results. Features Big capacity for batch testing Available to test 26 samples per batch, big capacity crucible(35ml) can be applied to test a variety of samples. Easy operation The sample loader of a proper height and large touch screen make operation easier and more convenient. High Automation and efficiency After loading, the procedure will carry on as required to control temperature change, weigh sample, and calculate constant weight and test result. Operator can walk away and control more than one equipment simultaneously.

 Open the catalog to page 1
TGM226 Automatic Moisture Analyzer-2

1. Built-in balance can weigh samples continuously in real time without cooling down. External balance is available for option as well. 2. Flexible setting function, capable to set temperature, atmosphere and report content as required. 3. Thermal weight curve (temperature -mass changes curve) can be drawn and display in real time. Avoid human mistakes of traditional method in weighing, transmission and calculation through selfinspection, the analyzer guarantees a more accurate and precise test result. Sample Application Classification Constant temperature(°C) Incubation time(min) Constant weight...

 Open the catalog to page 2

All CKIC / Changsha Kaiyuan Instruments Co., Ltd catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.