Safety Warnings
The document highlights critical safety measures for using the Geomembrane Dielectric Withstand Testing Kit. It warns against use by individuals with electronic implants and in combustible environments. Only trained personnel should operate the equipment, ensuring it is properly earthed to prevent electric shocks. Adequate ventilation is necessary to avoid ozone exposure.
Declarations of Conformity
The Geo Pro2 adheres to EU and UK regulations, including directives on electronic compatibility and hazardous substances.
Specifications
The Geo Pro2 features an output voltage range of 0.9kV to 40kV, with a maximum output current of less than 1mA. It supports various testing standards and has a battery life of 8 hours.
Maintenance
Regular cleaning and annual calibration are recommended. Batteries should be removed and charged when not in use.
Product Registration
Registering the product online offers benefits such as service reminders and updates on new products.
Useful Advice
Guidance is provided on setting instrument sensitivity and managing alarm suppression to avoid false alarms.
Kit Contents
The kit includes a main detector unit, various electrodes, extension rods, a wristband, and a carry case. Users should verify all components are present and undamaged upon delivery.
Risk Assessment
Users must conduct a risk assessment before use, considering personal safety, explosion risks, and potential electromagnetic interference.
Factory Settings
Default settings include a 10kV output voltage and a sensitivity of 200µA. Users may need to adjust these settings for specific applications.
Setup
Instructions for setting up the instrument include turning it on and adjusting settings such as voltage and sensitivity.
Setting Interface Language
Access the main settings menu by pressing 'b', navigate to 'Set Language' using 'd' or 'e', select the desired language with 'b', and return to the main display with 'c'.
Setting Output Voltage
Enter the main settings menu with 'b', scroll to 'Adjust Voltage', adjust to the desired voltage using 'd' or 'e', confirm with 'b', and return to the main display with 'c'.
Setting Sensitivity Threshold
Access the settings menu, navigate to 'Adjust Sensitivity', adjust the threshold using 'd' or 'e', confirm with 'b', and return to the main display with 'c'.
Setting Alarm Suppression
Enter the settings menu, scroll to 'Alarm Suppression', adjust the setting using 'd' or 'e', confirm with 'b', and return to the main display with 'c'.
Setting Test Thickness via a Standard
Access the settings menu, navigate to 'Test Standard', select the desired standard, input the test material thickness, and observe the calculated test voltage. Test the instrument by holding the ON button and contacting a conductive item; the alarm should trigger.
Testing Conditions
Users should establish appropriate test conditions with clients, considering the material tested. In the absence of specific requirements, NACE SP0274 is recommended as a general-purpose standard, but users must ensure safe and reliable conditions.
Contact Information
Manufacturer: Buckleys (UVRAL) Ltd., Address: Buckleys House, Unit G, Concept Court, Shearway Business Park, Shearway Road Folkestone, Kent CT19 4RG, UK, Tel: +44 (0)1303 278888, Email: [email protected], Website: www.buckleysinternational.com
See more

Catalog excerpts

Quick start guide: Geo Pro2-1

Geomembrane Dielectric Withstand Testing Kit Quick Start Guide Please read this manual before using the equipment

 Open the catalog to page 1
Quick start guide: Geo Pro2-3

Safety Warnings IMPORTANT: We strongly advise that individuals with pacemakers, cochlear implants, or trans-dermal electronic implants e.g., glucose monitors do not use our highvoltage test equipment under any circumstances. IMPORTANT: This product must not be used in combustible atmospheres or ATEX Hazardous Areas, such as fuel stations or in proximity to any combustible gas (and/ or oxygen concentrations exceeding normal atmosphere). All high-voltage testing equipment must be operated by suitably trained, responsible, authorised personnel only. IMPORTANT: This product can cause radio interference...

 Open the catalog to page 3
Quick start guide: Geo Pro2-4

Quick Start Guide EC Declaration of Conformity We, Buckleys (UVRAL) Ltd., as sole manufacturer of the apparatus listed below, declare that the product Geo Pro2 is manufactured in conformity with the following directives: 2014/30/EU, 2014/35/EU, 2015/863/EU and .and 2011/65/EU (RoHS). Date: 01/07/2021 Authorised by: J P Hoveman CEO, Buckleys (UVRAL) Ltd. UKCA Declaration of Conformity We, Buckleys (UVRAL) Ltd., as sole manufacturer of the apparatus listed below, declare that the product Geo Pro2 is manufactured in conformity with the following UK legislation: Electronic Compatibility Regulations...

 Open the catalog to page 4
Quick start guide: Geo Pro2-5

Maintenance The Geo Pro2 is designed for minimum maintenance by the user, however the following periodic checks/actions are recommended: The exterior of the unit should be cleaned regularly. Stubborn dirt may be removed with a moistened cloth (99% water +1% mild detergent). Mild solutions may be used to remove greasy dirt. Do not allow cleaning fluid to enter the unit. The use of other cleaning agents may lead to damage to the detector’s surface areas. Our service team will ensure your detector’s performance is within our strict tolerances and carry out a thorough inspection of the entire kit;...

 Open the catalog to page 5
Quick start guide: Geo Pro2-6

Quick Start Guide Kit Contents Upon delivery of your Geo Pro2, please take the time to ensure that the kit is complete, and all components shown below are present and undamaged: Main detector unit Test probe handle and nosecone Interconnecting lead Electrostatic discharge wristband Calibration certificate 150mm stainless steel drum brush electrode 5m earth cable on reel D-cell rechargeable batteries x4 Battery charger with mains power adaptors and in-car charger* 450mm angled phosphor-bronze brush electrode Pair of knurled locknuts Pointed probe electrode Padded carry bag with shoulder strap...

 Open the catalog to page 6
Quick start guide: Geo Pro2-7

Risk Assessment It the user’s responsibility to complete a risk assessment before using Buckleys equipment. The following points offer some guidance but must not be assumed to be complete or sufficient. Personal safety • Have all users been trained in the correct and safe use of the instruments? • Are they aware that the instrument produces high voltages, and can give electric shocks if used incorrectly? • Is the instrument in good condition, undamaged? Is the instrument dry? Are the accessories similarly in good condition, dry and undamaged? • Is it possible to ensure that persons with pacemakers,...

 Open the catalog to page 7
Quick start guide: Geo Pro2-9

Setup (continued) Instrument settings Switch the instrument on by pressing a. The screen displays the following information (see fig.8): a Battery level indicator b Menu symbol c Voltage currently applied to electrode d Current draw at electrode e Output lock ON/OFF f Sensitivity threshold g Set test voltage h Sensitivity threshold graph i Handle temperature Setting interface language Press b to access the main settings menu and press d or e to scroll to Set Language and press b. Press d or e to scroll to the desired language and press b. Press c to return to the main display screen. Setting...

 Open the catalog to page 9
Quick start guide: Geo Pro2-10

Quick Start Guide Basic Testing Press and hold the ON button

 Open the catalog to page 10
Quick start guide: Geo Pro2-11

Basic Testing (continued)

 Open the catalog to page 11
Quick start guide: Geo Pro2-12

Contact Details Manufacturer Address Buckleys House Unit G, Concept Court, Shearway Business Park, Shearway Road Folkestone, Kent CT19 4RG, UK Tel: +44 (0)1303 278888 Email: [email protected] www.buckleysinternational.com