Operation Manual Milliohmmeter RESISTOMAT® Model 2316
124Pages

{{requestButtons}}

Catalog excerpts

Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 1

OPERATION MANUALMilliohmmeter RESISTOMAT® Model 2316 © 2024 Manufacturer: burster praezisionsmesstechnik gmbh & co kg burster Talstr. 1 - 5 P.O.Box 1432 praezisionsmesstechnik gmbh & co kg 76593 Gernsbach 76587 Gernsbach All rights reserved Germany Germany Software version V 201801 Valid from: 04.03.2024 Email: info@burster.com

Open the catalog to page 1
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 2

Exclusion of warranty liability for operating manuals All information in the present documentation was prepared and compiled with great care and reproduced subject to effective control measures. No warranty is provided for freedom from errors. We reserve the right to make technical changes. The present information as well as the corresponding technical data can change without notice. Reproduction of any part of this documentation or its processing or revision using electronic systems is prohibited without the manufacturer’s prior written approval. Components, devices and measured value...

Open the catalog to page 2
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 3

RESISTOMAT® Model 2316The measurement solution. EU-Konformitatserklarung (nach en iso/iec 17050-1:2010) EU-Declaration of conformity (in accordance with EN ISO/IEC 17050-1:2010) Name des Ausstellers: burster prazisionsmesstechnik gmbh & co kg Issuer’s name: Anschrift des Ausstellers: Talstr. 1-5 Issuer’s address: 76593 Gernsbach, Germany Gegenstand der Erklarung: Milliohmmeter RESISTOMAT® fur Fertigung und Labor Object of the declaration: Milliohmmeter RESISTOMAT® for Production and Laboratory Model number / type: Diese Erklarung beinhaltet obengenannte Produkte mit alien Optionen This...

Open the catalog to page 3
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 4

6.5.3.1 Handling inductive loads e.g. reactors, cables on reels, motors,

Open the catalog to page 4
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 9

RESISTOMAT® Model 2316 1 Safety instructions On the device RESISTOMAT® 2316 and in this manual the following symbols warn about risks: The following signal words are used in the operating manual according to the specified hazard classification. DANGER indicates a hazard with a high level of risk which, if not avoided, will result in death or serious injury. A WARNING WARNING indicates a hazard with a medium level of risk which, if not avoided, could result in death or serious injury. A CAUTION CAUTION indicates a hazard with a low level or risk which, if not avoided, could result in minor...

Open the catalog to page 9
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 10

IMPORTANT: Read the operation manual carefully before using the equipment, and keep for future reference. 2.1 Use Fast and accurate measurements of ultra-small resistances can be made using the RESISTOMAT® type 2316 milliohmmeter. With its rugged table-top case and membrane keypad, this instrument is designed for both laboratory use and harsh industrial environments. Temperature-compensated resistance-testing of wires and coils is possible using a Pt 100 sensor or pyrometer to measure the temperature of the device under test. The instrument then corrects the resistance to e.g. 20 °C...

Open the catalog to page 10
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 11

3.1 Unpacking the unit The instrument weighs 3.5 kg and is packaged accordingly to protect against shock. Unpack the instrument carefully and verify that all items are present. This normally includes: 1 RESISTOMAT® model 2316 milliohmmeter 1 power lead 1 copy of this manual Inspect the instrument carefully for damage. If you suspect that the instrument has been damaged during shipping, notify the delivery company immediately. The packaging should be retained for examination by a representative of the manufacturer and/or the delivery company. The RESISTOMAT® model 2316 should be shipped only...

Open the catalog to page 11
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 12

3.4 Power supply and signal-lead connectors USB interface Measurement input either via 5-pin bayonet socket or 4x laboratory safety sockets (4mm ø) Mains fuse Ethernet interface Power switch Mains socket RS232 interface IMPORTANT: additional protection of the measurement current using a super quick-acting 10 A fuse 6.3 x 32 mm, 600 VAC, 50 kA breaking capacity (or greater) RS Components #209-9406 (in Germany) Only use this fuse NOTICE • Use a suitable connecting cable with a dual-shield construction (aluminum foil cladding plus braided shield) for the communications port connection and the...

Open the catalog to page 12
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 13

RESISTOMAT® Model 2316Block diagram c <D .Q 3 CO <u ®

Open the catalog to page 13
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 14

RESISTOMAT® Model 2316 3.6 Setup and installation • Ensure that there is an adequate supply of air to prevent heat building up in the instrument. • Do not place the instrument on surfaces such as carpets or cloths, or near materials such as curtains or wall hangings that could prevent the air circulating. • Do not place the instrument at an inclined angle. It should always be used in a horizontal position. • Keep the instrument away from apparatus, equipment, machines and installations that generate strong magnetic fields. • Do not place heavy objects on the instrument. • Condensation can...

Open the catalog to page 14
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 15

RESISTOMAT® Model 2316 3.10 Safety instructions A DANGER Whilst the hardware and software has been developed and tested in accordance with the state of the art, they cannot be guaranteed totally free of errors. Thus this instrument or part of this instrument must not be used to influence a control system from which risk to life or property can arise directly or indirectly without additional protection. Maintenance and repair work must only be performed by trained, competent technical personnel familiar with the associated risks. • The instrument has two measurement inputs connected in...

Open the catalog to page 15
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 16

4 Controls 4.1 Front panel RESISTOMAT 2316 Front panel with backlit LCD display and integral membrane keypad with tactile feedback [START] : In the measurement menu this button starts a measurement In the Configuration menu this button is assigned different functions depending on the text shown on the display above the button (soft key). [STOP] : In the measurement menu this button stops a measurement. In the Configuration menu this button is assigned different functions depending on the text shown on the display above the button (soft key). [ft] : In the measurement menu and for manual...

Open the catalog to page 16
Operation Manual Milliohmmeter RESISTOMAT® Model 2316 - 17

RESISTOMAT® Model 23164.2 Rear panel4.2.1 Description of connector socketsMeasurement input + U + I Analog GND - I - U PE (protective ground) potential burster model 9900-V172 Note: The current branch is protected by a fuse 6.3 x 32 [mm] 10AFF. (rear side of unit) Connector shell : PE (protective ground) potential Mating connector : burster model 4291-0 Two-wire technology is possible if the relevant conductors are joined together at the sensor. Note: NEVER connect the cable shield to the connector shell if the grounding at the sensor end is unclear. Otherwise, if there is a ground...

Open the catalog to page 17