1. Catalogs
  2. CETA Testsysteme GmbH
  3. Volume Flow Tester CETATEST 925

Volume Flow Tester CETATEST 925

Volume Flow Tester CETATEST 925

Volume Flow Tester CETATEST 925

Product catalog summary
System Specifications
The industrial PC features a Quad Core CPU at 1.8 GHz, 4 GB RAM, and a 128 GB SSD. It includes a micro-controller system for test process control with a 32-bit ARM μC at 84 MHz and a fast 24-bit A/D converter for real-time processing of measurement signals.
Pressure Sensors and Regulators
The system uses gauge and differential pressure sensors. The electronic pressure regulator (EPR) offers pressure ranges from -1 bar to 6 bar, while the mechanical pressure regulator (MPR) ranges from -1 bar to 6 bar, with other ranges available on request. Maximum deviations for gauge pressure are 1% and for volume flow 2%, both referred to full scale value.
Flow Tester CETATEST 925
The CETATEST 925 is designed for volume flow measurement using compressed air. It employs laminar flow elements (LFE) to measure pressure differences, which are directly proportional to volume flow. The device supports both direct and indirect measurement methods and offers extended applications through electronic pressure regulation and flow calibrations.
Test Options and Functions
Standard test options include direct method and nominal pressure optimization. With EPR, additional functions like prefill, pulsing, and smooth filling are available. Optional features include indirect method, bypass, program series, and exponential extrapolation. The device features a 7-inch touch screen, dual jog dial, and password-protected user levels.
Interfaces and Compatibility
The CETATEST 925 includes interfaces such as RS-232, USB, Ethernet, and optional Profibus DP, Profinet, and EtherCAT. It is fully compatible with CETATEST 915 and supports Industry 4.0 requirements. The device can store over 1 million measurement results and 256 test programs.
Power and Pneumatical Connections
The test device operates at 24 VDC with a maximum power consumption of 50 W. It requires a compressed air supply at least 1 bar above test pressure, with quality according to ISO 8573-1:2010 [1: 4: 1]. Pneumatical connections include a 6 mm plug-in fitting for input and options for up to three pneumatically driven outputs.
Calibration and Warranty
All devices come with a DAkkS-accredited calibration certificate. Special calibrations are available for different test pressures. The warranty is 3 years with yearly maintenance, extendable to 5 years.
Accessories
Optional accessories include a UL-certified power supply, filter combination, master jet, D-Sub connector, leak-tight valve, RS-232/Ethernet adapter, CETA Soft 2G software, and more.
See more

Catalog excerpts

Volume Flow Tester CETATEST 925-1

Flow Tester CETATEST 925 - for volume flow measurement The CETATEST 925 test device is suitable for flow testing using compressed air as test medium. The measuring method uses the fact that in a laminar flow, the pressure difference between two measuring points is directly proportional to the volume flow. Laminar flow elements (LFE) are used to laminarise the air flow. The pressure difference is measured with a differential pressure sensor. In the direct method, the air flow supplied by the pressure regulator first flows through the LFE and then through the test part. In the indirect method, the LFE is located in the outlet flow of the test part. The use of an electronic pressure regulator and flow calibrations at different LFE pressures results in an extended range of applications. System | Signal processing Industrial PC: Quad Core CPU, 1.8 GHz, 4 GB RAM, 128 GB SSD. Micro-controller system for the test process control: 32-bit ARM pC/ 84 MHz. Fast 24-bit A/D converter, real-time processing of measurement signals Pressure sensors Gauge pressure sensor (test pressure) Differential pressure sensor (pressure difference at LFE) Pressure ranges Electronical pressure regulator (EPR): -1 bar, -1/+1 bar, -1/+6 bar, 200 mbar, 1 bar, 6 bar Mechanical pressure regulator (MPR): -1 bar, -1000 mbar/+150 mbar, 150 mbar, 1 bar, 6 bar, Other ranges on request Maximum Deviations Specification for combined devices based on positive gauge pressure Volume flow: 2 % * Result units ml/min, ml/h, l/min, l/h, mbanl/s Measurement ranges of the laminar flow elements Type of LFE Flow at positive gauge pressure Test connections * referred to full scale value

 Open the catalog to page 1
Volume Flow Tester CETATEST 925-2

Leak Tester Flow Interface for optional extensions Test output Option: bypass Optional module for additional interfaces such as Profibus DP, Profinet, EtherCAT or USB (Slave) Controlled pressure output (option) Digital signal inputs RS-232 interface USB Host Ethernet Power supply Test options | Additional functions Standard: Direct method, nominal pressure optimisation With EPR: prefill, pulsing, smooth filling Optional: Indirect method, bypass, program series, exponential extrapolation, free programmable control valves Display | Operation 7-inch touch screen, dual jog dial, separate start/stop-buttons,...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.