Overview
The 5E-IRSII Automatic Infrared Sulfur Analyzer is designed to measure the total sulfur content in coal and coke using infrared absorption. It is widely used in various industries such as power plants, coal mines, metallurgy, and chemical industries.

Standard Configuration
  • Computer
  • Main analyzer
  • Standard Reference Material (GBW)
  • A/C adapter
  • H2O sorb reagent
  • Silica wool
  • Crucibles Boat stop
  • O-ring kit
  • Tool kit
  • Outer and Inner combustion tubes
  • Unique 'Quick Start' Button

Key Features
  • Stability and Accuracy: Utilizes a top-quality ultra-low drift infrared cell for precision.
  • Reliable Heating: Features single Si-C spiral tube heating components.
  • Gas Tightness: Unique structure to prevent SO2 interference from air.
  • Easy Operation: Automatic sample mass transfer to computer and a 'Quick Start' button for simplified operation.

Specifications
  • Model: 5E-IRSII
  • Conforms to ASTM D4239, ISO19579, GB/T25214
  • Max. Sample Loading: 1 sample per batch manually
  • Analysis Method: Infrared absorption
  • Analysis Resolution: 0.001% or 10ppm
  • Sulfur Range: 0.01%-30% (customizable)
  • Analysis Time per Sample: ≤120s
  • Analysis Temperature: 1300°C
  • Temperature Control Precision: ± 1°C
  • Sample Mass: 200mg-400mg
  • Power Supply: Single phase, AC220±10%, 50/60Hz, ≤4kW
  • Net Weight: 60kg
  • Dimensions: 540mm x 700mm x 610mm

Test Data
  • Sample Name: Control 6H0160-1 and 6H0160-2
  • Sample Weight: 0.3092g and 0.3077g
  • Average Sulfur Content: 2.624%
  • Reference Value: 2.625%
  • ASTM D4239-10 Repeatability Limit: 0.099
  • ASTM D4239-10 Reproducibility Limit: 0.256
  • Repeatability: 0.015
  • Reproducibility: 0.001

Conclusion
The 5E-IRSII Infrared Sulfur Analyzer exceeds ASTM precision requirements, ensuring reliable and accurate sulfur content analysis.
See more

Catalog excerpts

CKIC 5E-IRSII Infrared Sulfur Analyzer Sulfur analyzer / coal / combustion / infrared absorption-1

5E Series Elemental Analyzer Automatic Infrared Sulfur Analyzer Standard Configuration Computer Tool kit Main analyzer A/C adapter H2O sorb reagent Silica wool Crucibles Outer combustion tube Inner combustion tube Boat stop O-ring kit Standard Reference Material(GBW) Unique ”Quick Start” Button Application 5E-IRSII Infrared Sulfur Analyzer is used to determine the total sulfur content in coal, coke by infrared absorption, which is widely applied in power plants, coal mines, metallurgy, chemical industry, commercial inspection, scientific research, etc. Features Stability and Accuracy 1. Top quality ultra-low drift infrared cell to ensure stability, precision and accuracy. 2. Reliable single Si-C spiral tube heating components. 3. Unique gas tightness structure to avoid the effect of SO2 in air. Easy Operation 1. The sample mass can be automatically sent to the computer by balance connection. 2. Upgraded gas circuit design and reliable components to minimize the maintenance work. 3. Unique “Quick Start” Button to simplify the operation

 Open the catalog to page 1
CKIC 5E-IRSII Infrared Sulfur Analyzer Sulfur analyzer / coal / combustion / infrared absorption-2

Test Data Sample Name Sample Weight Reference Value ASTM D4239-10 Repeatability Limit (r) ASTM D4239-10 Reproducibility Limit (R) Conclusion: 5E-IRSII Infrared Sulfur Analyzer exceeds the ASTM precision requirement Specification Model 1 sample per batch manually Analysis Method Infrared absorption Analysis Resolution Sulfur Range 0.01%-30% customized range available Analysis Time per Sample Analysis Temp Temp. Control Precision* Sample Mass 200mg-400mg depends on sample condition Power Supply Dimensions (L×W×H) * Varies with age of heating element

 Open the catalog to page 2

All CKIC / Changsha Kaiyuan Instruments Co., Ltd catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.