Underwater Corrosion Probes UCP1A & UCP1B Operating Instruction October 2022 English Please read carefully before use

Open the catalog to page 1
User Manual: UCP1A and UCP1B Probes - 4

Instruction Manual Description The UCP 1A and 1B contact probes are versatile instruments which are used to carry out under water cathodic potential measurements. The unit is designed for use with manned or remotely operated vehicles (ROVs). There are two types available for use: UCP1A - Contact Probe Operation of the UCP1A is carried out by way of a probe tip making contact with the surface under test, and reading the resulting voltage on a suitable high impedance (> 10M Ohms) or digital voltmeter. The UCP1A has two connections: • Black Wire - Stainless Steel Probe Tip • White Wire -...

Open the catalog to page 4
User Manual: UCP1A and UCP1B Probes - 5

Underwater Corrosion Probes Unpacking Remove the UCP from its packaging and check for damage. If any part is damaged, the carrier and supplier should be notified immediately. All packing material should be kept for inspection and the UCP should not be used. The package will contain one of the following items depending on which UCP probe that you purchased: UCP1A – Contact Probe, or UCP1B – Proximity Probe If the probe type is not what you expect please contact your supplier. Maintenance Replacement of electrode Unscrew the two recessed nylon screws on either side of the body. Carefully pull...

Open the catalog to page 5
User Manual: UCP1A and UCP1B Probes - 6

Instruction Manual Long Term Storage Flush well in clean seawater to remove any contamination from the Nose Cone and allow to dry. If salt encrustation is seen on the nose cone, remove the Ag/AgCl half-cell and briefly wash in fresh water but immediately after, thoroughly rinse in salt water (3% salinity). Recommissioning Soak the UCP for 2 hours in a seawater solution (3% salinity). Carry out verification checks (see page 7). Calibration There are two types of UCP probe, the 1A ‘Contact Probe’ and the 1B ‘Proximity Probe’.

Open the catalog to page 6
User Manual: UCP1A and UCP1B Probes - 7

Underwater Corrosion Probes Both types use the same type of replaceable silver/silver chloride/seawater half-cell. Verification of the Ag/AgCl Half-Cell Plastic container filled with 3% Salinity artificial seawater solution. K-series or K-SAT Reference Electrode. (Use a glass type or plastic screw in with an adaptor) Note 1: Remember to remove the plastic cap on the plastic reference electrode). Note 2: No bare wires should be exposed to the seawater.

Open the catalog to page 7
User Manual: UCP1A and UCP1B Probes - 8

Instruction Manual Screw the adaptor lead (6004-0018) onto the reference electrode (plastic screw on type). Submerge the UCP and reference electrode in the 3% seawater solution. Connect the white lead from the UCP to the Positive terminal of the DVM (see Fig. 3 on page 7). Connect the lead from the K-series or K-SAT reference electrode to the negative terminal of the DVM. Allow the electrodes to reach a stable temperature and potential (15 - 30 minutes). The voltage difference between the reference electrode and the internal silver/slver chloride electrode can be read directly off the DVM...

Open the catalog to page 8
User Manual: UCP1A and UCP1B Probes - 9

Underwater Corrosion Probes For the UCP1B connect the white lead from the UCP to the positive terminal of the mV meter and negative terminal of the meter to a zinc block. Note: The connection to the zinc block needs to be above the waterline. When the UCP1B is submerged in the water in close proximity to the zinc block the readings will be as shown for the UCP1A above. The UCP1B should also be soaked for 2 hours in the 3% seawater solution prior to taking any measurements. Product Variations The UCP1A is available with varying lengths of underwater cables as follows: Cable Length Part...

Open the catalog to page 9
User Manual: UCP1A and UCP1B Probes - 10

Instruction Manual Disposal information Producer registration number: WEE/HJ0051TQ This Product must be disposed of in accordance with UK WEEE Producer Responsibility Regulations, or in accordance with your local WEEE guidance. For further information on UK WEEE regulations click on: http://www.gov.uk/government/collections/producer-responsibility-regulations

Open the catalog to page 10
User Manual: UCP1A and UCP1B Probes - 11

Contact details Manufacturer: Address: Buckleys (UVRAL) Ltd Buckleys House Unit G, Concept Court Shearway Business Park Shearway Road Folkestone Kent CT19 4RG, UK Distributor details

Open the catalog to page 11
User Manual: UCP1A and UCP1B Probes - 12

Product registration Thank you for choosing a Buckleys product, we are sure it will provide you with many years of reliable service. Please register this product via Buckleys’ website and download the Warranty Registration Certificate. Register your product in 5 minutes Once your product is registered, you will receive the following benefits: • FREE annual service & calibration reminders by email Latest industry news relating to your product Be the first to hear about our new products We strive to improve the quality of our products and service. Registering your product helps us monitor...

Open the catalog to page 12