Quick start guide: Lab Pro2

Quick start guide: Lab Pro2
1 / 12 PagesView full catalog

Quick start guide: Lab Pro2

Product catalog summary
Safety Warnings
Individuals with electronic implants should avoid using the equipment. It is unsuitable for combustible atmospheres or ATEX Hazardous Areas. Only trained personnel should operate it, ensuring proper grounding to prevent electric shocks. The equipment may cause radio interference and ozone formation, necessitating adequate ventilation.

Declarations of Conformity
The Lab Pro2 complies with EU directives and UK legislation on electronic compatibility and safety, as declared by Buckleys (UVRAL) Ltd.

Specifications
The device operates on a 90-250VAC supply, with an output voltage range of 0.9kV to 40kV and a maximum output current of less than 1mA. It meets several test standards and has specific dimensions and weight.

Maintenance
Regular cleaning is advised, and annual calibration by an approved provider is required. Users should register their product for service reminders and updates.

Kit Contents
The kit includes the Lab Pro2 detector unit, test probe handle, interconnecting lead, electrostatic discharge wristband, calibration certificate, power supply cable, stainless steel drum brush electrode, earth cable, and pointed probe electrode.

Risk Assessment
Users must conduct a risk assessment before use, considering personal safety, explosion and fire risks, EMC and RFI risks, and other potential hazards. Proper grounding and safety measures are essential.

Factory Settings
Default settings include an output voltage of 10kV, alarm suppression set to 0, manual testing standard, sensitivity at 200µA, alarm volume at 5, and screen contrast at 5.

Setup
Instructions are provided for setting the instrument's language, output voltage, sensitivity threshold, alarm suppression, and test thickness via a standard. The setup process involves navigating the settings menu using specific buttons.

Instrument Testing Procedure
To ensure the instrument functions correctly, verify that the voltage is displayed below the set test thickness. Once settings are confirmed, test the instrument by pressing the ON button on the test probe handle and briefly touching the electrode to a conductive, grounded item. The alarm system should activate as indicated in Figure 9.

Establishing Test Conditions
Users should collaborate with clients to determine appropriate test conditions based on the material being tested. In the absence of specific customer requirements, the NACE SP0274 standard is recommended as a general-purpose guideline. However, it is the user's responsibility to ensure safe and reliable testing conditions.

Contact Information
Manufacturer: Buckleys (UVRAL) Ltd.
Address: Buckleys House, Unit G, Concept Court, Shearway Business Park, Shearway Road, Folkestone, Kent CT19 4RG, UK
Tel: +44 (0)1303 278888
Email: [email protected]
Website: www.buckleysinternational.com
See more

Catalog excerpts

Quick start guide: Lab Pro2-1

Quick Start Guide Please read this manual before using the equipment

 Open the catalog to page 1
Quick start guide: Lab Pro2-3

Safety Warnings IMPORTANT: We strongly advise that individuals with pacemakers, cochlear implants, or trans-dermal electronic implants e.g., glucose monitors do not use our highvoltage test equipment under any circumstances. IMPORTANT: This product must not be used in combustible atmospheres or ATEX Hazardous Areas, such as fuel stations or in proximity to any combustible gas (and/ or oxygen concentrations exceeding normal atmosphere). All high-voltage testing equipment must be operated by suitably trained, responsible, authorised personnel only. IMPORTANT: This product can cause radio interference...

 Open the catalog to page 3
Quick start guide: Lab Pro2-4

Quick Start Guide EC Declaration of Conformity We, Buckleys (UVRAL) Ltd., as sole manufacturer of the apparatus listed below, declare that the product Lab Pro2 is manufactured in conformity with the following directives: 2014/30/EU, 2014/35/EU, 2015/863/EU and 2011/65/EU (RoHS). Date: 01/07/2021 Authorised by: J P Hoveman CEO, Buckleys (UVRAL) Ltd. UKCA Declaration of Conformity We, Buckleys (UVRAL) Ltd., as sole manufacturer of the apparatus listed below, declare that the product Lab Pro2 is manufactured in conformity with the following UK legislation: Electronic Compatibility Regulations 2016,...

 Open the catalog to page 4
Quick start guide: Lab Pro2-5

Maintenance The Lab Pro2 is designed for minimum maintenance by the user, however the following periodic checks/actions are recommended: The exterior of the unit should be cleaned regularly. Stubborn dirt may be removed with a moistened cloth (99% water +1% mild detergent). Mild solutions may be used to remove greasy dirt. Do not allow cleaning fluid to enter the unit. The use of other cleaning agents may lead to damage to the detector’s surface areas. Our service team will ensure your detector’s performance is within our strict tolerances and carry out a thorough inspection of the entire kit;...

 Open the catalog to page 5
Quick start guide: Lab Pro2-6

Quick Start Guide Kit Contents Upon delivery of your Lab Pro2, please take the time to ensure that the kit is complete, and all components shown below are present and undamaged: Electrostatic discharge wristband Calibration certificate Mains power supply cable Stainless steel drum brush electrode 5m earth cable on reel Interconnecting lead Test probe handle and nosecone Lab Pro2 detector unit with desk stand Pointed probe electrode

 Open the catalog to page 6
Quick start guide: Lab Pro2-7

Risk Assessment It the user’s responsibility to complete a risk assessment before using Buckleys equipment. The following points offer some guidance but must not be assumed to be complete or sufficient. Personal safety • Have all users been trained in the correct and safe use of the instruments? • Are they aware that the instrument produces high voltages, and can give electric shocks if used incorrectly? • Is the instrument in good condition, undamaged? Is the instrument dry? Are the accessories similarly in good condition, dry and undamaged? • Is it possible to ensure that persons with pacemakers,...

 Open the catalog to page 7
Quick start guide: Lab Pro2-9

Setup (continued) Instrument settings Switch the instrument on by pressing a. The screen displays the following information (see fig.8): A Battery level indicator (not in use) B Menu symbol C Voltage currently applied to electrode D Current draw at electrode E Output lock ON/OFF F Sensitivity threshold G Set test voltage H Sensitivity threshold graph I Handle temperature Setting interface language Press b to access the main settings menu and press d or e to scroll to Set Language and press b. Press d or e to scroll to the desired language and press b. Press c to return to the main display screen....

 Open the catalog to page 9
Quick start guide: Lab Pro2-10

Quick Start Guide Basic Testing Press and hold the ON button

 Open the catalog to page 10
Quick start guide: Lab Pro2-11

Basic Testing (continued)

 Open the catalog to page 11
Quick start guide: Lab Pro2-12

Contact Details Manufacturer Address Buckleys House Unit G, Concept Court, Shearway Business Park, Shearway Road Folkestone, Kent CT19 4RG, UK Tel: +44 (0)1303 278888 Email: [email protected] www.buckleysinternational.com

 Open the catalog to page 12