1. Catalogs
  2. CAMOZZI AUTOMATION
  3. Integrated valve manifold for gas generator EN

Integrated valve manifold for gas generator EN

Integrated valve manifold for gas generator EN

Integrated valve manifold for gas generator EN

Product catalog summary
Overview: The document describes an integrated valve manifold system used in gas generators, specifically oxygen concentrators. These systems are primarily used in electro-medical applications and can be customized for specific requirements, including clean room environments.
System Components:
  • Molecular Sieves: Two sieves are used to separate oxygen, nitrogen, and water from air.
  • Compressor: Pumps air into the sieve columns, increasing the pressure to facilitate gas separation.
  • Control Unit: Configurable to provide continuous oxygen supply, such as for respiratory therapy.
  • Regulator: Reduces and controls the pressure of oxygen at the output port.
Functionality: The concentrator increases the oxygen concentration from approximately 21% to 90% by filtering atmospheric air and compressing it. The compact manifolds allow for fast switching and high flow rates, ensuring reliability and longevity.
Technical Specifications:
  • In-line Poppet Valve 2x3/2 Solenoid Valves:
    • Pressure: 1 to 3 bar
    • Media: Air/Oxygen
    • Flow: kv = 13 l/min, 600 Nl/min air at 2.5 bar free flow, 510 Nl/min air at 2.0 bar free flow
    • Voltage: 6-12-24 V DC
    • Power Consumption: 2x1 Watt
    • Cycles: On/off 7 seconds, life 30,000 hours (5 years), 20 million cycles
  • In-line Poppet Valve 4x2/2 Solenoid Valves:
    • Pressure: 1 to 6 bar
    • Media: Air/Oxygen
    • Flow: kv = 10 l/min, 460 Nl/min air at 2.5 bar free flow, 400 Nl/min air at 2.0 bar free flow
    • Voltage: 6-12-24 V DC
    • Power Consumption: 4x1 Watt
    • Cycles: On/off 5 seconds, life 30,000 hours (5 years), 20 million cycles
Additional Features: Other manifold functions are available upon request, and the system includes components such as a flip block sieve control, inlet filter, muffler, reservoir tank, and flow meter.
Conclusion: The integrated valve manifold system is designed for high efficiency and reliability in oxygen concentration, suitable for various medical applications.
See more

Catalog excerpts

Integrated valve manifold for gas generator EN-1

INTEGRATED VALVE MANIFOLD FOR GAS GENERATOR Air is a mixture of oxygen and nitrogen; an oxygen concentrator separates these two gases and increases the proportion of oxygen from about 21% to about 90%. Atmospheric air is trapped by the concentrator and filtered. A compressor then raises the pressure of the air. The concentrator consists of two molecular sieves, a compressor and a control unit The compact manifolds for gas management feature an on/off or proportional fluid control function which can be used for fast switching and high flow rates while assuring high reliability and long life. Typically used in electro-medical applications, these systems can be customised to meet detailed specifications or requirements, including for use in a clean room environment. The control unit can be set to provide, for example, a continuous oxygen supply for respiratory therapy In use the compressor pumps air into the two sieve columns where oxygen, nitrogen and water are separated out from one another A regulator’s function is to reduce and control pressure of the oxygen on the output port

 Open the catalog to page 1
Integrated valve manifold for gas generator EN-2

Manifold integrated solution Flip block sieve control F Inlet filter Valves (passive) Flow meter Reservoir tank In-line poppet valve 2x3/2 solenoid valves To sieve General data Pressure Media Flow Voltage Power consumption Cycles 1 ÷ 3 bar air / oxygen kv =13 l/min 600 Nl/min air at 2.5 bar free flow 510 Nl/min air at 2.0 bar free flow 6-12-24 V DC 2x1 Watt on/off 7 seconds life 30,000 hrs (5 years), 20 Mcycles Other manifold functions are available on request. In-line poppet valve 4x2/2 solenoid valves EXH 2 General data Voltage Power consumption Cycles 1 ÷ 6 bar air / oxygen kv =10 l/min 460...

 Open the catalog to page 2

Archived catalogs

  1. Air treatment

    234  Pages

  2. Handling

    132  Pages

  3. SERIES 63

    12  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.