pH / ORP ANALYZER - 1

Continuously measures pH and/or ORP with temperature readout Patented polymer gel double junctionђ > pH/ORP Automatic temperature compensation Includes clear constant-head flowcell Multiple analog and digital outputs Modbus communication (RS485 & TCP/IP) Datalogging (1 year at 15 min. intervals) Graphical trend of last 60 logged values Detailed sensor diagnostic alarms Capability to add on second pH/ORP sensor Automatic calibrator SMS text alarms Low flow detector PID control capability Remote access via internet > Extended life by 50% Intuitive menu and programming functions > Faster response and better stability Lower maintenance requirements > Online pH/ORP monitoring and control for: Water Treatment Wastewater Treatment De-Ionized Water Boiler Water > Chemtrac Systems, Inc.6991 Peachtree IndustrialBoulevard, Building 600Norcross, GA 30092 USA PH: 770.449.6233 US: 800.442.8722 FX: 770.447.0889 www.chemtrac.com At the heart of the PHO6300 pH and ORP Analyzer is a K-series, polymer filled, glass electrode. The innovative design makes the electrodes more responsive and sustain a longer life than other electrodes. Unlike many pH analyzers, you don't need to search through dozens of pH electrodes to find one that works in your process. The K-series electrode works in all environments from pH 2 to 12, and in conductivity from 0 to 10,000s. To help greatly simplify maintenance, the PHO6300 also offers self-cleaning and auto-calibration. >

Open the catalog to page 1
pH / ORP ANALYZER - 2

pH Sensor Probe Analyzer Type: W-type rod membrane pH electrodeReference Type: PatentedpH Range: 0-14 Slope: 95-102%Pressure Range: 0 - 45 psi (0 - 3 bar) Long Term Stability (drift): < 0.01 pH/hour Reproducibility: < 0.01 pHEO: -25 mV to +25 mVElectrode Resistance: 100-500 megaohms Response Time: 95% of step pH2 to pH12 < 5 secTemperature Range: 32 - 104а F (0 - 40а C) Cable Length: 20 ft. (6 m) Any length optionalShelf Life: 12 monthsTemperature Compensation: Automatically by integrated thermistor PT100 (ATC)Estimated Life: 12 - 18 months application dependent Power: 110 - 240 VAC 24 VDC...

Open the catalog to page 2

All Chemtrac catalogs and technical brochures

  1. UVP1000

    1 Pages

  2. UVM5000

    2 Pages

  3. PC6

    2 Pages

  4. PC5000

    2 Pages

  5. PC4400

    2 Pages

  6. PC3400

    2 Pages

  7. PC3

    2 Pages

  8. HydroACT 600

    2 Pages

  9. CCA3100

    2 Pages

  10. PM2500

    2 Pages

  11. UV1

    2 Pages