Catalog excerpts

Data Sheet Precision Load Cells Series 85040 and 85070-1

Series 85040 and 85070 for compressive load, for tensile and compressive forces For static and dynamic forces High linearity from ± 0.1 % F.S. Very low sensitivity to lateral forces Models 85073 / 85075 suitable for extremely high dynamic stress, series 85070 up to 109 load cycles Extended temperature compensation range -55 °C to 120 °C (optional) Measuring ranges from 0 ... 20 N to 0 ... 2 MN Protection class IP68 (optional) These load cells feature an outstanding ability to withstand static and dynamic stress and have exceptional precision and service life. These products have a huge range of applications in industry, R+D and testing thanks to the optional IP68 degree of protection and low sensitivity to lateral forces. The force-sensitive diaphragm with fitted strain gauges is located between the central part, where the force is applied (force application), and the outer ring (containing fixing holes). Examples of applications would be the measurement of: ►► ►► ►► ►► Insertion forces Reference measurement Weights (e.g. silos, skips) Tensile forces (tension in cables, chains etc. with load centering plate) ►► Materials testing For the compressive force sensors (models 85043 and 85073) the force must be applied through a plane plate, hard enough for the range of forces being measured, or a piston. Technical changes reserved. All data sheets at www.burster.com Two support diaphragms are additionally fitted, above and below the actual measuring element, to desensitize the sensors to extraneous lateral forces or moments. (see „Permitted external forces“ on page 2). Models 85041/85075 (tension/compression) are calibrated in the tensile direction. The characteristic figure for the compression direction can nominally vary by ± 0.25 % from the figure for the tensile direction. The output signal is positive when the applied force is tensile. The 8507X series (models 85073 and 85075) is designed for the highest possible dynamic stress and service life. This is > 109 cycles from zero up to the full value of the measuring range in the tensile or compressive direction and tolerates a maximum operational force of up to 200 % of the nominal range. burster praezisionsmesstechnik gmbh & co kg Germany Talstr. 1-5 . Gernsbach 76593 . Phone +49-7224

 Open the catalog to page 1
Data Sheet Precision Load Cells Series 85040 and 85070-2

Dimensions Compressive load cell model 85043 - for static and dynamic operations Order Code Measurement Range Compressive load cell sensor model 85073 - also suitable for highest dynamic operations Order Code Measurement Range Dimensional drawings models 85043 and 85073 from 50 kN Permitted external forces As a result of the two stabilizing membranes with which these load cells have been designed, they only have very low sensitivity to forces that do not act centrally on the sensor. The effect of these undesirable external forces cannot be expressed in blanket figures; it depends on the sensor‘s...

 Open the catalog to page 2
Data Sheet Precision Load Cells Series 85040 and 85070-3

Dimensions Tensile and compressive load cell model 85041 - for static and dynamic operations Order Code Tensile and compressive load cell model 85075 - also suitable for high dynamic operations Order Code Number Natural of Holes Frequency in ø G [kHz] * These values are nominal values Dimensional drawing models 85041 and 85075 as well as 85041 and 85043 to 20 kN Notes: 1. A hole in the mounting surface allows force to be applied from below. If the sensor is mounted on a surface that has not been drilled, it is necessary to ensure that the threaded bolt that is screwed in from above does not press...

 Open the catalog to page 3
Data Sheet Precision Load Cells Series 85040 and 85070-4

Technical Data Order Information Electrical values Bridge resistance: full bridge (foil strain gauges) 350 Ω, nominal1 Calibration resistor: 59 kΩ ± 0.1 % The bridge output voltage caused by a shunt of this value is given in the calibration protocol. Excitation voltage: recommended 10 V DC or AC maximum 15 V DC or AC Nominal value: 2 mV/V, nominal1 Models 85041/85043, > 100 N: 3 mV/V, nominal1 Isolation resistance: > 109 Ω at 50 VDC 1 ) Deviations from the stated value are possible. Environmental conditions Range of operation temperature: - 55 °C ...120 °C Range of nominal temperature (compensated...

 Open the catalog to page 4