Mercury Determination for Biomass
1 /3Pages

Mercury Determination for Biomass

Mercury Determination for Biomass
1 /3Pages

Catalog excerpts

Mercury Determination for Biomass-1

Mercury Determination for Biomass Sample Preparation Boats should be pre-baked at 550°C for an hour or analyzed (without a sample) before loading a sample Samples may be analyzed as air dried Crush and grind to make the sample size <0.2mm Instrument 5E-HGT2320 Automatic Mercury Analyzer Instrument Settings Oxygen : Flow 300ml/min; purity >99.9% Pressure :0.1 Mpa Main Furnace Drying Temp.: 300°C Time: 90s Combustion Temp.: 800C Time: 180s Catalyst Furnace Temp.: 600C Purge Time: 60s Nominal Sample Mass: 50mg~70mg Analysis Time: 6 minutes/per test

 Open the catalog to page 1
Mercury Determination for Biomass-2

Calibration Standard Standard Mercury Solution or equivalent Certified Reference Materials. Method Approval GB 16659-2008 / ASTM D6722-11 Procedure 1. Prepare instrument as outlined in the operator’s instruction manual, (check the Inlet pressure, Outlet2 pressure, Furnace temperature, Catalytic tube temperature and Voltage.) 2. Keep the instrument in “prepare” state for at least half an hour 3. Do three blank samples test to reduce the blank level; (if the blank LC_peak is higher than 0.001, more blank sample is suggested) 4. Weigh one standard sample (60mg) in nickel boat, input the sample weight...

 Open the catalog to page 2
Mercury Determination for Biomass-3

8. Repetition of point 6 to point 8 until all samples are processed Note: Do Calibrating refer to operator' s instruction manual for details. Typical Results

 Open the catalog to page 3

All CKIC / Changsha Kaiyuan Instruments Co., Ltd catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.