Benefits
  • Reliable and accurate measurement of particles down to parts per trillion (ppt) levels.
  • Advanced firmware and menu functions eliminate the need for external software for setup and calibration.
  • User-friendly calibration routine reduces maintenance costs.
  • Multiple display modes including particle size channels, total counts, and particulate concentration.
  • External sensor design keeps water away from electronics, ensuring easy cleaning.
Applications
  • Boiler/Condensate systems
  • Water and Wastewater Treatment
  • Reverse Osmosis Pretreatment
  • Membrane Filtration
  • Parts Washing
  • Food & Beverage
  • Pharmaceutical
Description

The PC4400 is designed for online monitoring of steam cycle corrosion product transport and filter performance. It can detect particulate concentrations as low as 0.01 ppb, providing early detection of filter breakthrough. This technology enhances water quality monitoring beyond traditional methods like turbidimeters.

Standard Features
  • Particle counting from 2-750 microns and sizing from 2-125 microns.
  • Displays particle concentrations in parts-per-billion (ppb).
  • Supports up to 8 user-programmable size ranges.
  • Large display with 8-channel readout and graphical trending.
  • MODBUS RTU communication and laser condition readout.
Optional Features
  • TracWare data acquisition software.
  • Analog outputs, inputs, and alarm relays.
  • Multiple communication protocols.
Technical Specifications
  • Laser Type: Solid-state laser diode (780 nm).
  • Detection Range: 2-750 microns.
  • Flow Rate: 75 mL/min.
  • Sample Temperature: 32° - 120° F (0° - 50° C).
  • Resolution: Better than 10% at 10 microns.
  • Data Storage: 65,000 sample strings with user-defined logging intervals.
  • Power Requirements: 100-240 VAC, 200mA, 47 - 63 Hz.
General Specifications
  • Operating Temperature: 32° - 120° F (0° - 50° C).
  • Dimensions: 12” W x 11.5” H x 5.25” D.
  • Weight: 7 lbs (3.2 kg).
See more

Catalog excerpts

PC4400-1

Particle Counter SCHEMTRAC PARTICLE COUNTER PC4400 Corrosion Product / Steam Cycle Contaminant Detection Series STANDARD FEATURES • Counts particles in liquid from 2-750 microns • Sizes particles from 2-125 microns • Parts-per-billion (ppb) display for particle concentrations* • Reports total counts, counts/mL, or counts/100mL • Up to 8 user-programmable size ranges • Large display with 8-channel readout & graph • User friendly, menu driven calibration • MODBUS RTU communication • Laser & cell condition readout (0-100%) • External sensor with sapphire optics OPTIONAL FEATURES • TracWare data acquisition software • Analog Outputs, Inputs, & Alarm Relays • Multiple communication protocols supported BENEFITS • Provides a reliable and accurate measurement of particles down to ppt levels • Advanced firmware and menu functions eliminates need for external software to perform setup and calibration • User friendly calibration routine makes calibration fast, simple, and lowers maintenance cost • Display modes include: all 8 particle size channels at once; total counts; ppb (Insoluble Material Volumetric Concentration) of particulate > 2 microns; graph of last 64 readings; and other sensor performance info • External sensor is easy to clean and water is kept away from electronics APPLICATIONS Boiler/Condensate Water Treatment Wastewater Treatment Reverse Osmosis Pretreatment DESCRIPTION The PC4400's performance reliability, ease of use, and measurement capability make it an ideal choice for online steam cycle corrosion product transport monitoring, and online filter performance monitoring and optimization. This is most clearly demonstrated in the PC4400's ability to report as low as a 0.01 ppb level of IMCV>2 (Insoluble Material Volumetric Concentration of particulate > 2 microns) in boiler/condensate sytems, or in the case of water treatment applications, being able to detect filter breakthrough often hours in advance of any turbidity (NTU) change. Close to two decades of customer experience with Chemtrac's Particle Counters has proven the effectiveness of this technology to take water quality monitoring to the next level. Chemtrac's PC4400 is the right choice if the optimization demands of your application are not being met by relative and less sensitive measurements of water quality like those provided by turbidimeters or SDI test systems. Chemtrac has experience with many applications that require the measurement of liquid borne particles. Our team of professionals can help with just about any application. Particle Detection Membrane Filtration Parts Washing Food & Beverage Pharmaceutical

 Open the catalog to page 1
PC4400-2

Laser Type: Cell Material: Viewing Windows: Detection Range: Sizing Range: ppb Range: Flow Rate: Sample Temperature: Resolution: Coincidence Limit: Signal to Noise Ratio: Size Channels: Alarms: Laser Diode Life: Measurement Type: Solid-state laser diode (780 nm) Nituff coated aluminum PEEK plastic (optional) Sapphire 2-750 microns 2-125 microns 0.01 - 10,000 ppb* 75 mL/min 32° - 120° F (0° - 50° C) Better than 10% at 10 micron (ASTM-F658) 20,000 cnts/mL (2 micron) Better than 5:1 8 - user selectable and total counts Sensor diagnostic, particle count limit MTBF 75,000 hours @ 55° C Light obscuration,...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.